ACP 90 — instruction manual

Technical Reference Summary

Covered Models & Document Scope

Pfeiffer Vacuum ACP 90 multi-stage air-cooled Roots pump operating instructions. Covers two versions: ACP 90 SD (standard, clean gases, part number V9SAbScZde02) and ACP 90 G (traces of corrosive gases, part number V9GAbScZde02). Ordering matrix encodes: pumping application (V9SA clean / V9GA corrosive trace), inlet tube (A without / B with), power supply (A single-phase 230 V, B three-phase 230 V, C three-phase 440 V), gas ballast dilution (B closed with plug, F permanent filter, M ON/OFF manual), purge dilution (R inert gas purge, G version only), and mounting (F feet, R casters). Not all option combinations exist as products.

Section-by-Section Technical Breakdown

pp. 13-15
Product Description and Human/Machine Interface

The ACP 90 is a multi-stage Roots pump for clean, dry vacuum, capable of pumping up to atmospheric pressure and cyclic duty, driven by an IPM synchronous motor with an integrated frequency converter. The HMI includes gas ballast controls (SD), purge connection (G), control pad with display, RUN/STOP keys and indicators; the 4-digit display shows frequency, current, voltage, and 'L-AL' alerts (x10 indicator means displayed value ×10).

  • Scope of delivery: pump, cover plug for auto-start, feed through, operating instructions and safety labels
  • Labels: INLET, PUMP EXHAUST, DILUTION, LIFTING EYE, MAINTENANCE indicator, REMOTE PORT, RS PORT, CPC earth
  • Rating plate example: AC-IN 200–240 VAC, 50/60 Hz, 16.4 A
pp. 16-18
Installation and Pumping Line Connection

Pump must be horizontal, resting on feet, pumping axis vertical, inlet on top, with ≥80 mm clearance on all sides and unobstructed ventilation grids. Inlet line must be short, straight, at least the inlet flange diameter; connected chambers must withstand 1·10⁻³ hPa absolute. Exhaust must connect to an extraction system; G version purge uses dry filtered nitrogen at ≤300 hPa relative pressure at the purge input.

  • Minimum 80 mm clearance from fixed walls on all sides
  • Pumping line leak test mandatory after installation; O-rings under blanking plates may need replacement per application
  • Chambers connected to inlet must withstand 1·10⁻³ hPa absolute negative pressure
  • Install isolation valves on inlet/exhaust and flexible tubes for vibration decoupling
  • Purge: 1/4 BSPT connector, isolation valve near pump, ≤300 hPa relative at purge input
pp. 18-21
Electrical Connection and Protection

Power supply options: single-phase 230 V, three-phase 230 V, three-phase 440 V. Customer must provide a curve D circuit breaker (IEC 60947-2) with ≥10 kA short-circuit capacity within 7 m of the pump, plus a differential circuit breaker. Earth stud torque 3 N·m; earth impedance pump-body-to-ground <0.1 Ω at 25 A. The pump has no integrated emergency stop — the host equipment must switch it off.

  • Earth wire longer than conductor wires; crimp ferrules on conductors, ring terminal on earth
  • US/Canada: wiring per ANSI/NFPA 70 (NEC) and CSA C22.1 (CEC Part I)
  • Wait 5 minutes after power-off before working on the pump (capacitors charged >60 VDC)
  • CPC earth terminal marked with IEC 60417 #5019 symbol
pp. 21-25
Remote Control Connector (REMOTE PORT) Wiring

15-pin Male D-Sub HD connector for start/stop, speed setting, and status dry contacts; dry contact outputs rated 30 V / 1 A max, SELV circuits only. Logic inputs S1 (start/stop), S2 (serial link enable), S3 (LOCAL/REMOTE), S4/S5 (speed steps). Cover plug enables auto-start in LOCAL mode.

