HiScroll 46 — instruction manual 1ph atex
Technical Reference Summary
Covered Models & Document Scope
Pfeiffer Vacuum HISCROLL 46 scroll vacuum pump operating instructions (82 pages, EN, translation of original). Covers two variants: HiScroll 46 standard (corrosive gas version) incl. ATEX, and HiScroll 46 without gas ballast incl. ATEX. Part number on rating plate: P/N 2000075017. Options addressed: manual or automated gas ballast valve, retrofit RPT 010 pressure sensor, 'remote' 15-pin interface, RS-485.
Section-by-Section Technical Breakdown
Risk assessment-based safety instructions for transport, installation, operation, maintenance and malfunctions, plus the ATEX classification (Directive 2014/34/EU) and limits of use. Defines qualified personnel levels and improper use exclusions.
- ATEX marking: II 3/-G Ex h IIC T4 Gc X, +5 °C ≤ Ta ≤ +40 °C; Category 3G (Zone 2 indoor only), installation only outside Ex zones
- X condition: medium ignition temperature must stay above +135 °C at 3500 hPa abs on exhaust side; only ATEX-approved accessories permitted
- Limits of use: air pressure 750-1060 hPa, altitude max. 3000 m, installation surface flatness ±10°, rel. humidity max. 90% non-condensing, protection class I, overvoltage category II, pollution degree 2
- After switching off, wait 5 minutes for capacitor discharge before maintenance
- IP40 per DIN EN 60529:2014 — no water ingress protection; RCD fuse protection recommended in hazardous areas
- Improper use includes: pumping fluids, dust, corrosive media (except C version), use as pressure generator, use in ionizing radiation, separating/reusing the electronic drive unit
Explains the dry-running scroll principle (stationary + orbiting spiral), integrated drive/control unit, and design components. Describes drive (INFORM sensor-free synchronous motor), air cooling, hermetically separated ball bearings, gas ballast versions and the retrofittable RPT 010 pressure sensor, plus electronic drive unit connections.
- Suction chamber completely free of lubricants (dry operation)
- Drive: highly efficient synchronous motor with sensor-free control (patented INFORM procedure per Prof. Schrödl)
- Shaft bearing: 2x maintenance-free deep groove ball bearings, hermetically separated from working chamber
- Gas ballast: 3-stage system; automated GB valve and RPT 010 sensor are NOT ATEX-approved — explosive media not permitted with these options
- Connections: mains input, 15-pin D-sub 'remote' socket, standby button [P:002], On/Off button [P:010], 2 accessory sockets (C, D)
- Nominal pumping speed 43 m³/h for both versions; scope of delivery includes pump, protective caps/covers, operating instructions
Covers setting up the pump, connecting the vacuum side, exhaust side, external gas ballast supply and mains power. Safety requirements include earthing, double-insulated cables per IEC 61010/60950, and inert-gas-only ballast supply.
- Use only 3-pin mains cable with proper PE; connect mains cable before all other cables
- Mains voltage/frequency must match rating plate (100-127 or 200-240 VAC ±10%)
- For explosive media: protective strainer at vacuum connection; inert gases only for ballast; closed inert gas system
- Exhaust line must be routed without shut-off units; observe max. permissible exhaust pressure
Describes the 15-pin 'remote' D-sub inputs/outputs and the RS-485 interface with the Pfeiffer Vacuum protocol (telegram frame, telegram examples, data types), plus control unit connection.
- 15-pin D-sub 'remote' connection for inputs/outputs (Tbl. 6, p.34)
- RS-485 interface features (Tbl. 7, p.35); Pfeiffer Vacuum protocol with telegram frames and examples (pp.36-37)
Documents the Pfeiffer Vacuum parameter set: general parameter table [P:xxx], control commands, status requests, and reference value inputs for remote operation via RS-485/remote interface.
