TORRI BD 0100 A — instruction manual torri bd 0100 0300 0600 a
Technical Reference Summary
Covered Models & Document Scope
Busch TORRI BD series Multi-Roots dry vacuum pumps, models BD 0100 A (100 m³/h), BD 0300 A (300 m³/h) and BD 0600 A (600 m³/h). BD 0300/0600 combine a backing pump (DP) with a Roots booster (MB). Options: LCD controller, Safety Circuit External Enclosure (SEMI S2, EMO), N2 flow regulator, cooling water valve. Two supply voltage ranges: 200-220 V and 380-415 V.
Section-by-Section Technical Breakdown
Verifies nameplate specification and delivered accessories, and describes moving the pump via four caster/height-adjustment-foot units. Lifting must be done with a hoist using lifting brackets; forklift lifting from underneath damages the bottom-mounted silencer.
- Supplied accessories: instruction manual, one power connector (female), KF50 inlet centering ring, one set of RC 1/4" cooling water quick-connect couplers
- Before moving, set all four height-adjustment feet to their highest positions
- Lifting brackets are included in the pump package
Step-by-step commissioning sequence: verify supply voltage and NFB/wire sizing, level the pump on its adjustable feet (wheels 3-5 mm off floor, M16 nuts), check gear oil level (1/2-2/3 of sight glass), connect vacuum/exhaust/cooling water, wire power, then start and verify current and temperatures after 4 hours.
- NFB/wire sizing: BD 0100 A: 15 A / 10 A, 14 AWG (105 °C); BD 0300 A and BD 0600 A: 30 A / 20 A, 14 AWG (105 °C) for 200-220 V / 380-415 V
- Cooling water minimum: 1 L/min (15-28 °C) for BD 0100 A; 2 L/min (15-28 °C) for BD 0300 A / BD 0600 A
- For SEMI remote operation, set LCD controller control mode to "SEMI" and wire per Section 3.5.2 Tool Interface Connector
- After 4 hours of running, check motor and pump casing temperature
The pumps are dry, positive-displacement rotary Multi-Roots pumps; non-contacting rotors require no grease for lubrication/sealing, eliminating oil back-permeation. 5-stage, 2-lobe rotor design for low power consumption. Component layout and cooling water routing are given for each model; stainless-steel piping prevents cooling-water contamination.
- BD 0100 A: inlet NW50, exhaust NW25, 3HP motor; BD 0300/0600 A: inlet ISO80, exhaust NW40
- BD 0100 A cooling path: inlet → gear box & inverter → motor → water flow meter → outlet
- BD 0300/0600 A cooling path: inlet → gear boxes of DP & MB → inverter → motor → flow meter → outlet
- Insufficient water flow causes warning when motor/casing temperature is normal, and alarm (pump stop) when temperature exceeds the default setting
Control consists of frequency inverter and controller with LCD controller (option), Modbus TCP and Ethernet ports. Two protection levels: Warning (pump keeps running) and Alarm (immediate shutdown). Warning/alarm history is recorded for traceability.
- LCD controller provides parameter setting, status display, warning/alarm messages and start/stop
- Ethernet port allows remote start/stop with external signals; Modbus TCP (RJ45) allows PC monitoring - protocol available from Busch
Full specification tables for the three models, covering pumping speed, ultimate pressure, motor data, connections, cooling water requirements, dimensions, weight, noise, leak rate, environment and lubrication.
- BD 0100 A: 1667 L/min (100 m³/h), ultimate pressure 1×10⁻² mbar, 2.2 kW, 11 A (220 V)/6.3 A (380 V), 450×230×275 mm, 60 kg
- BD 0300 A: 5000 L/min (300 m³/h), ultimate 1×10⁻³ mbar; BD 0600 A: 10000 L/min (600 m³/h), ultimate 1×10⁻³ mbar; both 530×230×520 mm, 60 kg
- BD 0300 A motor frequencies: booster 110.0 Hz / backing 183.3 Hz; BD 0600 A: booster 233.3 Hz / backing 183.3 Hz
- Cooling water: max pressure 4.0 kg/cm² (57 PSI), differential 1.0 kg/cm², min flow >1.0 (BD0100) / >2.0 L/min (BD0300/0600), 15-28 °C, RC 1/4" connection
- Noise <60 dB(A); system max leak rate 1×10⁻⁵ mbar·l/s; ambient 5-40 °C, ≤90% RH; oil: Busch YLC 250 B, 0.1 L (level 1/2-2/3 of indicator)
Indoor installation with ventilation, ≥900 mm clearance around/top, ≥300 lux illumination, ambient ≤40 °C, on a floor able to carry the weight with shock-absorbing pads. Piping rules minimize conductance losses; bellows recommended at inlet and exhaust; helium leak check limit 1×10⁻⁵ mbar·l/s. Electrical connection tolerances and SEMI F47 compliance are specified, plus 50-pin D-sub interface pin assignments and Modbus TCP port.
