ATH 500 M — instruction manual _mt
Technical Reference Summary
Covered Models & Document Scope
Pfeiffer Vacuum ATH 500 M and ATH 500 MT magnetically levitated turbomolecular pumps with integrated electronic drive unit. Variants per part number matrix (Vxxxxx00 = ATH 500 M light-duty; Vxxxxx05 = ATH 500 M nickel plated heavy-duty; Vxxxxx03 = ATH 500 MT heavy-duty; Vxxxxx01 = ATH 500 MT nickel plated). MT = temperature management system (TMS) version with water solenoid valve and heater band; M available air- or water-cooled. Water valve option in copper or inox per ordering digits. HHR remote and 48 VDC power supply are optional accessories. Optional EtherCAT or Profibus fieldbus interfaces depending on ordering code.
Section-by-Section Technical Breakdown
Magnetically levitated turbopump with active magnetic bearings (AMB) and integrated electronic drive unit, for coating, dry-etching plasma, glass coating and R&D. M version has no TMS; MT version has temperature management system with water solenoid valve and heater band for condensable gases. Rotation speed selectable between minimum and nominal (stand-by speed configurable).
- Scope of delivery: pump with integrated drive unit, splinter shield in inlet flange, water solenoid valve (MT), manual, safety labels
- HHR remote and 48 VDC power supply not included unless ordered
- Purge connection: 1/8 BSPP female with filter (M version), 1/4 VCR (MT version); 50 sccm neutral gas purge
- Water connections: 1/8 NPT female (M), 1/4 double-ring (MT)
- Drive unit has RS-232/RS-485 serial link, REMOTE connector, HHR service connector, optional EtherCAT/Profibus
- Electronic drive unit does not power the pump; external 48 VDC supply required
Covers frame rigidity and securing per ISO 27892 rotor-burst tests, pump positioning (any position allowed), pumping line connection, water cooling circuit, air cooling accessory, nitrogen purge, leak check and electrical connection. Fastening screws, grade and torques are mandatory per flange type. Frame must absorb kinetic energy of sudden rotor blockage (24 kJ at 50000 rpm).
- Rotor blockage loads: max torque 9 kN·m (1 ms), bending moment 6 kN·m (0.5 ms), axial force 0–110 kN (0.3 ms)
- Enclosure withstands 11·10³ hPa static overpressure for 1 minute
- DN160 ISO-F/K: 8× M10 screws grade 12-9, 30 N·m each, length ≥35 mm; DN100 ISO-F/K: 8× M8, 20 N·m; DN160 CFF: 20× M8, 20 N·m; DN100 CFF: 16× M8, 20 N·m
- ISO-K claw clamps: use same number of stainless claw clamps as recommended screws
- Uniform magnetic field up to 0.5 mT guaranteed; 0.5–5 mT depends on cooling/gas load; >5 mT requires shielding; standalone pump withstands 10³ Gy radiation
- Inlet-side parts must withstand negative pressure of 1·10⁻³ hPa absolute
- Manual valve within 3 m on water and nitrogen circuits; LEL detection at 25% LEL for flammable gas pumping
Describes powering on, pumping start-up, restart after emergency stop, shutdown, prolonged stoppage and operation monitoring with status LEDs on the interfaces. Pump starts at atmospheric pressure transiently but requires a backing pump for evacuation.
- Start/stop/standby via REMOTE connector, RS-232/RS-485, HHR or fieldbus
- LED tables given for EtherCAT interface and Remote/RS232-RS485/Profibus interface status
- Lock the power connector before use; never disconnect while pump is running
Pump temperature management system (MT version) and pump braking parameters controlled via the integrated electronic drive unit.
- Temperature management (TMS) controls pump and gas temperature to minimize condensation (MT version)
- Pump braking function available via electronic drive unit
Control via HHR hand-held remote (DISPLAY/SETUP/SER NUM menus), REMOTE connector with logic inputs/outputs for dry contacts or direct voltage, RS-232/RS-485 serial link with command list, and fieldbus (EtherCAT/Profibus per dedicated manuals). RS-485 enables multi-pump control from one computer.
