HiLobe 11003 — product leaflet hilobe 8103 14003 global
Technical Reference Summary
Covered Models & Document Scope
Pfeiffer Vacuum HiLobe series three-lobe vacuum boosters datasheet covering three models: HiLobe 8103, HiLobe 11003 and HiLobe 14003, all driven by the RC 15000 electronic drive unit. Accessories/options listed: Profinet converter, barrier gas kit (plastic/stainless steel), gear chamber evacuation, spray-lance flushing device, temperature/pressure monitoring (inlet and exhaust), extension cable package up to 30 m, OmniControl, operating fluids, and electronic drive options (box version, switch-cabinet mounting plate, without drive electronics).
Section-by-Section Technical Breakdown
The HiLobe vacuum boosters are positioned as reliable, high-performance, efficient roots-type boosters. Key design points are intelligent condition monitoring via integrated sensors, hermetically sealed IE4 canned-motor design without shaft seals, variable speed drive, optimized rotor geometry and extended maintenance intervals. A list of optional accessories and drive configurations is given.
- High up-time due to intelligent condition monitoring via integrated sensors
- Hermetically sealed synchronous motor (IE4), canned motor design, no shaft seals
- Variable speed drive and optimized rotor geometries; extended maintenance intervals
- Options: Profinet converter; barrier gas kit (plastic or stainless steel); gear chamber evacuation; flushing device (spray lance); temperature and pressure monitoring (inlet and exhaust side)
- Extension cable package possible up to 30 m; OmniControl; operating fluids
- Electronic drive options: box version, mounting plate for switch cabinet integration, or without drive electronics
This section provides the dimensional drawing, pumping speed curves for 50/60 Hz, and a comparison table of the three models. Nominal pumping speeds range from 2700 to 13540 m³/h depending on model and speed, with differential pressure limits decreasing at higher speeds. All models share the same DN 250 flanges, water cooling, and RC 15000 electronic drive.
- Nominal pumping speed: 8103: 2700-8120 m³/h; 11003: 2700-10830 m³/h; 14003: 2700-13540 m³/h
- Rotational speed: 8103: 1500-4500 rpm; 11003: 1500-6000 rpm; 14003: 1500-7500 rpm
- Max. differential pressure at max. rotational speed: 25 hPa (8103), 20 hPa (11003), 10 hPa (14003); at min. rotational speed: 30 hPa all models
- Main and secondary flanges: DN 250 ISO-F | DN 250 PN 16 (all models)
- Input voltage: 380-480 V AC (±10%), 50/60 Hz; three-phase motor with PTC protection
- Rated power at full rotational speed: 9 kW (8103), 12 kW (11003), 15 kW (14003)
- Electronic drive unit: RC 15000; recommended mains fuse 50 A; interfaces PV.can and RS485
- Cooling: water cooled; ambient temperature 5-40 °C
This table lists operating environment limits, cooling water requirements, noise level, operating fluid, leak rate, dimensions and weight. All three models share identical values for these parameters.
