DuoVane 1.6 — instruction manual duovane 16
Technical Reference Summary
Covered Models & Document Scope
Pfeiffer Vacuum DUO 1.6 rotary vane pump (standard version) and DUO 1.6 M (M version with non-contact, wear-proof magnetic coupling; hermetically sealed pump housing). Operating instructions cover all models/variants of the DuoLine two-stage oil-sealed rotary vane pumps; some features (e.g., magnetic coupling, voltage variants) may not be present on every unit. Related variants mentioned: C (corrosive gas) and MC versions via abbreviations.
Section-by-Section Technical Breakdown
Two-stage, oil-sealed rotary positive displacement pumps for coarse and medium vacuum with vacuum safety valve. Describes rotor/vane operating principle, exhaust valve oil sealing, safety valve and gas ballast function, and oil tasks.
- Vacuum safety valve seals the chamber at standstill and vents the pump to prevent oil migration; opens with delay after switch-on
- Gas ballast blends ambient air/inert gas to expel vapor mixtures without condensation
- Oil functions: lubrication, dead-volume filling, gap sealing, heat transfer
- Delivery includes pump with motor, operating fluid, centering rings with O-rings/cone sieve, locking caps, manual
Placement limits (±10° inclination, >35 mm clearance), vacuum and exhaust line connection via circlip flanges, mains connection with fuse values and voltage selector switch, and initial oil filling procedure.
- Max inclination ±10°, min. clearance >35 mm; keep sight glass and gas ballast valve accessible
- Exhaust pressure max 1500 hPa absolute; exhaust pressure must be ≥250 hPa above intake pressure; negative exhaust pressure only permitted on M (magnetic coupling) version
- Vacuum/exhaust line cross-section ≥ nominal flange diameter; route exhaust downwards, support piping
- Voltage selector: '115'/'1' = 115 V ±10%, '230'/'2' = 230 V ±10% (100/200 V motor accordingly)
- Recommended slow-blow fuses: 100 V 50 Hz → 4 A; 105 V (115–125 V) → 4/5 A; 100/200 V → 6/4 A; 115/230 V → 6/4 A; 220–240 V → 2.5 A
- Rated currents: e.g., 2.1 A (100 V/50 Hz), 1.15 A (220–240 V/50 Hz)
- Filler screw tightening torque: max 0.3 Nm; initial cold fill max 3/4 of min./max. sight glass range
- Approved fluids: P3 (standard), F4 (corrosive gas version), D1 (special applications); use Allen key WAF 5
Covers commissioning checks, switch-on procedure, warm-up, operating conditions, gas ballast use for vapors, refilling of operating fluid and switch-off.
- Warm up with vacuum flange closed approx. 30 minutes before process start
- At intake pressure ~300 hPa, current can exceed rated current; limit 1.5× rated current to max 2 minutes (EN 60034-1)
- Cycle operation: max 10 cycles per hour; continuous operation is the optimal condition
- Check fluid level daily in continuous operation; keep level at middle of sight glass
- Pump condensable vapors only when warm and with gas ballast open; run pump 30 min after process end
- Lowest ultimate pressures achieved with gas ballast valve closed
- After power failure/overheating the motor restarts automatically
Maintenance instructions and intervals, special notes for magnetic coupling versions, fluid change procedure including aging determination, rinsing/cleaning, gas ballast valve cleaning and changing fluid type.
- Maintenance interval tables and checklist for inspection/maintenance provided
- Special section for M/MC magnetic coupling versions
- Fluid change includes degree-of-aging determination for P3 fluid, draining, rinsing and cleaning the pump
- Gas ballast valve cleaning procedure with cap removal described
Shutting down for longer periods, recommissioning checks and disposal of the pump.
- Recommissioning requires re-checking settings (e.g., voltage range) after downtime or oil change
- Dispose of pump and contaminated fluid per local regulations
Troubleshooting table for rotary vane pumps (Tbl. 7) covering fault causes and remedies.
- Troubleshooting table on p.42-43
Spare parts packs by maintenance level: Maintenance kit 1 (level 1), radial shaft seal ring set (level 2), Maintenance kit 2 extended (level 2), overhaul kit (level 3), set of vanes, coupling set for magnetic coupling versions.
- Maintenance levels 1–3 mapped to spare part kits
- Coupling set available for magnetic coupling (M) versions
Ordering information for accessories (Tbl. 9) and consumables (Tbl. 10), e.g., oil mist filters, separators, operating fluids.
- Accessory ordering via tables 9-10 (pp. 48-49)
General data, unit conversion tables, media-wetted materials, technical data tables for Duo 1.6 (Tbl. 14) and Duo 1.6 M (Tbl. 15), and dimension drawings for standard and M versions across voltage variants.
