ATH 1600 MT — instruction manual ath 1603 2303 m

Technical Reference Summary

Covered Models & Document Scope

Pfeiffer Vacuum operating instructions covering two magnetically levitated (active magnetic bearing) turbomolecular pump models for clean applications: ATH 1603 M (part numbers Yxxx2100 / YNxx215A / YLxx215A / YJxx215A) and ATH 2303 M (Xxxx2100 / XNxx215A / XLxx215A / XJxx215A / TMBxxxxxxxxx). 'M' denotes version without Temperature Management System (no TMS). Variants may include an OBC electronic drive unit mounted on the pump, optional copper gasket, purge solenoid valve, air inlet solenoid valve, screw/bolt set; HHR hand-held remote is NOT included and must be ordered separately. Applicable documents: Magpower operating instructions (P/N 120932), OBC V4 operating instructions (P/N 124470).

Section-by-Section Technical Breakdown

pp. 17-18
Product description and man/machine interface

Identifies the pump, scope of delivery and applications. These AMB turbopumps handle inlet pressures on the order of 2 hPa or inlet flow around 4000 sccm, suited to coating/dry-etch plasma and R&D. Rotation speed is selectable between minimum and nominal speed; interface components (splinter shield, inlet flange, purge, exhaust, water, electrical connector, grounding) are shown per model.

  • Scope of delivery: pump, splinter shield in inlet flange, 2 lifting rings, label sheet, manual; optionally copper gasket, purge/air inlet solenoid valves, screws
  • Selected (stand-by) speed can be set between minimum and factory-set nominal speed
  • M version = no temperature management system (no TMS)
  • Controlled via OBC (pump-mounted) or Magpower drive unit
pp. 19-29
Installation: mounting, loads, positioning, pumping line, water, nitrogen, purge and leak check

Detailed mechanical and utility installation. Frames must absorb rotor-blockage kinetic energy (ISO 27892 verified); mandatory screw types, grades and torques are tabulated. Covers positioning limits (magnetic field, radiation), pumping-line connection rules, water and nitrogen circuit hookup with connector tightening procedures and purge setup.

  • Rotor-blockage rotor energy: 76 kN·m (ATH 1603 M) / 124 kN·m (ATH 2303 M); max torque 30 / 52 kN·m; max bending moment 21 / 35 kN·m; max axial force 0–398 / 0–650 kN
  • DN200 ISO-F / DN250 ISO-F: M10 screws, 12 pcs, grade 12-9, 30 N·m each; DN200 CFF: M8, 24 pcs, 20 N·m; DN250 CFF: M8, 32 pcs, 20 N·m
  • Pump enclosure withstands max 11·10³ hPa static overpressure for 1 min; connected chamber must withstand −1·10⁻³ hPa relative
  • Uniform magnetic field limit 0.5 mT guaranteed, 0.5–5 mT load-dependent, >5 mT needs shielding; radiation tolerance up to 10³ Gy
  • Pump can operate in any position; transient start at atmospheric pressure possible but a backing pump is required
  • Water connectors: 1/4 NPT with sealing tape/glue; single-ferrule nut tightened 1/2 turn; twin-ferrule 3/4 turn
  • Purge flow reducer guarantees 50 sccm at 1.1·10³ hPa; purge pressure max 1–1.5·10³ hPa absolute; maintain purge while rotor runs
pp. 32-35
Operation: start-up, shutdown and monitoring

Covers preliminary precautions, powering on, pumping start-up, restart after emergency stop, pumping shutdown, powering off, prolonged stoppage and operation monitoring (per table of contents; details beyond page 28 not included in extract).

  • Restart procedure defined after emergency equipment stop
  • Prolonged stoppage procedure defined (see also decommissioning)
pp. 36-41
Maintenance and exchange procedure

Maintenance safety instructions, maintenance frequency and on-site maintenance, plus exchange by replacement product: disconnection, water circuit draining and shipping preparation (per table of contents; details beyond page 28 not included in extract).

  • Includes full disconnection and water-circuit draining procedure before removal
  • Shipping preparation steps for the exchanged pump
pp. 42
Decommissioning, recommissioning and disposal

Shutting down for longer periods, recommissioning after storage and disposal of the product (per table of contents; details beyond page 28 not included in extract).

  • Used-pump storage references the prolonged-stoppage procedure
pp. 43-47
Service solutions, accessories and spare parts

Lists Pfeiffer Vacuum service offerings, accessories (including HHR remote control, ordered separately) and spare parts sections (per table of contents; details beyond page 28 not included in extract).

  • HHR hand-held remote control is not part of the standard order
pp. 48-55
Technical data and dimensions

Conversion tables for pressure and throughput units, general/technical/environmental characteristics, cooling water, nitrogen, electrical characteristics (pump and customer network), plus dimension drawings for both models with and without OBC and solenoid valve dimensions (per table of contents; detail tables beyond page 28 not included in extract).