  • Speed via S4/S5: 0/0 = 6720 min⁻¹ (112 Hz); 0/1 = 4800 (80 Hz); 1/0 = 3600 (60 Hz); 1/1 = 2400 (40 Hz)
  • Frequency converter functions for analog mode: F01=1, F18=0, E02=21 or 1021, E03=100 (logic: F01=5, F18=100, E02=2, E03=100)
  • Outputs: 2-6 isolation valve control (opens on STOP, closes 2 s after RUN); 9-13 pump running; 10-14 maintenance threshold reached
pp. 23-24
Analog Input Speed Control

Rotation speed adjustable via 0–10 V analog input: E02=21 maps 0–10 V to 2400–6720 min⁻¹ (increasing); E02=1021 maps to 6720–2400 min⁻¹ (decreasing). Speeds between 4800 min⁻¹ (80 Hz) and 6540 min⁻¹ (109 Hz) are prohibited — the pump does not operate in this range.

  • Analog input A1 on pins 8-10, 0–10 V signal
  • Prohibited band 4800–6540 min⁻¹ (80–109 Hz) applies to all speed control modes
pp. 25-27
Serial Link (Modbus RS-485)

RJ45 RS PORT provides Modbus RS-485 control: start/stop, speed setting 2400–6720 min⁻¹, and parameter reading. Settings: 9600 baud, no parity, 1 stop bit, 8 data bits; default pump address 1; function codes 03 (read) and 10 (write) with CRC16.

  • Pinout: 4 = D0-, 5 = D1+, 6 = RTS, 8 = GND; others not connected
  • Key Modbus addresses: 07 06 W start/stop; 07 05 W speed (300 × Hz); 08 09 R frequency; 0F 15 R power (0.1 × W); 04 5E R running time (0.1 × h)
  • Max speed 6720 min⁻¹ = write '83 40' to address 07 05
  • Ferrite recommended (e.g., Wurth 74271112) since network output cannot be shielded
pp. 28-34
Operation: Precautions, Control Modes, Gas Ballast and Purge

Before start: verify inlet connection, unclogged exhaust line, inlet pressure ≤ atmospheric, ambient temperature in range. Thermal sensor blocks operation/start-up when pump body temperature >40 °C; pump auto-restarts when the fault clears. Noise may exceed 70 dB(A) during fore-pumping at high pressure. Control modes: LOCAL with/without auto-start, REMOTE, and SERIAL LINK.

  • Wait 5 min after power-off before any work (charged capacitors)
  • Oil charge delivered filled — do not modify level or drain (done only at service overhaul)
  • Noise >70 dB(A) in fore-pumping phase: connect exhaust to pipe/chimney, wear hearing protection
  • G version purge protects bearings and dilutes corrosive gases
pp. 35-38
Maintenance and Replacement

Maintenance sections cover safety during maintenance, operating time tracking, maintenance frequency, on-site maintenance, and the exchange/disconnection procedure including preparation for shipping. The MAINTENANCE indicator lights when the maintenance threshold is reached (also output 10-14 closed).

  • Pump oil is not user-serviceable — draining performed only during service center overhaul
  • Disconnect procedure: isolate from installation and prepare pump for shipping with blanking plates
pp. 39
Decommissioning, Extended Immobilization and Disposal

Covers extended immobilization, recommissioning after storage, and disposal of the product.

  • Run-in procedure for long-term storage: 30 min gas ballast open + 30 min at ultimate pressure, at least annually
pp. 40-42
Malfunctions and Fault History

Includes fault history function and troubleshooting: 'the pump runs incorrectly' and 'the pump does not work'. The control pad displays 'L-AL' for alerts and alarm codes in alarm mode; holding FUNC/DATA acknowledges alerts.

  • Alarm mode on control pad displays the active alarm code
  • Thermal safety fault clears automatically and pump restarts
pp. 46-53
Technical Data and Dimensions

Sections 12.1–12.6 contain general, technical, environmental, nitrogen, electrical characteristics and dimensions, plus ETL mark and EC/UK declarations of conformity. (Full data tables are on pages 46–51 of the document.)

  • Nitrogen characteristics given for the G-version purge circuit
  • Electrical characteristics per supply variant (single-phase 230 V, three-phase 230 V / 440 V)
  • ETL mark confirms UL/CSA compliance
pp. 29
Gas/Application Matrix

A matrix recommends pump configuration per gas type: SD version with gas ballast closed for neutral/inert gases; G version for traces of corrosive gases. Users must verify gas compatibility with pump materials and apply monitoring actions (e.g., gas ballast use for condensable vapors).