- Materials in contact with process media listed (Tbl. 13)
- Dimension drawings for voltage variants: 100 V/50 Hz, 105 V/115–125 V, 100/200 V, 115/230 V, 230–240 V, standard and M versions
- Pressure and gas throughput conversion tables included
Key Integration, Operating & Safety Requirements
| Specification / Condition | Engineering Implication | Applies To |
|---|---|---|
| Pump is a two-stage, oil-sealed rotary vane pump for coarse and medium vacuum; example rating plate shows max pumping speed 9/10.5 m³/h (50/60 Hz), oil P3, 0.5 l, mass ~17 kg p.8 ↗ | Suitable for backing/roughing duty; check speed and oil capacity against application | Duo 1.6 series |
| Exhaust pressure limit is 1500 hPa absolute; exhaust pressure must be at least 250 hPa higher than intake pressure p.23 ↗ | Size and route exhaust lines accordingly; never fit shut-off units in exhaust line | whole series (negative exhaust pressure only tolerated on M version) |
| Negative pressure in the exhaust line is permitted only on pumps with magnetic coupling (M version) p.23 ↗ | For standard version, ensure exhaust never sees negative pressure or pump damage/malfunction results | Duo 1.6 standard vs. Duo 1.6 M |
| Ambient conditions: installation altitude max 2000 m, ambient +12 °C to +40 °C, relative humidity max 85%, intake medium temperature max +40 °C, surface inclination ±10% p.14 ↗ | Verify site conditions before installation; pump must be mounted nearly level to safeguard oil supply | whole series |
| Recommended slow-blow fuses: 100 V/50 Hz → 4 A; 115–125 V → 4–5 A; 100/200 V → 6/4 A; 115/230 V → 6/4 A; 220–240 V → 2.5 A (rated currents 2.1, 1.7, 2.4/1.5, 2.1/1.4, 1.15 A respectively at 50 Hz) p.25 ↗ | Provide on-site fuse protection per table to protect motor and supply cable | whole series, per motor voltage variant |
| Voltage selector switch: position '1' = 115 V ±10% (or 100 V), position '2' = 230 V ±10% (or 200 V); change only with pump disconnected from mains p.26 ↗ | Check/set voltage range on every relocation or recommissioning to avoid motor destruction | reversible single-phase motor variants |
| Built-in thermal protection switch interrupts motor current on excess temperature but does not provide permanent shutdown; motor restarts automatically after power failure/overheating p.13 ↗ | Secure against restart before any work near vacuum flange or rotating parts | whole series |
| At intake pressure ~300 hPa, motor current may exceed rated current; limit to 1.5× rated current for max 2 minutes (EN 60034-1) p.29 ↗ | Avoid prolonged pumping near atmospheric pressure against high counterpressure; monitor current | whole series |
| Cycle operation is limited to max 10 cycles per hour; continuous operation is the optimal operating condition p.29 ↗ | Design process so pump runs continuously or with long operating phases | whole series |
| Warm up pump ~30 minutes with vacuum flange closed before process start; run 30 minutes more after process end when pumping condensable vapors p.29 ↗ | Prevents condensation in the pump and oil contamination; use gas ballast open when pumping vapors | whole series |
| Approved operating fluids: P3 (standard), F4 (corrosive gas version), D1 (special applications like higher temperatures); only fluid from initial filling is permissible without Pfeiffer consultation p.26 ↗ | Use only the fluid specified on the rating plate; wrong fluid voids warranty and performance data | whole series |
| Oxygen concentration ≥21% requires perfluorinated oils (F4, F5, A113); D1 or mineral oil forbidden above 21% O2 p.14 ↗ | Mineral oils are combustible and oxidize in pure oxygen — safety-critical fluid selection | whole series |
| Filler screw tightening torque max 0.3 Nm; initial cold fill level max 3/4 of min./max. sight glass range p.27 ↗ | Prevents O-ring damage and overfilling; use Allen key WAF 5 and calibrated torque wrench | whole series |
| Magnetic coupling minimum distance for pacemaker/medical implant wearers is 2 m; coupling provides no overload protection for the motor p.12 ↗ | Mark M-version rooms accordingly; provide external fusing since coupling torque transmits overload | Duo 1.6 M only |
| Vacuum and exhaust line cross-sections must be at least equal to the nominal flange diameter; piping must be supported so no forces act on pump flanges p.23 ↗ | Prevents flange loading and conductance losses; connect via centering rings and circlips | whole series |
| Surface temperature can exceed 70 °C in operation and 105 °C in fault conditions; keep lines/cables away from surfaces >70 °C p.12 ↗ | Provide touch protection and cable routing accordingly | whole series |
| Storage >2 years requires changing the operating fluid; store filled to upper sight glass edge with flanges closed and gas ballast valve closed p.20 ↗ | Preserves internal corrosion protection during long storage | whole series |
| Pump includes a vacuum safety valve that seals the vacuum chamber at standstill and vents the pump so oil cannot migrate; it opens with a delay after switch-on p.17 ↗ | Chamber stays protected on unexpected stop; account for delayed pumping start | whole series (depending on pump type) |
| Removing or damaging the factory closure seal voids the warranty p.8 ↗ | Do not open sealed pump sections; limit field maintenance to defined levels/kits | whole series |
| Vacuum safety valve, lowest ultimate pressures achieved with gas ballast closed; cycle operation ≤10 starts/hour p.29 ↗ | Close gas ballast for final pressure; avoid rapid cycling duty | whole series |