- Voltage within ±10%, voltage unbalance within 10%, frequency within ±5%
- System passed SEMI F47-0812R; automatic recovery run within 1 second of voltage interruption; harmonics per IEC 61000-3-2 Class A
- Mains connection: Han® C module, 40 A, 2.5-8 mm²
- Interface: 50-pin female D-sub (SEMI E73-0299); relay contacts RL1-RL5 rated 5 A 250 VAC / 30 VDC; remote start via pins 37/38 (dry contact or open collector), Process ON via pins 41/42; EMO on pins 39/40 (with SEMI S2 box)
- Cooling water: never stop flow during operation; keep cooling ≥10 minutes after pump stop; never connect pump loops in series; below 0 °C purge residual water with high-pressure air before shutdown
- Exhaust: open exhaust valves before starting; bellows and O-ring-capture centering rings recommended; exhaust must not vent to atmosphere with explosive/corrosive/toxic gases
Simple control panel provides POWER/ALARM/WARNING LEDs, START/STOP buttons and a LOCAL lockout button (5 s press). MMI controller (RJ45) loads pump parameters. LCD panel shows currents, temperatures, water flow, N2 mass flow, exhaust pressure; user setup 12.1.x (units, MB, N2 flow, PC link) and factory settings 12.2 for warning/alarm thresholds.
- Local mode (12.1.2 = LOCAL): START/STOP keys on LCD panel; SEMI mode: external signal control only
- Temperature unit selectable (°C/°F) at 12.1.18; pressure units Torr/mbar/kPa/psi/kgf at 12.1.19
- Water-flow protection without temperature (12.2.4.12 DI=0): low flow gives W06 warning; if motor/casing temperature exceeds warning setting, A06 alarm stops the pump
- After stopping process gases, do not stop the pump for at least 30 minutes to prevent corrosive gas/byproduct retention
Table of W/A message codes with causes and corrective actions, covering current-high, motor/casing over-temperature, water-flow-low, maintenance renewal due, communication errors and motor overload/fail codes for DP (backing) and BP (booster) pumps.
- W01(A01) DP Current High: exhaust pressure rise, rotor contact, power supply failure
- W03(A03)/W31(A31) motor temp high, W04(A04)/W32(A32) casing temp high: check cooling water flow/temperature
- W06(A06) Water Flow Low: leaks, small differential pressure, clogged piping, reversed in/outlet, flow meter failure
- W09/W10 DP/BP Renew Time: overhaul due; W23 Comm. Error: replace PLC/VFD; A16/A17 motor fail (lose speed): power off/restart or overhaul; A19/A20 motor overload
Safety rules for qualified personnel including lockout/tagout procedure, PPE, and handling of toxic residue and fluorinated-material breakdown products (dangerous if pump overheated to ≥260 °C). Oil change procedure, plate dismantle/assembly, cooling-water cleaning, decontamination and scrap procedures, and a periodic maintenance plan (weekly to 24000 hr).
- Gear box oil level must be 1/2-2/3 of indicator; check monthly; replace oil every year or half-year depending on process; oil is fluorinated lubricant (Busch YLC 250 B) - no substitution allowed; never fill oil unless pump is at atmospheric pressure
- O-rings must not be reused; replacements in Busch PM kit package
- Replace oil at 8000 hr per maintenance plan; bearings and O-rings complete overhaul at 2 years/24000 hr; weekly checks of controller, temperatures, current, water flow; monthly: piping, oil level/color, inlet flange, silencer
- Keep cooling water running ≥10 min after pump stop; close cooling water and N2 supply before removing piping
- Pump is only for load lock - no special decontamination procedures required; shut down ≥30 min after process gas stop
The Safety Circuit External Enclosure option provides full EMO function per SEMI S2 with leakage circuit breaker and electromagnetic contactor; power reset button restarts pump. EU and UK Declarations of Conformity are issued by Busch Manufacturing Korea for all three models under Machinery, EMC and RoHS directives.
