ASM 390 — instruction manual 392
Technical Reference Summary
Covered Models & Document Scope
Operating instructions (EN, 102 pages) for Pfeiffer Vacuum ASM 390 and ASM 392 mobile helium/hydrogen/helium-3 leak detectors with dry pumps. Part numbers: CSGB01GxMM9x (ASM 390, tracer gas pumping speed >10 l/s) and ESGB02GxMM9x (ASM 392, two turbomolecular pumps, fastest tracer gas pumping speed). Options/accessories covered by reference: standard and smart sniffer probes, spray gun, RC 10 remote control, bypass, ASM 39x-HVM bottle holder (separate documents listed).
Section-by-Section Technical Breakdown
Identification, scope of delivery, variants, connection interface and control panel. ASM 390/392 are mobile detectors with dry pumps for ultra-clean, high pumping speed applications (semiconductors, coating).
- ASM 390: tracer gas pumping speed >10 l/s; ASM 392: two turbomolecular pumps for fastest tracer gas speed
- Scope of delivery includes mains cable (EU FR/DE and/or US), calibration certificate for internal calibrated leak, maintenance kit, hose holder, Smart sniffer probe adaptor, D-Sub connector kit
- Interfaces: Ethernet, USB, 9-pin D-Sub RS-232, 15-pin or 37-pin D-Sub I/O, sniffer probe connectors, purge/inlet vent connector with protection filter, SD card slot
- Control panel: magnetized, START/STAND-BY button, two levels of function keys, touch screen
Site requirements, ventilation, purge/inlet vent and neutral gas line connection, exhaust connection, electrical connection and connection of the part under test.
- Leave ≥10 cm free space all around; work surface load ≤50 kg
- Neutral gas supply: 200 hPa relative excess pressure, 5000 sccm flow at 1 bar abs inlet, dry and filtered; use pressure 0–0.3 bar relative (1–1.3 bar abs); manual valve within 3 m of product
- Exhaust pressure must not exceed 200 hPa (relative); exhaust line must be slightly negative; never obstruct exhaust
- Inlet connection: max permitted weight at inlet 18 kg, max torque 25 N·m; use flexible metal pipework, never rigid/compressed-air style plastic tube
- Connected parts must withstand negative pressure of 1·10³ hPa relative to atmospheric
- Class I equipment — must be earthed; use only original manufacturer mains cable; shielded cables in interference-prone environments
Start-up sequence (cable, breaker to I, set language/unit/date/time, wait for Stand-by), powering off, and control panel familiarization.
- Power off: breaker to O, disconnect cable, wait 5 minutes before any work
- After mains failure the detector shuts down and restarts automatically when power is restored
Use conditions, test start/stop for hard vacuum and sniffing methods, automatic test after start-up, and detailed calibration procedures with internal or external calibrated leaks. Operation monitoring via warnings, errors and critical errors (E244).
- Never use tracer gas with H2 content >5%; recommended mix is 95% N2 / 5% H2
- Two test methods: Hard Vacuum and Sniffing; reject points settable per method
- In Stand-by mode the displayed leak rate corresponds to the detector background
- Internal calibrated leak is 4He, in the 10⁻⁷ mbar·l/s range, supplied with calibration certificate; default tracer gas is 4He
- Mandatory external calibration when changing tracer gas (4He, 3He, H2) or test method; manufacturer does not supply 3He or H2 calibrated leaks
- Calibration recommended at least once a day and at start of each work session for intense operation (e.g. 8-hour shifts); detector prompts calibration 20 minutes after switching on
- Critical error E244 requires stopping the detector and contacting service center
Touch-screen menu structure covering set points (audio alarm, pollution function, reject points, probe clogged threshold), test parameters (test method, correction factor, test mode, probe type, regeneration, massive mode), spectro (tracer gas, filament, calibrated leak), maintenance timers/history, configuration (language, passwords, screen), and advanced parameters (start-up timer, background suppression, crossover pressures, serial links, I/O connector, SD card).