- Key parameters: [P:002] standby, [P:010] On/Off, [P:019]/[P:024] output/input config, [P:063] interface selection, [P:060] accessory configuration
- Correction factors apply to displayed pressure below 1 hPa (Tbl. 15, p.48)
Commissioning and operation of the pump: switching on/off, configuring digital I/O and interfaces, operating modes, speed modes (normal, standby, rotation speed setting, auto standby, auto start), pressure regulation with RPT 010, and gas ballast valve control with/without pressure sensor. LED status indication is documented.
- Speed modes: normal operation, standby, rotation speed setting, auto standby, auto start
- Gas ballast valve has snap positions '0', '1', '2' only — intermediate positions leak process media
- Pressure-dependent automatic gas ballast valve control possible with pressure sensor
- Operating mode display via LEDs of the electronic drive unit (Tbl. 17-18, pp.52-53)
Maintenance intervals, inspection checklist, and step-by-step procedures for replacing valves (with fan cover removal), replacing the gas ballast valve, and changing the tip seals (spiral housing disassembly/reassembly), plus final inspection.
- Maintenance intervals table (Tbl. 19, p.55)
- Procedures: valve replacement (pp.55-58), gas ballast valve replacement (pp.58-60), tip seal replacement via spiral housing (pp.61-63), final inspection (p.64)
- Before any maintenance: disconnect from mains, wait for standstill (f = 0) plus 5 min capacitor discharge, vent to atmosphere, secure against restart
Troubleshooting table for scroll pumps, error messages and warning messages of the electronic drive unit, and warning/malfunction messages when operating with control units.
- Troubleshooting table (Tbl. 20, p.68)
- Error messages of electronic drive unit (Tbl. 21, p.69)
- Warning messages (Tbl. 22, p.70) and messages with control units (Tbl. 23, p.70)
Lists spare part packages (tip seal maintenance set, valve set) and accessories for HiScroll 46 with ordering information. Accessories used with explosive media must comply with Directive 2014/34/EU.
- Spare part packages: tip seal maintenance set (Fig. 33, p.73), HiScroll valve set (Fig. 34, p.74)
- Accessories for HiScroll 46 (Tbl. 26, p.76); RPT 010 pressure sensor retrofittable
Technical data tables for HiScroll 46, conversion tables for pressure and gas throughput units, substances in contact with process media, dimension drawing, and EC/UK Declarations of Conformity.
- Technical data for HiScroll 46 (Tbl. 29, p.78)
- Materials in contact with process media (Tbl. 30, p.78); FKM/FFKM/PTFE abbreviations apply
- EC Declaration of Conformity (p.80) and UK Declaration of Conformity (p.81) included
Key Integration, Operating & Safety Requirements
| Specification / Condition | Engineering Implication | Applies To |
|---|---|---|
| Nominal pumping speed S(N2) is 43 m³/h (max. 44 m³/h on rating plate) at max. 1800 1/min for both HiScroll 46 variants. p.9 ↗ | Sizing of backing/roughing vacuum systems; select pump based on 43 m³/h nominal N2 throughput. | whole series |
| Power supply: 100-127 VAC ±10% 50/60 Hz at 15.0 A, or 200-240 VAC ±10% 50/60 Hz at 8.0 A; SCCR 500 A @ 250 VAC, 10 kA @ 125 VAC. p.9 ↗ | Dimension mains supply and branch protection accordingly; dual voltage range requires correct cable/fuse selection per range. | whole series |
| ATEX marking: II 3/-G Ex h IIC T4 Gc X with permissible ambient temperature +5 °C ≤ Ta ≤ +40 °C. p.18 ↗ | Pump may be used in Zone 2 (Category 3G) indoors only; installation location itself must be non-Ex; medium ignition temperature must exceed T4 limit. | whole series (ATEX variants) |
| When installed with exhaust-side backpressure, the medium's ignition temperature must remain above +135 °C at 3500 hPa abs. p.14 ↗ | Limit exhaust-side pressure build-up when pumping flammable media; route exhaust without shut-off units. | whole series |
| Protection degree IP40 per DIN EN 60529:2014 — no water ingress protection. p.13 ↗ | Do not install where water contact is possible; provide on-site RCD protection; in water damage events disconnect all power. | whole series |