  • Nominal speeds: ATH 1603 M 39000 min⁻¹, ATH 2303 M 31000 min⁻¹ (from load table)
  • Electrical supply per rating plate example: 44 Vrms, 1300 Hz, 7 A
  • Weight example ATH 1603 M: 39 kg

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication Applies To
Inlet flange fastening is mandatory: DN200 ISO-F and DN250 ISO-F use 12× M10 screws grade 12-9 at 30 N·m each; DN200 CFF uses 24× M8 grade 12-9 at 20 N·m; DN250 CFF uses 32× M8 grade 12-9 at 20 N·m (screw lengths 35/35/65/65 mm; total clamping forces 161500/161500/266000/355000 N). p.21 ↗ Frame/flange design and bolt torque must follow these exact values; under- or over-torque risks screw loosening or damage and voids warranty/liability. ATH 1603 M and ATH 2303 M
Worst-case rotor blockage (rotor split in two at nominal speed) produces max torque 30 kN·m (ATH 1603 M) / 52 kN·m (ATH 2303 M), max bending moment 21 / 35 kN·m, and max axial force up to 398 / 650 kN; rotor kinetic energy is 76 / 124 kN·m. p.21 ↗ Equipment frame and anchoring must be dimensioned to absorb these transient loads per ISO 27892-validated testing. ATH 1603 M: 39000 min⁻¹; ATH 2303 M: 31000 min⁻¹ nominal
Maximum axial force only occurs when the pump inlet is blocked by rigid parts; no load on the equipment occurs with limited stiffness, e.g. a pendulum valve or >100 mm free space above the inlet. p.20 ↗ Avoid placing stiff parts close to the inlet flange; use pendulum valves or ≥100 mm clearance to eliminate axial blockage loads. Both models
Pump enclosure withstands maximum static overpressure of 11·10³ hPa for 1 minute when connected per manufacturer recommendations. p.19 ↗ Defines burst-safety margin for incident overpressure conditions of the installation. Both models
Chambers/parts connected to the inlet must withstand a negative pressure of 1·10⁻³ hPa relative to atmospheric pressure. p.24 ↗ Verify upstream chamber mechanical strength before connection. Both models
Pump operation is guaranteed in a uniform magnetic field up to 0.5 mT; 0.5–5 mT depends on cooling and gas loads; >5 mT can overheat the rotor and requires shielding. Radiation tolerance of the standalone pump is up to 10³ Gy. p.22 ↗ Site survey for magnetic fields and radiation needed before installation; add shielding above 5 mT. Both models
Purge flow reducer delivers 50 sccm at 1.1·10³ hPa; purge pressure must not exceed 1–1.5·10³ hPa absolute; purge must remain on as long as the rotor runs. p.27 ↗ Size N₂ supply regulator accordingly; stopping purge while rotating risks fore-vacuum gas migrating to the high-vacuum side and damaging bearing housings. Both models (with purge option)
LEL detection device in the extraction system must stop chemical supply when flammable gas exceeds 25% of LEL; oxidant concentration must not exceed 60% of the maximum permissible concentration (MOC) per NFPA 69-2019. p.12 ↗ Integrator must implement these interlocks and dilution limits for flammable/pyrophoric processes per EN 50495. Installations pumping flammable/pyrophoric gases
Manual lock-out valves for water and nitrogen supply must be installed within 3 m of the product. p.11 ↗ Plant layout must include accessible LO/TO shutoff points for both utilities. Both models
Threaded water connections use 1/4 NPT male connectors with oleo-waterproof glue or PTFE tape; single-ferrule nuts tightened 1/2 turn, twin-ferrule nuts 3/4 turn with open-ended spanner while holding the connector body. p.25 ↗ Standardized leak-free utility hookup procedure; deviation risks leaks in pressurized circuits. Both models, water and purge circuits
New pumps may be stored max 1 year, in protective envelope with blanking plates on inlet/exhaust/purge, in a clean dry non-contaminated area within permitted storage temperatures. p.16 ↗ Warehouse stock rotation must respect the 1-year limit; check environmental characteristics chapter for temperature range. Both models, new pumps
Nominal speeds are 39000 min⁻¹ (ATH 1603 M) and 31000 min⁻¹ (ATH 2303 M); selected stand-by speed can be set anywhere between minimum and nominal speed. p.21 ↗ Speed can be tuned for high-pressure/high-throughput process optimization via the drive unit. ATH 1603 M / ATH 2303 M respectively
Pumps handle inlet pressure on the order of 2 hPa or inlet flow around 4000 sccm, aimed at coating/dry-etch plasma and R&D clean applications. p.17 ↗ Confirms suitability for high-throughput process operation rather than generic clean-vacuum duty. Both models
Electrical rating per product rating plate example: 44 Vrms, 1300 Hz, 7 A, 39000 rpm, 39 kg (ATH 1603 M). p.8 ↗ Drive/power supply sizing and cable spec must match these values; verify against the actual rating plate of each unit. Example shown for ATH 1603 M (Y266215A)
Lifting requires devices rated for pump weight, slings with minimum arm length 500 mm, 2 EN-1492-1 approved multi-strand straps for horizontal handling, or the 2 rear M10 holes for inlet-up handling; always transported upright. p.15 ↗ Installers must prepare approved rigging; pump stays hanged/attached until the high vacuum flange is secured. Both models
The pump cannot evacuate at atmospheric pressure and must be combined with a backing pump; transient start at atmospheric pressure is allowed. An isolation valve (NC) must be installed between turbo and backing pump. p.22 ↗ Vacuum system design must include backing pump and NC isolation valve in the fore-vacuum line. Both models
Connected equipment must include an exhaust connection to a process-compatible backing pump venting to a dangerous-gas extraction system, with regular leak checks on exhaust pipework. p.11 ↗ Never vent the turbo exhaust to ambient in process applications; foreline leak checks become a recurring maintenance task. Both models
Blanking-plate O-rings are not compatible with all applications; users/integrators are responsible for fitting application-compatible O-rings, and avoiding reducing flanges at the inlet. p.24 ↗ Seal material selection is a user duty for chemical compatibility; do not use reducer flanges on the inlet. Both models
A filtered, dry nitrogen supply is required for optimum purge performance; when a mass flowmeter is used, the 50 sccm flow reducer must not be installed. p.27 ↗ Purge line architecture depends on flow measurement method; remove the restriction orifice when MFC/MFM controls flow. Both models with purge option