- Set points include hard vacuum reject point, sniffing reject point and probe clogged threshold
- Test menu: correction factor, automatic cycle end, inlet vent, memo function, bypass option, regeneration, massive mode
- Spectro menu: tracer gas selection (4He, 3He, H2), filament parameters, calibrated leak type
- Maintenance menu: timers, detector/pump information, event and calibration history, burn-in, secondary pump and analyzer cell maintenance
- Advanced menu: leak detection calibration, analyzer cell, internal Pirani gauge calibration, external gauge, serial link 1/2, I/O connector, SD card, service
- Password-protected access configuration available
Fault identification and remedy tables for operational issues during use.
- Faults classified as Warning, Error, or Critical error (E244) displayed on control panel
Maintenance and replacement procedures, with detailed instructions delegated to the separate Maintenance instructions document (part number 126348M).
Lists original manufacturer accessories (sniffer probes, remote control, filters, bypass, bottle holder, etc.) with part numbers.
- Only original manufacturer accessories permitted; accessory part numbers listed in this chapter
General technical characteristics, environmental conditions, pressure unit conversions, gas throughput conversions and dimensional drawings.
- Environmental conditions table (Tbl. 1) and pressure/throughput conversion tables (Tbl. 2, 3) included
- Dimensional drawings for installation planning
Tree diagram of the Settings menu with default setting tables for all menus, plus UL/CSA compliance, SEMI certificate, UK Declaration of Conformity and EC Declaration of Conformity.
- Default settings tables Tbl. 4–9 for Set Points, Test, Spectro, Maintenance, Configuration and Advanced menus
- Initial settings tables Tbl. 10–11 for graph screen parameters and recording parameters
Key Integration, Operating & Safety Requirements
| Specification / Condition | Engineering Implication | Applies To |
|---|---|---|
| Hydrogen tracer gas concentration must never exceed 5%; the recommended tracer gas is a 95% N2 / 5% H2 mix (hydrogenated nitrogen). p.14 ↗ | Use forming-gas-type mixtures only; higher H2 concentrations are classified as foreseeable misuse and void warranty/safety protection. | ASM 390 and ASM 392 |
| Tracer gases supported: 4He (default), 3He and H2; the internal calibrated leak is a 4He leak in the 10⁻⁷ mbar·l/s range. p.29 ↗ | Default operation in 4He requires no external calibration; using 3He or H2 mandates external calibration with a leak of the selected gas, which Pfeiffer does not supply for 3He/H2. | ASM 390 and ASM 392 |
| ASM 390 has tracer gas pumping speed >10 l/s; ASM 392 is equipped with two turbomolecular pumps for the fastest tracer gas pumping speed. p.18 ↗ | Select ASM 392 for highly demanding, high-throughput applications; ASM 390 for fast response with a single dry primary pump train. | ASM 390 vs ASM 392 variant distinction |
| Maximum permitted weight at the detector inlet is 18 kg and maximum torque is 25 N·m. p.25 ↗ | Limit connected pipework/fixture mass and tightening torque at the inlet flange to avoid damaging the vacuum circuit. | whole series |
| Parts/installations connected to the detector inlet must withstand a negative pressure of 1·10³ hPa relative to atmospheric; inlet pressure must not exceed atmospheric. p.25 ↗ | Verify mechanical integrity of the test part before hard vacuum connection to prevent implosion and detector damage. | whole series |
| Exhaust pressure must not exceed 200 hPa (relative); exhaust must never be obstructed and the customer exhaust line should be kept under slight negative pressure. p.24 ↗ | Check exhaust filter for clogging and ensure any connected exhaust line does not create back-pressure that damages the primary pump. | whole series |
| Neutral gas supply (e.g. nitrogen) specifications: relative excess pressure 200 hPa, flow rate 5000 sccm at 1 bar abs inlet; usable pressure range 0–0.3 bar relative (1–1.3 bar absolute); manual valve within 3 m of product; gas must be dry, filtered and different from the tracer gas. p.23 ↗ | A neutral gas line reduces detector background; overpressure can keep the inlet valve closed — set a pressure regulator within these limits. | whole series (purge/inlet vent connector) |