| After switching off, capacitors need 5 minutes to discharge before maintenance work. p.16 ↗ | Add a mandatory 5-minute dead time to all service SOPs before opening or touching electrical parts. | whole series |
| Installation limits: air pressure 750-1060 hPa, altitude max. 3000 m, surface flatness ±10°, rel. humidity max. 90% non-condensing, overvoltage category II, pollution degree 2, protection class I. p.19 ↗ | Verify site conditions before installation; high-altitude or humid sites require derating/consideration. | whole series |
| Surface temperature can exceed 70 °C during operation. p.14 ↗ | Install touch protection if pump is freely accessible; allow cool-down before any work; keep cables away from hot surfaces (>70 °C). | whole series |
| Gas ballast system is vacuum-sealed only in snap positions '0', '1', '2'; intermediate positions leak. p.15 ↗ | Only operate/change valve in fully snapped positions; mandatory when handling toxic media to prevent escape. | variants with gas ballast |
| The automated gas ballast valve (electromagnetic) and the RPT 010 pressure sensor have no ATEX approval. p.23 ↗ | Do not use automated GB or the pressure sensor when pumping explosive atmospheres; use manual GB with inert gas instead. | ATEX applications |
| Only inert gases may be used as gas ballast supply, especially with explosive media; the inert gas system must be closed. p.12 ↗ | Design ballast supply as a sealed inert-gas line (e.g. N2) to avoid explosive mixture formation and zone entrainment. | whole series |
| Working chamber is completely dry/lubricant-free; shaft bearing uses 2x maintenance-free deep groove ball bearings hermetically separated from the chamber. p.22 ↗ | No oil changes for the suction chamber; pump suitable for oil-sensitive processes; bearings not user-serviceable. | whole series |
| Drive is a highly efficient synchronous motor with sensor-free control (patented INFORM procedure per Prof. Schrödl); fan cooling is demand-regulated by the electronic drive unit. p.23 ↗ | Speed/cooling is managed internally; no external motor feedback or additional cooling provision needed. | whole series |
| Pump mass is 32 kg; lifting only via the factory-fitted eye bolt with a lifting device; never stack products. p.9 ↗ | Plan lifting equipment and rigging for 32 kg; improper lifting damages fan cover fastener and risks pump fall. | whole series |
| Before opening the vacuum side on explosive media service, a protective strainer must be fitted; gas ballast inlet must be cleaned by brief slight-overpressure purging with valve closed. p.12 ↗ | Prevent particle ingress as ignition source during commissioning with explosive gases. | ATEX applications |
| Mains cable and extensions must satisfy double isolation requirements per IEC 61010 and IEC 60950; use only 3-pin cables with PE; connect mains first. p.12 ↗ | Electrical installation must be executed by qualified electricians to these standards to avoid fatal shock hazards. | whole series |
| Corrosive media pumping is permitted only for the C (corrosive gas) version; corrosive use on standard version voids conformity. p.19 ↗ | Select the C variant for corrosive processes; standard variant restricted to dry, inert gases (moist gases with gas ballast). | HiScroll 46 C version vs standard |
| Separating, replacing or reusing the electronic drive unit is defined as foreseeable improper use; damaging the factory warranty seal voids warranty. p.20 ↗ | Never swap drive electronics between pumps; return to Pfeiffer Vacuum service for drive unit issues. | whole series |
| Key remote-control parameters: [P:002] standby button, [P:010] On/Off, [P:019]/[P:024] digital output/input configuration, [P:063] interface selection, [P:060] accessory configuration. p.24 ↗ | Use these parameter codes for PLC/remote integration via the 15-pin D-sub or RS-485 with the Pfeiffer Vacuum protocol. | whole series (electronic drive unit) |
| Automated gas ballast valve is documented in a separate operating instruction PU 0104 BN. p.8 ↗ | Order/download PU 0104 BN when using the automated gas ballast option. | variant with automated gas ballast |