A 100 L — instruction manual
Technical Reference Summary
Covered Models & Document Scope
Pfeiffer Vacuum A 100 L / A 100 L ES multi-stage dry Roots pump operating instructions (EN, 62 pp.). Covers part-number families A100L3XXXX and A100L4XXXX (load-lock/transfer chamber pumping) and A100L5XXXX (A 100 L ES model with 'Energy Saving' option reducing energy consumption by ~50%). Document covers all models/variants; features vary per unit.
Section-by-Section Technical Breakdown
Compact multi-stage dry Roots pump for load-lock/transfer chambers (semiconductor, photovoltaic) and turbomolecular roughing; no preventive maintenance required. HMI includes indicator lights, time counter, Loc/Rem switch, ON/OFF circuit breaker, remote and RS-232/RS-485 connectors, water/N2 connections. Electronic monitoring handles temperature alarm/warning.
- A 100 L ES 'Energy Saving' option cuts energy consumption by 50%, auto-activated per pumping conditions.
- Scope of delivery: pump, 2 water connectors, power cable with female connector, socket mounting flange, 3× CHc M8x16 screws + washers, lifting-hook guide, manual.
- Pump delivered with oil charge filled; do not modify oil level or drain (overhaul-only operation).
Horizontal installation on feet with spirit level; frame attachment with ≥4 M10x25 grade 8-8 screws at 40 N·m. Two pumps can be stacked using brackets (M8x16 screws, 22 N·m); seismic installation uses 4 anchored brackets (40 N·m fixing screws, only lower pump anchored when stacked). Water and nitrogen circuits have defined connector types; pumping-line best practice covers isolation valves, flexible tubes and leak testing.
- Frame fixing: ≥4 M10x25 grade 8-8 screws, torque 40 N·m.
- Stacking screws: 3× CHc M8x16 + washers, torque 22 N·m.
- Seismic brackets: 4 pcs at 'SEISMIC TIE DOWN' label positions, supplied screws torque 40 N·m; floor bolts are customer-supplied.
- Water inlet = female connector, outlet = male connector; use regulated (softened) water; filter inlet if solid pollution spec cannot be met.
- Nitrogen tubing: 6 mm OD / 4 mm ID pipe; N2 IN port is 1/8 NPT; CDA may replace N2 if pumped gases are inert.
- Water max pressure label: 101 PSI (7 bar).
- Connected chambers must withstand 1·10⁻³ hPa absolute negative pressure.
- Inlet line: as short/straight as possible, ID ≥ inlet flange ID.
- Exhaust line components must be rated above max pump-generated pressure.
Both connectors provide start/stop logic input, 0–10 V rotation frequency setpoint (0 V = max frequency, 10 V = min, requires Rem mode) and NO dry-contact status outputs. Contacts rated max 50 V / 1 A, must be wired per SELV rules. State tables map contact closures to pump running/stopped/warning/alarm conditions.
- 16-pin: S1 (1-2) start/stop (24 VDC = start, 0 V = stop); A2 (15-16) speed setpoint.
- 16-pin outputs: 3-4 pump status, 7-8 warning, 9-10 alarm, 11-12 reserved, 13-14 isolation valve authorization.
- 19-pin: S1 (A-B) start/stop (15–24 VDC = run); A2 (R-S) setpoint; +15 VDC (25 mA) and 15–24 VDC supplies available.
- 19-pin outputs add second warning contact (G-H) and pump failure contact (T-U).
- Contacts open on fault (NO logic); max 50 V / 1 A.
Pump has ON/OFF circuit breaker, motor thermal protection and 2× 0.1 A/250 VAC semi time-lag fuses (not user-replaceable). Customer must provide a breaker per IEC 60947-2 curve D with ≥10 kA short-circuit capacity within 7 m and in line of sight, plus a differential breaker and CPC earth conductor. Pump has no built-in emergency stop; host equipment EMS must switch it off.
- Restart procedure: breaker OFF, fix fault, wait ~15 s, breaker ON.