  • Neutral/inert gas (air, N2, CO2, noble gases): SD with gas ballast closed
  • Gases with condensable vapors: use gas ballast per dedicated section

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication Applies To
Supply variants: single-phase 230 V, three-phase 230 V, three-phase 440 V; example rating plate AC-IN 200–240 VAC, 50/60 Hz, 16.4 A p.90 ↗ Verify supply voltage against the product label before connection; breaker and cabling sized accordingly Whole series (per ordering matrix code position 'c')
Required customer circuit breaker: curve D per IEC 60947-2, short-circuit cut-off capacity ≥10 kA, within 7 m of pump, plus a differential circuit breaker p.90 ↗ Specify breaker type/rating and placement in the electrical design of the host installation Whole series
Earth stud torque 3 N·m; impedance pump body to ground must be <0.1 Ω at 25 A; earth wire longer than phase conductors p.90 ↗ Ensures protective earthing compliance; verify earth continuity at commissioning Whole series
Dry contact outputs on REMOTE PORT rated 30 V, 1 A max, SELV circuits only p.90 ↗ PLC/relay interfaces must stay within 30 V/1 A to avoid damaging the converter Whole series
Speed steps via logic inputs S4/S5: 6720 / 4800 / 3600 / 2400 min⁻¹ (112 / 80 / 60 / 40 Hz) p.90 ↗ Speed selection can be hardwired without analog signal Whole series
Speeds between 4800 min⁻¹ (80 Hz) and 6540 min⁻¹ (109 Hz) are prohibited — the pump does not operate there p.90 ↗ Avoid this band in analog setpoints, logic inputs and Modbus writes Whole series
Analog input 0–10 V maps to 2400–6720 min⁻¹ (E02=21) or 6720–2400 min⁻¹ (E02=1021) p.90 ↗ Configure converter functions F01/F18/E02/E03 per desired direction and mode Whole series
Modbus RS-485 on RJ45: 9600 baud, no parity, 1 stop bit, 8 bits, default address 1; speed write value 300 × Hz ('83 40' = 6720 min⁻¹) p.90 ↗ Define serial integration parameters and scaling factors for PLC communication Whole series
Isolation valve control output (pins 2-6): opens on STOP, closes 2 s after RUN command p.90 ↗ Valve actuation timing must tolerate the 2 s delay after start Whole series
Thermal sensor blocks operation/start-up when pump body temperature >40 °C; auto-restart after fault clears p.90 ↗ Ambient/cooling design must permit body temp below this threshold; no manual reset needed Whole series
Wait 5 minutes after power-off before any work — capacitors hold >60 VDC (up to mains voltage) p.90 ↗ Include discharge delay in maintenance procedures Whole series
Inlet-side connected chambers must withstand 1·10⁻³ hPa absolute negative pressure p.90 ↗ Check chamber mechanical rating before connection Whole series
Purge (G version): dry filtered nitrogen via 1/4 BSPT connector, max 300 hPa relative at purge input; manual lockable valve within 3 m of pump p.90 ↗ Nitrogen supply specification and safety valve placement in installation design ACP 90 G only
Minimum 80 mm clearance from walls on all sides; ventilation grids must not be obstructed p.90 ↗ Air cooling performance depends on clearance — reserve space in layout Whole series
Lifting: 2-strand strap, each strand >500 mm, load per strand >90 kg, by hoisting rings p.90 ↗ Implies pump weight >90 kg per strand; specify correct rigging Whole series
Storage: max 1 year in original packaging; if longer, run pump 30 min with gas ballast open + 30 min at ultimate pressure at least annually; service center review after 4 years before commissioning p.90 ↗ Plan periodic maintenance runs for stored spare pumps Whole series
Pump must be installed horizontal on feet, pumping axis vertical, inlet on top; no loads allowed on the inlet flange p.90 ↗ Chamber/pipework must be supported independently of the pump Whole series
Noise can exceed 70 dB(A) during fore-pumping at high chamber pressure p.90 ↗ Hearing protection and exhaust piping/chimney required in noise-sensitive areas Whole series
Oil charge is delivered filled and must not be modified or drained by the user — draining only during service-center overhaul p.90 ↗ No routine oil changes in user maintenance plan; category oil_capacity parameter is service-level only Whole series
Pump has no integrated emergency stop; host equipment EMS must switch off the pump when activated p.90 ↗ Integrate pump power into the machine's safety chain Whole series