SpeedAir 3050 — product leaflet
Technical Reference Summary
Covered Models & Document Scope
Pfeiffer Vacuum SpeedAir 3050 air-based leak detector for liquid products; single model, no ordering codes or variants given; sensitivity down to 1.0 µm, pilot-operated valve air supply, configurable options noted only as 'depending on configuration'.
Section-by-Section Technical Breakdown
Positioning of the SpeedAir 3050 as a flexible leak detector for a wide range of applications and liquid products, using proven Non-Destructive Testing (NDT) techniques. Claimed to be the most sensitive air-based technology for liquid products compared to helium leak detectors, with repeatable, reliable results and fewer false negatives/positives.
- Flexible: suitable for a wide range of applications and products
- Proven: based on global NDT analysis techniques (no damage to materials/components)
- Accurate: most sensitive air-based technology for liquid products vs helium leak detectors; repeatable, reliable results, fewer false results
Key operating data: sensitivity, clean dry air (CDA) supply requirements, calibration and venting gas conditions, electrical supply, environmental limits, and software/user interface features. Includes 10" touch HMI with Leak Rx software compliant with 21 CFR Part 11.
- Sensitivity: down to 1.0 µm
- CDA supply: air (pilot operated valves), quality 1.3.1 per ISO 8573-1
- CDA pressure: 65–145 psig / 4.5–10 bar rel.; typical consumption 2 Nl/cycle
- Calibration gas: CDA air, 1–20 Torr (pressure listing as printed: 1.3–27 bar abs.)
- Venting gas: N₂, 2–10 psig (as printed: 140–690 bar rel. — likely misprint)
- Power supply: 90–250 V AC, typical consumption 1200 W
- Operating temperature 59–83 °F / 15–45 °C, humidity 30–80 %
- 10" multi-touch full HD color screen; Leak Rx software (21 CFR Part 11, PDF GMP test/calibration reports, local or LDAP domain authentication)
- Operating system: Windows 10
Physical and interface data of the unit: network and USB connectivity, noise level, weight, and overall dimensions. Note that some values depend on configuration.
- Network: 1x LAN (RJ45); interfaces: USB, wired Ethernet
- Sound pressure level: <53 dB(A)
- Weight: approx. 280 lbs / 127 kg
- Dimensions (L x W x H): 36 x 24 x 55 inch / 922 x 601 x 1397 mm
- Values marked * depend on configuration
Key Integration, Operating & Safety Requirements
| Specification / Condition | Engineering Implication | Applies To |
|---|---|---|
| Leak detection sensitivity down to 1.0 µm p.2 ↗ | Sets the minimum detectable defect size for test-part qualification; defects smaller than 1 µm will not be detected. | SpeedAir 3050 |
| Requires CDA per ISO 8573-1 quality class 1.3.1, 65–145 psig (4.5–10 bar rel.), pilot operated valves p.2 ↗ | Facility compressed air must meet particle/water/oil class 1.3.1 and a pressure regulator within 4.5–10 bar rel. is mandatory before installation. | SpeedAir 3050 |
| Typical CDA consumption is 2 Nl per cycle p.2 ↗ | Allows sizing of plant air supply and estimating cost per test cycle. | SpeedAir 3050 |
| Calibration gas supply is CDA air at 1–20 Torr p.2 ↗ | No external calibration gas bottles needed; supply line must be regulated to the specified Torr range. | SpeedAir 3050 |
| Venting gas is N₂ at 2–10 psig p.2 ↗ | A regulated nitrogen line must be plumbed to the unit; pressure shown as '140–690 bar rel.' in the document appears to be a misprint and should be verified with the manufacturer. | SpeedAir 3050 |
| Power supply 90–250 V AC with typical consumption 1200 W p.2 ↗ | Wide-range AC input simplifies global deployment; provision ≥1200 W circuit capacity per unit. | SpeedAir 3050 |
| Operating temperature 15–45 °C (59–83 °F) and humidity 30–80 % p.2 ↗ | Installation environment (production floor, testing cell) must stay within these limits; not suitable for unconditioned hot/cold areas. | SpeedAir 3050 |
| Weight approx. 127 kg (280 lbs) and dimensions 922 x 601 x 1397 mm p.3 ↗ | Requires appropriate handling equipment or casters/foundation planning and floor space of roughly 0.92 x 0.60 m footprint with 1.4 m height clearance. | SpeedAir 3050 |
| Sound pressure level below 53 dB(A) p.3 ↗ | Can be installed near operators without additional acoustic enclosure. | SpeedAir 3050 |
| Connectivity: 1x LAN (RJ45), USB, wired Ethernet; runs Windows 10 p.2 ↗ | Plan plant network integration and IT security (Windows 10 endpoint) for data exchange and remote reporting. | SpeedAir 3050 |
| Leak Rx software with 21 CFR Part 11 compliance, PDF GMP test and calibration reports, local or LDAP domain authentication p.2 ↗ | Suitable for GMP-regulated (pharma) production environments; user management can be tied to corporate LDAP/Active Directory. | SpeedAir 3050 |
| HMI is a 10" multi-touch full HD color screen p.2 ↗ | Operator interface is touch-based; plan enclosure/access layout accordingly, no external monitor required. | SpeedAir 3050 |