- Customer-supplied power cable; rating plate example: 200–230 V Δ, 50/60 Hz, 3 phases/4 wires, full-load current 12 A (illustrative).
- Earth: second protective conductor cross-section ≥ conductor cross-section; IEC 60417 #5019 terminal marking.
- Wait 5 minutes after power-off before work (capacitors charged to >60 VDC).
- EMS of host equipment must switch off the pump when activated.
Three control modes: LOCAL (breaker ON/OFF), REMOTE (dry contacts + analog setpoint via 16/19-pin), SERIAL LINK (RS-232/RS-485); selected by Loc-Rem switch. Commissioning checklist: verify inlet connection, unclogged exhaust, water/N2 circuits on, then customer main switch. A 100 L ES power reduction is automatic at ultimate pressure.
- Local start: breaker ON → pump auto-starts, green LED on, temperature monitoring active.
- Local stop: breaker OFF → pump auto-stops.
- Loc-Rem switch: left = 'Loc', right = 'Rem'.
Contains maintenance safety instructions, maintenance frequency tables, on-site maintenance limits, and the exchange procedure for replacement products: disconnection from installation, draining the water circuit and preparing the pump for shipping. Pump design requires no preventive maintenance; major work is via Pfeiffer service centers.
- No preventive maintenance by design; overhaul work done at service centers.
- Exchange: disconnect, drain water circuit, prepare for shipping.
Covers shutdown for longer periods, recommissioning steps and disposal of the product.
Fault indication via indicator lights with troubleshooting for three cases: pump does not start; pump running with indicator light on; pump stops running. Alarm reset is by turning off the power switch (per status label).
- Indicator lights: PUMP RUNNING, PUMP ALARM, HAZARD TEMP WARNING.
- Alarm reset: turn OFF power switch.
Lists accessories including seismic brackets (referenced by label 'A 329 962 - 5') and replacement blanking plates/airtight accessories for nitrogen pressurization.
Contains general data, technical characteristics, and dimension tables/figures: environmental conditions, cooling water, nitrogen, CDA and electrical characteristics, pressure/gas-throughput unit conversion tables, load distribution tables per foot for single and stacked pumps, and load distribution on seismic brackets.
- Includes Tbl. 7 electrical network protection and Tbl. 8 main circuit breaker rating / cable cross-section.
- Tbl. 9/10: load per foot for single and 2-stacked pumps; Tbl. 11: seismic bracket loads.
- Dimensions with and without seismic brackets (Fig. 10, 11).
Installation of the serial link, pin connections, protocol configuration and full command lists for RS-232 and RS-485 communication, enabling SERIAL LINK control mode.
Key Integration, Operating & Safety Requirements
| Specification / Condition | Engineering Implication | Applies To |
|---|---|---|
| Pump families covered: A100L3XXXX, A100L4XXXX (load-lock/transfer) and A100L5XXXX (A 100 L ES with Energy Saving option). p.7 ↗ | Verify the 5th digit of the part number to identify whether the Energy Saving firmware/hardware variant is fitted. | whole series |
| Energy Saving option (A 100 L ES) reduces energy consumption by 50%, automatically activated at ultimate pressure per pumping conditions. p.17 ↗ | Size facility power for reduced average draw on A 100 L ES; standard A 100 L does not have this reduction. | A100L5XXXX (A 100 L ES) |
| Frame attachment requires at least 4× M10x25 grade 8-8 screws, tightening torque 40 N·m. p.18 ↗ | Customer must supply screws and torque to 40 N·m to prevent movement/vibration issues. | whole series |
| Stacking two pumps uses 3× CHc M8x16 screws + washers at 22 N·m; lifting brackets must never lift two stacked pumps. p.18 ↗ | Only the lower pump is hoisted during stacking; brackets are rated for a single pump only. | whole series |
