SmartVane 55 — instruction manual _70
Technical Reference Summary
Covered Models & Document Scope
Pfeiffer Vacuum SmartVane 55 and SmartVane 70 single-stage oil-sealed rotary vane pumps (operating instructions, 68 pages, EN translation). Covers both variants with electronic drive unit, 'remote' 15-pin D-sub interface, RS-485 with Pfeiffer Vacuum protocol, accessory connections, operating fluid variants H1 (standard) and F4 (oxygen >21%). Example rating plate shown: SmartVane 55, 0.2 hPa ultimate pressure, 800–1350 min⁻¹, DEM121290002 oil, order-related part 'PK D80 001'.
Section-by-Section Technical Breakdown
Risk assessment based on Machinery Directive 2006/42/EC, DIN EN ISO 12100 and DIN EN 1012-2, covering transport, installation, operation, maintenance and disposal hazards (electric shock, toxic process media, hot surfaces, tilting, exhaust line bursting). Defines proper/improper use and required personnel qualification levels. Lists installation limits such as ambient conditions and protection degree.
- Installation limits: indoors, dust-protected; altitude max. 4000 m; horizontal orientation with ±1° inclination; ambient +15 to +45 °C; gas inlet temp max. +40 °C; relative humidity max. 80%; protection degree IP20 / Nema Type 1 (UL 50E); degree of pollution 2
- Continuous operation permitted at pressures <15 hPa unlimited; exhaust at atmospheric pressure
- SmartVane 70: avoid 100–127 V ±10% mains range — 10 A input current limit prevents full rotation speed
- Improper use includes corrosive, radioactive, explosive media, condensable vapors, fluids, dust-laden gases, and use as compressor or climbing aid
- Oxygen concentration >21%: only perfluorinated synthetic oils (F4) permitted; mineral/H1 oils forbidden
- Surface temperature can exceed 70 °C in operation and 105 °C in fault conditions; touch protection required if accessible to untrained persons
- Motor restarts automatically after power failure or overheating standstill; secure against restart before work
- Maintenance level 1 (customer with technical education or Pfeiffer service); level 3 (Pfeiffer service technician only)
Describes the single-stage, oil-sealed rotary vane principle with air cooling and circulatory lubrication, including the non-return valve in the vacuum connection and integrated oil mist filter. Drive is a sensor-free controlled synchronous motor (INFORM procedure). Identifies connections, features and scope of delivery for both models.
- SmartVane 55: nominal pumping speed max. 50 m³/h; SmartVane 70: max. 70 m³/h; both with small-flange vacuum/exhaust connections and electronic drive unit
- SmartVane 70 has 2 axial fans; fan speed regulated by electronic drive unit according to cooling requirement
- Non-return valve closes intake line automatically at switch-off, preventing operating fluid backflow
- Connections: mains input, 15-pin D-sub 'remote', stand-by button [P:002], On/Off button [P:010], 2 accessory sockets
- Operating fluid tasks: lubrication, dead-volume filling under exhaust valve, sealing inlet/outlet gap and vane/chamber gap, heat transfer
- Scope of delivery: pump with electronic drive unit, operating fluid, locking caps/blank flange plugs, operating instructions
Setup requires a flat horizontal surface with ≥200 mm clearance on all sides, visible sight glass and rating plate. Vacuum and exhaust lines must be at least the nominal connection diameter and supported so no pipeline forces act on the pump. Operating fluid filling and mains connection procedures with torque and electrical safety requirements are given.
- Minimum 200 mm distance to lateral obstructions on all four sides
- Exhaust line routed downwards, no shut-off units; condensate separator with drain at lowest point recommended
- Threaded connector max. torque: 12 Nm (calibrated torque wrench, tightening factor ≤ 2.5)
- Filler screw tightening torque: 37 Nm (Allen key WAF 12 mm); optimal fill level is middle of sight glass
- Approved fluids: H1 (standard) and F4; use only type specified on rating plate; if factory-filled, level may exceed sight glass — tilt pump ~10° toward sight glass to check
- Mains: 1~ 100–127 V / 200–240 V ±10%, 50/60 Hz, 10 A, IP20 (rating plate example); 3-pin cable with PE, double insulation per IEC 61010/60950; connect mains cable first; overload protection required; drive unit starts with mains connection
The electronic drive unit offers a 15-pin 'remote' D-sub with configurable digital inputs/outputs and an RS-485 interface using the Pfeiffer Vacuum protocol. Parameters configure DI/DO behavior and interface selection; telegram frame, data types and examples are documented.
