COMBI WS 0245 WV2D — instruction manual combi wvd
Technical Reference Summary
Covered Models & Document Scope
Instruction manual for Pfeiffer Vacuum+Fab Solutions COMBI WVD / COMBI WS vacuum booster pumping units (Roots pump PANDA WV + two-stage DuoVane rotary vane backing pump on a frame). Variants: COMBI WVD 0220, 0235, 0245, 0400, 0440, 0900, 0470, 0950, 1800, and COMBI WS 0245 WV2D, WS 0440 WV2D, WS 0900 WV2D (twin backing pump versions). Options: control cabinet, measuring/control unit (OmniControl), dust separator (SAS/STP), condensate separator (KAS), vacuum safety valve (HV), oil mist filter OME/ODK; Roots pump standard or magnetic coupling; stationary or mobile (castors) frames.
Section-by-Section Technical Breakdown
Defines risk levels, general precautions, intended use (evacuation of dry, inert gases to medium vacuum) and foreseeable misuse (corrosive/explosive media, open flanges, liquids, pressure generation). Sets operating limits and defines qualification levels for mechanical, electrical and maintenance personnel.
- Installation limits: weatherproof indoor location, max. 2000 m altitude, horizontal with max. ±3° inclination, ambient +5°C to +45°C, max. 85% relative humidity
- Surface temperature may exceed 70°C; use finger guards as needed
- Electrical connection only by expert electrician per VDE 0105 / VDE 0100
- Misuse includes pumping corrosive/reactive gases, explosive media or oxygen above atmospheric concentration, liquids, and operation in potentially explosive areas
- Maintenance levels 1-3 defined; levels 2 and 3 require Pfeiffer technical education or service training
COMBI WVD units combine a Roots pump (PANDA WV) with a two-stage DuoVane rotary vane backing pump, OME oil mist filter, and frame (stationary with anchor bores or mobile with castors). A performance table gives pumping speeds, final pressures, flange sizes, weights and power consumption for all 12 variants.
- Supply scope: unit ready for connection on frame with connecting lines, counter flanges, operating fluid P3/D2/VSC100; optional control cabinet and OmniControl
- WVD 0220: 200 m³/h at 1 hPa (50 Hz), final pressure 1.5E-03 hPa, inlet DN100 ISO-F, outlet DN40 ISO-KF, 218 kg, 1.15 kW
- WVD 0440: 458 m³/h (50 Hz), final pressure 1.3E-03 hPa, inlet DN100 ISO-F, outlet DN63 ISO-F, 550 kg, 2.45 kW
- WVD 0900: 825 m³/h (50 Hz), final pressure 9.2E-04 hPa, inlet DN160 ISO-F, 650 kg, 4.87 kW
- WVD 1800 (largest): 1646 m³/h (50 Hz) / 2006 m³/h (60 Hz), final pressure 7.2E-04 hPa, inlet DN160 ISO-F, outlet DN100 ISO-F, 1025 kg, 5.65 kW
- WS WV2D versions use 2x DuoVane 70 backing pumps, e.g. WS 0900 WV2D: 849 m³/h, final pressure 8.2E-04 hPa, 570 kg
Covers installation location requirements, preparation (fluid filling, media supply), vacuum-side and exhaust-side piping, and mains electrical connection. Emphasizes short suction lines, strain-free piping, start-up strainer use, and motor protection switch settings.
- Vacuum line: keep as short as possible, at least the pump flange nominal diameter; use larger diameter for line lengths > 5 m
- Use a start-up strainer (cone strainer) at the intake during initial commissioning due to fouling risk; note pumping speed loss
- Exhaust line cross-section at least the pump exhaust nominal diameter; slope piping downward or fit condensate separator; no shut-off valves on exhaust side
- Excess pressure hazard threshold: > 1500 hPa abs. in lines; when extracting exhaust gases, exhaust pressure must be ≥ 250 hPa above intake-side pressure
- Three-phase motors shipped without connection accessories; motor protection switch setting must match motor rating plate; verify direction of rotation per component manual
- Control-cabinet versions are fully internally wired with motors rotating in correct direction; phase sequence relay provides monitoring; measuring tube power supply has internal 100 mA microfuse
Pre-start checks (fluid levels, oiler, line connections, rotation direction), switch-on/off procedures, and control-cabinet operation with panel elements. Specifies warm-up and post-process purge practices to protect the operating fluid.