- Safety enclosure includes EMO signal connection port, emergency stop button, power reset button, leakage breaker, magnetic contactor
- EU conformity: 2006/42/EC (Machinery), 2014/30/EU (EMC), 2011/65/EU (RoHS)
- Manufacturer: Busch Manufacturing Korea, Ltd., Icheon-si, Gyunggi-do, Republic of Korea
Key Integration, Operating & Safety Requirements
| Specification / Condition | Engineering Implication | Applies To |
|---|---|---|
| Pumping speeds: BD 0100 A = 100 m³/h (1667 L/min), BD 0300 A = 300 m³/h, BD 0600 A = 600 m³/h; ultimate pressures 1×10⁻² and 1×10⁻³ mbar respectively (BD0300/0600: 7.5×10⁻⁴ Torr) p.13 ↗ | Model selection by required throughput and base pressure; BD0300/0600 offer one decade lower ultimate pressure via booster stage | whole series |
| Power supply: 220 V ±10% or 380 V ±10%, rated current 11 A (220 V) / 6.3 A (380 V) per motor, rated power 2.2 kW per stage p.7 ↗ | Size supply, NFB (15/30 A at 200-220 V; 10/20 A at 380-415 V) and cabling (14 AWG, 105 °C rating) accordingly | whole series |
| Cooling water minimum flow: 1 L/min for BD 0100 A, 2 L/min for BD 0300/0600 A, at 15-28 °C; max pressure 4 kg/cm² (57 PSI); RC 1/4" quick couplers p.14 ↗ | Chiller/water circuit must guarantee these flows at 15-28 °C; do not exceed 57 PSI; do not串connect pump loops in series | whole series |
| Keep cooling water flowing at least 10 minutes after pump stop; never interrupt cooling during operation p.20 ↗ | Post-purge timer or interlock needed in facility water circuit to prevent overheating damage | whole series |
| Pump is not suitable for continuous duty at pressure ≥60 mbar; suitable for thermally evacuating a 300 L tank; Argon capacity up to 10 slm p.5 ↗ | Avoid sustained high-pressure operation in process design; cycle/load-lock applications are the intended duty | whole series |
| Ambient operating conditions: 5-40 °C, ≤90% RH, indoor installation with ≥900 mm maintenance clearance and ≥300 lux illumination p.14 ↗ | Installation site must meet these limits; ≥900 mm clearance required around and above pump | whole series |
| System max leak rate: 1×10⁻⁵ mbar·l/s (helium leak test recommended limit) p.14 ↗ | Acceptance criterion for post-installation leak checks | whole series |
| Connections: BD 0100 A inlet NW50 / exhaust NW25; BD 0300/0600 A inlet ISO80 / exhaust NW40 p.8 ↗ | Matching flange hardware and centering rings needed; bellows recommended at inlet and exhaust | BD 0100 A / BD 0300 A / BD 0600 A |
| Dimensions/weight: BD 0100 A 450×230×275 mm, 60 kg; BD 0300/0600 A 530×230×520 mm, 60 kg p.6 ↗ | Floor loading and footprint planning; lifting by hoist only (forklift from below damages silencer) | whole series |
| Noise <60 dB(A); system leak rate and environmental data per nameplate; gear oil Busch YLC 250 B, 0.1 L, level 1/2-2/3 of sight glass p.13 ↗ | Only fluorinated Busch oil may be used; refill when below 1/2 of indicator; check monthly | whole series |
| Electrical tolerances: voltage ±10%, voltage unbalance ≤10%, frequency ±5%; SEMI F47-0812R passed with auto-recovery within 1 s of voltage interruption; harmonics per IEC 61000-3-2 Class A p.21 ↗ | Facility power quality requirements; pump rides through F47 sag events without shutdown | whole series |
| Mains connector: Han® C module, 40 A, wire 2.5-8 mm²; remote interface: 50-pin female D-sub per SEMI E73-0299 with relay contacts rated 5 A 250 VAC / 30 VDC p.21 ↗ | Specify mating connectors and keep relay loads within contact rating when wiring tool interface | whole series |
| Remote start via pins 37/38 (dry contact or open collector), Process ON via pins 41/42; control mode must be set to SEMI in LCD controller p.22 ↗ | SEMI-mode tool integration wiring and mode setting required before remote operation works | whole series |
| Motor frequencies (canned motor): backing pump 183.3 Hz; booster 110.0 Hz (BD 0300 A) / 233.3 Hz (BD 0600 A) p.13 ↗ | Inverter-driven motors run above line frequency - do not judge speed by mains frequency; VFD parameter expectations differ per model | BD 0300 A / BD 0600 A |
| Stop the pump only ≥30 minutes after stopping process gas flow to prevent corrosive gas/byproduct retention p.31 ↗ | Include a 30-min post-run delay in the tool/pump sequencing logic | whole series |
| Cooling water protection logic: insufficient flow with normal motor/casing temperature → warning (pump keeps running); with high temperature → alarm and stop; protection behavior configurable at 12.2.4.12 DI setting p.11 ↗ | Water-flow warnings can be tolerated briefly, but temperature rise forces shutdown; set parameters per protection strategy | whole series |
| Oil replacement at 8000 hr; bearing and O-ring complete overhaul at 2 years / 24000 hr; oil level/color and inlet flange checked monthly p.37 ↗ | Build preventive maintenance schedule around these intervals; adjust frequency per process harshness | whole series |
| Below 0 °C ambient, remove cooling water from pipes at shutdown (purge with high-pressure air after disconnecting quick connectors) to prevent freeze burst p.4 ↗ | Required winterization step for unheated installations | whole series |
| EMO status (pins 39/40) requires the optional SEMI S2 Safety Circuit External Enclosure (leakage breaker + magnetic contactor); pump restart via power reset button p.24 ↗ | For SEMI S2 compliance, this enclosure must be ordered and wired into the EMO chain | whole series (option) |
| Fluorinated materials (oils, greases, seals) thermally decompose at ≥260 °C producing dangerous products - do not touch or inhale if pump overheated p.33 ↗ | Safety procedure for post-overheat or fire maintenance; use PPE (activated carbon mask, silicone gloves) | whole series |