| Leave at least 10 cm free space all around the detector; work surface load limit is 50 kg. p.21 ↗ | Plan installation clearance for ventilation and never exceed 50 kg of parts/accessories placed on the work surface. | whole series |
| Calibration is recommended at least once a day, at the start of each 8-hour work session for intense operation, and is mandatory whenever test method or tracer gas is changed; the detector prompts calibration 20 minutes after switch-on. p.29 ↗ | Embed daily/shift-start calibration into the test procedure to keep measurement accuracy; changing 4He↔3He↔H2 or hard vacuum↔sniffing always requires recalibration. | whole series |
| After switching the main breaker to O, wait 5 minutes before working on, removing the cover of, or moving the detector; some components between mains input and breaker remain live. p.12 ↗ | Lockout maintenance procedures must include the 5-minute wait and mains cable disconnection to avoid electric shock. | whole series |
| Product must not be placed on an incline greater than 3° (6%) and must be immobilized with rear-wheel brakes during use. p.16 ↗ | Floor flatness and brake engagement are installation prerequisites; tilting risk otherwise exists. | whole series |
| Transport of components weighing more than 20 kg requires suitable lifting gear; internal surfaces can exceed 70 °C during operation. p.12 ↗ | Plan handling equipment for >20 kg loads and allow cool-down plus EN ISO 21420 gloves before opening covers. | whole series |
| Use only flexible (metal) pipework to connect the test part; rigid or flexible plastic (compressed-air style) tube is prohibited; pipes should be short, sealed and of the same diameter as the inlet. p.25 ↗ | Plastic tubing causes permeation/desorption errors and outgassing; specify ISO-KF metal flexible lines (DN 40 ISO-KF hose holder provided). | whole series |
| The purge connector ships with a protection filter; the purge maintains an internal airflow, and activating the 'Pollution' function is recommended in addition to the purge. p.23 ↗ | Keep the filter fitted or switch to a neutral gas line; enable the Pollution function to protect the analyzer from tracer gas contamination. | whole series |
| External calibration should use a calibrated leak in the same leak-rate range as the leaks to be measured (i.e., near the application reject point), especially when the reject point is far from the internal 10⁻⁷ mbar·l/s leak. p.29 ↗ | Stock an external calibrated leak matched to your reject threshold for accurate measurements far from 10⁻⁷ mbar·l/s. | whole series |
| Critical error E244 requires stopping the detector and contacting the service center; warnings and errors are displayed via pictograms on the control panel. p.27 ↗ | Differentiate operator-resettable warnings/errors from E244, which always requires service intervention. | whole series |
| Detector is Class I equipment and must be earthed; use only the supplied or original-manufacturer mains cable. p.25 ↗ | Grounding integrity is a compliance and safety requirement; replacement cables must be original Pfeiffer parts (ref. in Maintenance instructions 126348M). | whole series |
| For extended storage, run a hard vacuum test in the most sensitive mode with the blank-off flange fitted and inlet vent disabled, then shut down — the detector remains under vacuum, reducing degassing time at next start-up. p.17 ↗ | Apply this procedure for long idle periods to shorten restart-to-measurement time. | whole series |
| Standby displayed leak rate corresponds to detector background; the measurement should be allowed to stabilize after reaching the most sensitive test mode before reading the leak rate. p.28 ↗ | Use the Standby reading as background check and enforce stabilization time before pass/fail judgment in test procedures. | whole series |
| Ordering part numbers: ASM 390 = CSGB01GxMM9x, ASM 392 = ESGB02GxMM9x. p.7 ↗ | Use these codes (with the x wildcards for variant options) when ordering and for service identification via the rating plate. | ASM 390 / ASM 392 |