- REMOTE connector supports START/STOP/STANDBY and remote status via dry contacts or direct voltage
- RS-232/RS-485 command list provided in section 8.4.3
- Fieldbus operation requires separate manuals: Profibus 004432, EtherCAT 131580
Maintenance safety instructions, frequency table, on-site maintenance and replacement procedure including disconnection from installation, water circuit draining and shipping preparation. Magnetically levitated pumps are characterized as maintenance-free in normal operation.
- Replacement exchange procedure: disconnect pump, drain water circuit, prepare for shipping
- Nitrogen purge of inlet, pump and exhaust lines required before maintenance on flammable-gas installations
Shutdown for longer periods, recommissioning and disposal instructions.
- Storage of a used pump per long-stoppage shutdown procedure
- Disposal section for end-of-life handling
Fault indication and malfunction table for diagnosing pump faults via the control interfaces.
- Malfunction and fault indication via control interface status outputs
Lists ordering information for accessories (HHR, 48 VDC power supply, air cooling accessory, purge set, screw sets) and spare parts table for ATH 500.
- 48 VDC power supply and HHR must be ordered separately if not in scope of delivery
- Recommended original screw sets for flange fastening available as accessories
Technical characteristics, environmental conditions, cooling water and nitrogen specifications, electrical characteristics of pump and customer network, 48 VDC external power supply characteristics, plus dimension drawings for ATH 500 M, air-cooled M and MT versions. Nominal rotation speed 50000 rpm; rated plate example: 23 kg, 50000 rpm, 29 Vrms, 1666 Hz, 9.5 A.
- Nominal speed 50000 min⁻¹; rotor kinetic energy 24 kJ
- 48 VDC, 12 A power supply connector rating
- Pressure and gas throughput unit conversion tables included
- Dimension figures for M, air-cooled M and MT variants (pages 76–77)
Key Integration, Operating & Safety Requirements
| Specification / Condition | Engineering Implication | Applies To |
|---|---|---|
| Nominal rotation speed is 50000 min⁻¹ with rotor kinetic energy of 24 kJ; selectable stand-by speed can be set between minimum and nominal speed p.23 ↗ | Mounting frame and fasteners must be dimensioned for worst-case 24 kJ rotor-blockage event; speed reduction can be used for high-pressure pumping optimization | whole series |
| Rotor blockage transient loads: torque max 9 kN·m (duration 1 ms, delay 1 ms), bending moment max 6 kN·m (0.5 ms), axial force 0–110 kN (0.3 ms); no axial load with limited stiffness e.g. a pendulum valve or >100 mm free space p.23 ↗ | Frame rigidity and flange bolting must absorb these loads; keep a pendulum valve or ≥100 mm clearance above inlet to eliminate axial impact force | whole series |
| Mandatory inlet flange fastening: DN160 ISO-F/K = 8× M10 grade 12-9 at 30 N·m, screw length ≥35 mm; DN100 ISO-F/K = 8× M8 at 20 N·m; DN160 CFF = 20× M8 at 20 N·m (≥40 mm); DN100 CFF = 16× M8 at 20 N·m (≥35 mm) p.24 ↗ | Non-compliant torque or screw grade voids warranty and risks pump ejection; use torque wrench and original Pfeiffer screw sets | whole series |
| Pump enclosure withstands maximum static overpressure of 11·10³ hPa for 1 minute when connected per manufacturer recommendations p.22 ↗ | Defines venting/overpressure incident limit for the pump housing in the integrated equipment design | whole series |
| Magnetic field limits: guaranteed operation in uniform field up to 0.5 mT; 0.5–5 mT depends on cooling and gas loads; >5 mT causes excessive rotor heating and requires shielding. Radiation withstand up to 10³ Gy (standalone) p.25 ↗ | Measure ambient fields near magnets/ion sources; add magnetic shielding otherwise; unsuitable for high-radiation areas above 1000 Gy | whole series |