- Shipping temperature: -10 to 40 °C; sound pressure level 70 dB(A) (EN ISO 2151) at 1 hPa inlet pressure
- Operating fluid: D2, filling volume 3 l
- Protection degree: IP54; type 12
- Cooling water: pressure 2000-6000 hPa, flow 120-360 l/h, temperature 5-35 °C
- Integral leak rate: 1E-06 Pa m³/s
- Dimensions (L x W x H): 1115 x 583 x 468 mm; weight approx. 620 kg (all models)
Key Integration, Operating & Safety Requirements
| Specification / Condition | Engineering Implication | Applies To |
|---|---|---|
| Max. differential pressure at maximum rotational speed is 25 hPa (HiLobe 8103), 20 hPa (11003) and 10 hPa (14003); at minimum rotational speed all models tolerate 30 hPa. p.2 ↗ | Pressure differential limits depend on speed: when staging the booster with a backing pump, the drive speed must be reduced or the backing pump sized so the differential across the booster stays within limits at high speeds. | whole series (values differ per model) |
| Rotational speed ranges: 8103: 1500-4500 rpm, 11003: 1500-6000 rpm, 14003: 1500-7500 rpm. p.2 ↗ | The higher-speed 14003 achieves the largest pumping speed but with the lowest allowed differential pressure; selection must trade speed headroom against differential tolerance. | whole series |
| All models use the RC 15000 electronic drive unit with a recommended 50 A mains fuse. p.2 ↗ | Specify RC 15000 drive and 50 A fuse in the supply-side electrical design regardless of model chosen. | whole series |
| Input voltage is 380-480 V AC ±10%, 50/60 Hz, three-phase motor with PTC motor protection. p.2 ↗ | Suitable for standard 400/480 V industrial networks; PTC sensors must be wired into the motor protection circuit. | whole series |
| Rated power at full rotational speed: 9 kW (8103), 12 kW (11003), 15 kW (14003). p.2 ↗ | Sizes the motor supply cable, breaker and drive electronics per model. | whole series |
| Cooling is water-based: pressure 2000-6000 hPa, flow 120-360 l/h, water temperature 5-35 °C. p.3 ↗ | A closed-loop or facility water circuit meeting these flow/pressure/temperature limits is required at installation. | whole series |
| Ambient temperature range is 5-40 °C (operation) and -10 to 40 °C for shipping. p.2 ↗ | Do not install or transport the pump outside these limits; freeze protection needed below 5 °C in operation. | whole series |
| Main and secondary flanges are DN 250 ISO-F or DN 250 PN 16 on all models. p.2 ↗ | Inlet/outlet connections are identical across models; pipework and gaskets can be standardized, with a choice of clamped ISO-F or PN16 flange mounting. | whole series |
| Integral leak rate is 1E-06 Pa m³/s. p.3 ↗ | Tightness is sufficient for high-vacuum applications; leak-up rate of the chamber can be budgeted with this value in mind. | whole series |
| Operating fluid is D2 with a filling volume of 3 l. p.3 ↗ | Stock Pfeiffer D2 fluid for oil changes; 3 l per pump for oil service planning. | whole series |
| Protection degree IP54, type 12. p.3 ↗ | Suitable for indoor industrial environments with dust and dripping-water protection; not for wash-down or outdoor use without additional enclosures. | whole series |
| Sound pressure level is 70 dB(A) (EN ISO 2151) at 1 hPa inlet pressure. p.3 ↗ | Noise is moderate and identical for all models; standard hearing-protection zoning likely sufficient. | whole series |
| Dimensions 1115 x 583 x 468 mm (L x W x H) and weight approx. 620 kg for all models. p.3 ↗ | Lifting gear rated for >620 kg and identical foundation/footprint can be used for any variant; models are physically interchangeable. | whole series |
| I/O interfaces are PV.can and RS485; a Profinet converter is available as an option. p.1 ↗ | Remote monitoring/condition monitoring integration via RS485 or PV.can; Profinet integration requires the optional converter. | whole series |
| Extension cable package allows drive-motor separation up to 30 m. p.1 ↗ | Pump can be located up to 30 m from drive electronics; longer distances need a different solution. | whole series |
| Drive electronics options: box version, mounting plate for switch cabinet integration, or without drive electronics. p.1 ↗ | Enables either decentralized installation next to the pump or central integration into an existing switch cabinet. | whole series |
| Canned-motor design is hermetically sealed with no shaft seals, IE4 synchronous motor. p.1 ↗ | Eliminates a common leak/wear path (shaft seals), reducing maintenance and improving containment in process applications. | whole series |
| Barrier gas kit (plastic or stainless steel), gear chamber evacuation and spray-lance flushing are available options. p.1 ↗ | For dusty/condensable processes, specify barrier gas, flushing or gear evacuation at ordering time since these are factory options. | whole series |