- Run unit ~30 minutes after process completion with intake closed and gas ballast valve open to prevent oil contamination by condensation
- Control cabinet: master switch 9.10 to '1', ON key 9.6 / OFF key 9.7; switch-on/off sequence dictated by controller; panel includes lamp test 9.9 and internal operating hours counter
- Lockout: main switch off, secured with lock, key carried by responsible supervisor before maintenance
- Critical: if Roots pump is switched off and stationary for > 5 minutes, wait at least 120 minutes after ventilation before restart to equalize rotor/housing temperatures
- Exhaust-side shut-off units must open before or simultaneously with pump start
Fault table listing problems (unit does not start, switches off after a while, does not attain ultimate pressure, pumping speed too low, fluid loss, unusual noises) with causes and remedies referencing component operating instructions.
- Failure to start: check mains voltage/fuse, motor switch, thermal protection, dirty/damaged suction or pumping systems, defective motor
- Low ultimate pressure: dirty pump/accessories, dirty operating fluid, system leak, dirty/damaged Roots pump overflow valve, low backing pump fluid level
- Low pumping speed: undersized intake line, excessive exhaust pressure
- Unusual noise: dirty suction chamber or silencer, damaged bearings/gear wheels, dirty overflow valve, defective motor bearing; magnetic-coupling Roots pump case: motor runs but pump does not
Maintenance intervals follow the individual component manuals; shorter intervals for extreme loads must be clarified with Pfeiffer service. Components may be removed for customer maintenance; safety precautions for contamination and electrical isolation are given.
- Tighten connecting screws between backing and Roots pump regularly to 40-50 Nm (risk of loosening by vibration)
- Unit is voltage-free only when mains plug disconnected; secure against restart before work
- Decontaminate parts contaminated by pumped media before maintenance
- Use lifting device for the heavy backing pump; gloves per EN 420, safety shoes per EN 347
- Operating individual disassembled components is improper use and invalidates the declaration of conformity
For shutdowns longer than one year, store indoors at -10°C to +40°C; in moist/aggressive atmospheres shrink-wrap air-tight with desiccant. Re-start after checking for contamination and moisture; dispose of materials per local regulations.
- Long-term storage > 1 year: airproof shrink-wrapping plus desiccant bag required in moist or aggressive atmospheres
- Before re-start: check for contamination/moisture; have Pfeiffer service clean if necessary
- Change lubricants and operating fluids before restart after storage (corrosion/aging risk)
- Decontaminate contaminated parts before disposal
References Pfeiffer service offerings, spare parts (section 11) and accessories (section 12), plus EU and UK Declarations of Conformity (sections 13-14). Detailed part lists are in the referenced sections.
- Optional accessories: SAS/STP dust separator, KAS condensate separator, HV vacuum safety valve, OME/ODK oil mist filter
- Documentation for accessories is order-specific
- EU and UK Declarations of Conformity included in the manual
Key Integration, Operating & Safety Requirements
| Specification / Condition | Engineering Implication | Applies To |
|---|---|---|
| Installation limits: max altitude 2000 m above sea level, ambient +5°C to +45°C, max relative humidity 85%, horizontal orientation within ±3°, weatherproof indoor location. p.7 ↗ | Exclude outdoor or high-altitude installations without mitigation; foundation must be level within 3°. | whole series |
| Pumping speeds range from 200 m³/h (WVD 0220) to 1646 m³/h at 1 hPa/50 Hz (WVD 1800), with final pressures (gas ballast) from 1.7E-03 down to 7.2E-04 hPa. p.14 ↗ | Selection basis for matching unit size to required pump-down time and ultimate pressure in medium vacuum. | whole series |
| Inlet flanges: DN100 ISO-F for WVD 0220-0470 and WS 0245/0440, DN160 ISO-F for WVD 0900/0950/1800 and WS 0900; outlet flanges DN40 ISO-KF to DN100 ISO-F depending on variant. p.14 ↗ | Vacuum plumbing and flange hardware must match the specific variant's connection sizes. | per variant, see table |