- Parameters for configuration: [P:019], [P:024] (digital outputs), [P:063] (digital input), [P:060] (interface selection)
- RS-485 enables cross-linking of multiple pumps via connection cables and accessories
- Accessories connect via two dedicated accessory sockets on the drive unit
Parameter set covers control commands, status requests and reference value inputs via the remote/RS-485 interface. Commissioning includes switching on, fluid level checks, interface configuration, operating mode and speed mode selection (normal, standby, speed actuator), with LED indication of operating state.
- Speed modes: normal operation, standby mode, speed actuator operation
- Stand-by button [P:002]; On/Off button [P:010] switches pump and configured components
- LEDs on drive unit (incl. standby LED) indicate operating mode; LED behavior tables provided (Tbl. 15/16)
- Rated speed range 800–1350 min⁻¹ (rating plate example)
Maintenance includes fluid level checks and changes, rinsing/cleaning, and servicing the operating fluid separator (exhaust filter removal and remounting). Maintenance intervals are given in a checklist.
- Fluid change: drain, fill, rinse and clean pump; procedures with figures provided
- Exhaust filter (operating fluid separator) dismantling and mounting procedures; separator locking screws only opened with pump switched off (hot oil mist hazard)
- Maintenance interval checklist (Tbl. 17) provided
Shutdown procedures for longer periods and recommissioning steps, plus general disposal guidance and pump-specific disposal including decontamination of toxic process media residues.
- Decontaminate parts exposed to toxic process media before maintenance/disposal
- Observe locally applicable regulations for disposal of contaminated operating fluid
Troubleshooting tables for the rotary vane pumps plus error and warning message lists of the electronic drive unit allow diagnosis and remedy of faults.
- Troubleshooting table for pumps (Tbl. 18)
- Error messages (Tbl. 19) and warning messages (Tbl. 20) of the electronic drive unit
Accessory list with ordering information (including operating fluids) and technical data tables for SmartVane 55/70, pressure/gas-throughput unit conversion tables, and dimension drawings for both models. Declarations of Conformity (EC and UK) are appended.
- Technical data table Tbl. 25 for SmartVane 55 / SmartVane 70; pressure and gas throughput conversion tables
- Dimension drawings: Fig. 17 (SmartVane 55), Fig. 18 (SmartVane 70)
- Example rating plate data (SmartVane 55): 37 kg mass, 1.75 L oil quantity, S(N!)max = 50 m³/h
Key Integration, Operating & Safety Requirements
| Specification / Condition | Engineering Implication | Applies To |
|---|---|---|
| Nominal pumping speed: SmartVane 55 max. 50 m³/h; SmartVane 70 max. 70 m³/h; rated speed 800–1350 min⁻¹ (rating plate example) p.19 ↗ | Size backing/roughing pump selection by variant; speed mode configurable between 800 and 1350 rpm | SmartVane 55 / 70 |
| Ultimate total pressure 0.2 hPa (mbar) abs per rating plate example p.8 ↗ | Defines achievable base pressure for process design | SmartVane series |
| Mains supply 1~ 100–127 V / 200–240 V ±10%, 50/60 Hz, 10 A, IP20 p.8 ↗ | Requires 10 A-rated supply circuit; IP20 means no water protection — install in dry indoor locations, use RCD in hazardous areas | whole series |
| SmartVane 70 must avoid the 100–127 V ±10% mains range: at max. 10 A input current and temperature limits, full rotation speed is not achieved p.16 ↗ | For full 70 m³/h performance, connect the 200–240 V range on SmartVane 70 | SmartVane 70 only |
| Ambient temperature +15 to +45 °C, gas inlet max. +40 °C, relative humidity max. 80%, altitude max. 4000 m, horizontal installation with ±1° inclination p.15 ↗ | Verify installation site conditions before commissioning; pump must be leveled within 1° | whole series |
| Continuous operation permitted at pressures <15 hPa without limitation; exhaust at atmospheric pressure p.15 ↗ | Do not operate continuously above 15 hPa inlet pressure; no compression duty (use as compressor forbidden) | whole series |