| Seismic installation: 4 brackets at 'SEISMIC TIE DOWN' positions (label A 329 962 - 5), supplied fixing screws torqued to 40 N·m; only lower pump anchored when stacked; floor bolts customer-supplied. p.19 ↗ | Plan anchor bolt selection per floor type and load tables (Tbl. 11) before installation. | whole series |
| Lifting requires a 3-section strap, 1 m per section, with loads of 261 N (bracket 2a), 236 N (2b), 570 N (2c). p.15 ↗ | Hoist and strap capacity must exceed the highest single-section load (570 N); use only the provided safety hooks. | whole series |
| New pumps are shipped pressurized with nitrogen; storage limited to 1 year; if blanking plates were removed, re-pressurize to 200 hPa relative with nitrogen before storage. p.15 ↗ | Prevent internal corrosion in storage; never store a used pump — return to service center. | whole series |
| Water circuit max pressure label: 101 PSI (7 bar); regulated, softened water required, with manual lockout valve 3 m from pump. p.8 ↗ | Facility water must meet spec and be valved for LO/TO; unregulated mains water causes limescale clogging. | whole series |
| Nitrogen connection: 6 mm OD / 4 mm ID pipe, 1/8 NPT ports; CDA may replace N2 when pumped gases are inert. p.21 ↗ | Stock correct tubing/fittings; CDA substitution simplifies gas infrastructure for inert processes. | whole series (N2/CDA purge, ES function) |
| Remote control dry contacts rated max 50 V / 1 A, SELV wiring required. p.23 ↗ | Do not switch higher-voltage loads directly; use interposing relays for PLC/valve circuits. | whole series |
| Rotation frequency setpoint via 0–10 VDC analog input: 0 V = maximum frequency, 10 V = minimum frequency; active only in Rem mode. p.23 ↗ | Note the inverted logic when programming the host analog output; switch Loc-Rem to 'Rem'. | whole series |
| Start/stop logic input: 24 VDC (16-pin) or 15–24 VDC (19-pin) = pump run; 0 V = stop. p.23 ↗ | Wire the equipment start interlock to deliver the correct voltage level for a running command. | whole series |
| Mains protection: IEC 60947-2 curve D breaker, ≥10 kA short-circuit cut-off, located ≤7 m from pump and in line of sight, plus a differential breaker. p.25 ↗ | Include this breaker in the facility electrical design; rating per Tbl. 7/8 of the technical data. | whole series |
| Rating plate example: 200–230 V Δ, 50/60 Hz, 3 phases/4 wires, full-load current 12 A (illustrative value). p.9 ↗ | Size supply cable and breaker per the actual rating plate of the delivered unit (see Tbl. 8). | whole series |
| Electronic circuit protected by two 0.1 A/250 VAC semi time-lag fuses, not replaceable by the installer; internal capacitors hold >60 VDC after power-off. p.25 ↗ | No field fuse replacement; observe 5-minute wait after power-off before any intervention. | whole series |
| Pump has no built-in emergency stop; host equipment EMS must switch the pump off when activated; breaker restart after fault requires ~15 s wait. p.25 ↗ | Integrate the pump into the equipment's EMS circuit and document restart delay in the equipment procedures. | whole series |
| Oil charge is filled at delivery and must not be modified or drained; oil change is done only during factory overhaul. p.28 ↗ | No user-side lubrication maintenance — any oil work goes through a Pfeiffer service center. | whole series |
| Permitted pumped gases: argon, helium, nitrogen and air only; pumping liquids, dust, condensable vapors, corrosive/explosive media is misuse. p.13 ↗ | Confirm process gas compatibility before selection; deviation voids warranty and conformity. | whole series |
| Connected chamber parts must withstand 1·10⁻³ hPa absolute negative pressure; exhaust line components must be rated above the pump's maximum generated pressure. p.22 ↗ | Verify pressure ratings of chambers and exhaust line components during system design. | whole series |
| Manual lockout valves must be installed at 3 m from the pump on water and nitrogen circuits; full LO/TO required before maintenance. p.11 ↗ | Add these valves and LO/TO steps to installation and maintenance procedures to remain compliant with the safety instructions. | whole series |