| Weights range from 218 kg (WVD 0220) to 1025 kg (WVD 1800); max power consumption 1.15-5.65 kW. p.14 ↗ | Lifting equipment, load-bearing capacity of floor and electrical supply sizing must be chosen per variant. | per variant |
| Vacuum-side connection line must be at least the pump flange nominal diameter; use a larger diameter for line lengths > 5 m. p.16 ↗ | Undersized or long suction lines cause pumping-speed loss; upsize piping for long runs. | whole series |
| A start-up strainer (cone strainer) must be used at the intake during initial commissioning because system fouling must be expected even in clean processes. p.16 ↗ | Install the strainer for first startup; remove it only when solid-particle ingestion risk is excluded; expect some pumping speed loss. | whole series |
| Exhaust line must have no shut-off valves; excess pressure above 1500 hPa abs. in lines requires accident-prevention measures; extracted exhaust pressure must be at least 250 hPa above intake-side pressure. p.17 ↗ | Design exhaust system unobstructed; downstream extraction must maintain the ≥250 hPa differential. | whole series |
| Motors are three-phase, shipped without electrical connection accessories; a suitable motor protection switch is mandatory with setting matching the motor rating plate; rotation direction must be verified. p.17 ↗ | Specify motor protection switch per motor nameplate and confirm rotation before first run. | whole series |
| In control-cabinet versions, the phase sequence relay provides monitoring; the measuring-tube power supply contains an internal 100 mA microfuse. p.18 ↗ | Include phase-sequence faults in diagnostics; stock/check the 100 mA microfuse when a gauge is connected. | control-cabinet variants |
| If the Roots pump is switched off and stationary for more than 5 minutes, restart only after ventilation and a waiting period of at least 120 minutes to equalize rotor and housing temperatures. p.21 ↗ | Prevents rotor seizure from differential cooling; embed in control interlocks or SOPs for rapid cycling. | whole series (Roots pump) |
| After each process, run the unit approximately 30 minutes with intake line closed and gas ballast valve open before shutdown. p.19 ↗ | Prevents condensation/corrosion contamination of the operating fluid; add to automatic purge routines. | whole series |
| Connecting screws between backing pump and Roots pump must be checked regularly and tightened to 40-50 Nm. p.23 ↗ | Vibration can loosen them causing the Roots pump to fall; add torque check to maintenance schedule. | whole series |
| Storage and long-term shutdown: indoors at -10°C to +40°C; after storage > 1 year, change operating fluids and lubricants before commissioning. p.12 ↗ | Plan fluid change after any extended storage; in humid/ aggressive atmospheres use shrink-wrap with desiccant. | whole series |
| Intended use is restricted to evacuation of dry and inert gases to medium vacuum; corrosive/reactive gases, explosive media, oxygen above atmospheric concentration, liquids and sublimating substances are prohibited misuse. p.8 ↗ | Process gases outside this envelope require prior consultation with Pfeiffer Vacuum+Fab Solutions. | whole series |
| Surface temperature of the pumps can exceed 70°C depending on operating/ambient conditions. p.7 ↗ | Provide finger guards and hot-surface warnings; use EN 420 protective gloves for maintenance after cooldown. | whole series |
| Approved operating fluid/lubricant types are P3 / D2 / VSC100; use of non-approved fluids limits warranty and may prevent achieving rated performance. p.13 ↗ | Only specify Pfeiffer-approved fluids; consult before any application-specific fluid substitution. | whole series |
| COMBI WS WV2D variants use two DuoVane 70 backing pumps (e.g. WS 0900 WV2D: 849 m³/h at 1 hPa/50 Hz, final pressure 8.2E-04 hPa, DN160 inlet). p.14 ↗ | Dual-backing-pump configurations offer redundancy/parallel capacity at defined performance points. | COMBI WS 0245/0440/0900 WV2D |
| The operator must provide a mains isolation device (main switch or emergency-stop) enabling emergency shutdown of the unit. p.17 ↗ | Include lockable mains disconnect in the site electrical design; lockout key carried by supervisor during maintenance. | whole series |