JetClean Operating Manual

Technical Reference Summary

Covered Models & Document Scope

Agilent JetClean Operating Manual (G7077-90052). Covers the JetClean system for GC/MS ion sources with two modes of operation: Acquire & Clean mode and Clean Only mode. Includes configuration, parameter setpoints, hydrogen supply plumbing, and troubleshooting. No model/size variants or ordering codes are printed.

Technical Data

Parameter Value Condition
Hydrogen flow (Clean Only mode), lower setpoint 0.13 mL/min p.20 ↗ Clean Only mode
Hydrogen flow (Clean Only mode), upper setpoint 3.52 mL/min p.20 ↗ Clean Only mode
Hydrogen flow step size 0.069 sccm p.20 ↗ Clean Only mode
Emission current, lower setpoint 10 µA p.20 ↗ Clean Only mode
Emission current, upper setpoint 35 µA p.20 ↗ Clean Only mode
Clean duration, lower setpoint 1 min p.20 ↗ Clean Only mode
Clean duration, upper setpoint 120 min p.20 ↗ Clean Only mode
Source temperature, lower setpoint 150 °C p.20 ↗ Clean Only mode
Source temperature, upper setpoint 350 °C p.20 ↗ Clean Only mode
Hydrogen flow (Acquire & Clean mode), lower setpoint 0.13 mL/min p.20 ↗ Acquire & Clean mode
Hydrogen flow (Acquire & Clean mode), upper setpoint 0.49 mL/min p.20 ↗ Acquire & Clean mode

Operating Limits

  • Hydrogen flow range (Clean Only mode): 0.13 to 3.52 mL/min p.20 ↗
  • Emission current range: 10 to 35 µA p.20 ↗
  • Clean duration range: 1 to 120 min p.20 ↗
  • Source temperature range: 150 to 350 °C p.20 ↗
  • Hydrogen flow range (Acquire & Clean mode): 0.13 to 0.49 mL/min p.20 ↗