| Mandatory manual lockout valves on both water and nitrogen supply circuits within 3 m of the pump p.12 ↗ | Include these valves in installation design for LO/TO maintenance safety | whole series |
| Power supply is external 48 VDC, 12 A (connector label '48 V 12 A'); drive unit rating plate: 29 Vrms, 1666 Hz, 9.5 A, 23 kg p.9 ↗ | Size the external 48 VDC supply and cabling for 12 A; the drive unit does not power the pump itself | whole series |
| For flammable gases: LEL detection must stop chemical supply at 25% LEL; oxidant concentration must stay below 60% of the maximum permissible concentration (MOC) per NFPA 69-2019; safety systems must comply with EN 50495 p.13 ↗ | Integrator must design gas detection and inerting per these thresholds before operating on pyrophoric/flammable processes | whole series |
| Maximum water circuit pressure marked on pump: 7 bar / 101 psi (WATER MAX Pr label) p.9 ↗ | Regulate cooling water supply below this limit; use softened, filtered water to avoid clogging | whole series |
| Neutral gas purge rate is 50 sccm; purge connection is 1/8 BSPP female with filter (M version) or 1/4 VCR (MT version) p.19 ↗ | Set purge flow to 50 sccm for dilution/corrosion protection; choose correct fitting per variant | M: 1/8 BSPP; MT: 1/4 VCR |
| MT version includes TMS (temperature management system) with water solenoid valve and heater band, controlled by the integrated drive unit; M version has no TMS and is air- or water-cooled p.18 ↗ | Select MT for high-temperature/condensable processes; M must not be used with condensables (listed as misuse) | MT with TMS; M without |
| Manual lifting allowed only below 20 kg; lifting requires 2 M8 lifting rings in inlet housing holes, strap minimum length 230 mm, transport upright p.16 ↗ | At ~23 kg, use hoisting equipment with these rigging details; never tilt or push sideways | whole series |
| Parts connected to pump inlet must withstand negative pressure of 1·10⁻³ hPa absolute; avoid reducing flanges at the inlet p.27 ↗ | Specify upstream chamber/valve vacuum integrity accordingly; consult manufacturer if a reducer is unavoidable | whole series |
| For flanged exhaust-to-backing-pump connection, an NC isolation valve must be installed between turbo and backing pump; turbo cannot evacuate at atmospheric pressure and requires a backing pump (transient atmospheric start allowed) p.27 ↗ | Design fore-line with NC valve for safe start/stop sequencing and process isolation | whole series |
| ISO-K flanges cannot prevent pump rotation on rotor rupture; if claw clamps are used, install the same number of stainless-steel claw clamps as the recommended screw count for the ISO-F flange p.24 ↗ | Use rotating flanges only, or full claw-clamp count, to prevent line damage on rotor failure | ISO-K flange installations |
| Interfaces: RS-232/RS-485 serial link and REMOTE connector as standard; RS-485 supports multi-pump control; EtherCAT (manual 131580) or Profibus (manual 004432) per ordering option; HHR accessory with DISPLAY/SETUP/SER NUM menus p.20 ↗ | Choose control architecture accordingly; fieldbus configuration requires the dedicated interface manuals | whole series; fieldbus per ordering code |
| Never disconnect the pump power connector while the pump is running (LOCK label); connector must be locked before use p.9 ↗ | Interlock procedures and operator training must enforce this to avoid drive damage | whole series |
| Max storage period for a new pump is 1 year, wrapped in protective envelope with blanking plates and connector plugs in place, within permitted storage temperatures p.17 ↗ | Plan stock rotation/recommissioning inspection for pumps stored longer than one year | whole series |