| Minimum 200 mm clearance required on all four lateral sides of the pump p.23 ↗ | Reserve installation footprint for airflow and service access | whole series |
| Filler screw tightening torque 37 Nm; exhaust threaded connector max. 12 Nm; torque wrench tightening factor ≤ 2.5 p.26 ↗ | Prevents thread/seal damage during fluid filling and exhaust hookup; use calibrated torque wrench | whole series |
| Approved operating fluids: H1 (standard) and F4; for oxygen concentration >21% only F4 perfluorinated synthetic oil permitted p.16 ↗ | Never run oxygen-rich processes on H1/mineral oil — combustion/oxidation hazard; use rating-plate-specified fluid only or warranty is void | whole series |
| Oil capacity 1.75 L (rating plate example); optimal fill level is middle of sight glass; check level via sight glass, tilt pump ~10° toward sight glass if factory fill exceeds visible range p.8 ↗ | Maintain fill level mid-sight-glass during operation to ensure sealing and lubrication | SmartVane 55 (example); check rating plate per unit |
| Surface temperature can exceed 70 °C in normal operation and 105 °C in fault conditions p.12 ↗ | Provide touch protection for untrained-person access; allow cooldown before any work; keep cables >70 °C away from hot surfaces | whole series |
| Motor restarts automatically after power failure or overheating-related standstill; with open control input and mains applied the motor can start unexpectedly p.12 ↗ | Lockout/tagout and secure against restart before flange or maintenance work; integrate into emergency-off circuit | whole series |
| Exhaust line must be routed without shut-off valves, downwards from the pump, min. diameter equal to nominal exhaust connection (DN 25, G 1/2" internal thread); condensate separator recommended at lowest point p.11 ↗ | Prevents exhaust overpressure bursting hazard and condensate return into pump | whole series |
| Vacuum connection DN 40 with protective cap; vacuum line min. diameter equal to nominal connection diameter; pipeline must be supported so no forces act on the pump p.18 ↗ | Undersized lines reduce pumping speed; unsupported piping transmits stress to pump flanges | whole series |
| Non-return valve in the vacuum connection closes automatically at switch-off, preventing operating fluid backflow into the vacuum line p.18 ↗ | No external isolation valve strictly needed for pump-stop backstreaming, but still isolate for maintenance | whole series |
| Operating fluid usability is limited to max. 2 years; after ≥2 years of shutdown, change fluid and replace all elastomer parts before recommissioning p.22 ↗ | Plan storage life and elastomer kit replacement for spare pumps | whole series |
| 15-pin D-sub 'remote' interface with configurable DI/DO plus RS-485 (Pfeiffer Vacuum protocol); parameters [P:019]/[P:024] (outputs), [P:063] (input), [P:060] (interface selection) p.29 ↗ | Integrate pump control/monitoring into PLC via remote I/O or RS-485 bus; multiple pumps can be cross-linked on RS-485 | whole series |
| Operating modes: normal operation, standby mode (button [P:002]), speed actuator mode; On/Off via [P:010]; state shown by drive-unit LEDs p.19 ↗ | Use standby mode to reduce speed/consumption during idle process phases; diagnose state via LED tables | whole series |
| Pump mass 37 kg (SmartVane 55 example); lift only via crane lug; transport horizontally or without fluid; never stack p.8 ↗ | Plan lifting gear and horizontal handling; tipping the pump fills oil into the pumping system and can break vanes at next start | whole series (mass varies per model — check packaging/rating plate) |
| Sensor-free synchronous motor drive using the INFORM procedure (patented, Prof. Schrödl); SmartVane 70 has two axial fans whose speed is regulated by the drive unit p.18 ↗ | No encoder maintenance; cooling demand-adaptive fan speed affects noise/heat budget of the enclosure | whole series; dual fans on SmartVane 70 only |