Induztr
Engineering Digest Document Viewer Related Models

ATP 2300 M — product leaflet

Technical Reference Summary

Covered Models & Document Scope

Pfeiffer Vacuum ATP 2300 M magnetically levitated (magnetic bearing) turbomolecular vacuum pump. Covers a single model with flange variants (various inlet flange types; dimensions depend on flange variant) and options/accessories: installation kit, cooling water valve, purge gas valve, handheld remote controller, rack-mounted controller, inlet protection screen, heater unit, valve cable and coil, various user interfaces, on-board control options (Profibus, Remote). Remote controller type: Magpower; recommended backing pumps: ADP/ADS.

Section-by-Section Technical Breakdown

pp. 1
Product Overview and Key Features

The ATP 2300 M is a turbomolecular vacuum pump with fully magnetic bearing technology, making it maintenance-free with no wearing parts. It is aimed at vibration-sensitive applications such as semiconductor manufacturing and analytical instrumentation due to low noise and vibration. It is compatible with a variety of backing pumps and offers excellent compression ratio for light gases.

  • Maintenance-free magnetic bearing technology — no wearing parts, long maintenance intervals
  • Low noise and vibration, suited to vibration-sensitive semiconductor and analytical applications
  • Flexible compatibility with various backing pumps; excellent light-gas compression
  • Accessories/options: installation kit, cooling water valve, purge gas valve, handheld remote controller, rack-mounted controller, inlet protection screen, heater unit, valve cable and coil, various inlet flange types and user interfaces
pp. 2-3
Performance Data (Pumping Speed, Compression, Pressures)

Detailed vacuum performance figures per gas: pumping speeds of roughly 1400–2050 l/s depending on gas, very high N₂/Ar compression ratios (>1·10⁸), ultimate pressure below 4·10⁻⁹ mbar, and maximum forevacuum/inlet vacuum values per gas. Gas throughput at final speed ranges from 900 to >3000 sccm.

  • Pumping speed: N₂ 1450–1950 l/s, Ar 1400–1900 l/s, He 1700–2050 l/s, H₂ 1600–1700 l/s
  • Compression ratio (ISO 5302): N₂ and Ar > 1·10⁸; He > 1·10⁵; H₂ > 7·10³
  • Ultimate pressure < 4·10⁻⁹ mbar
  • Max. forevacuum: N₂ 2.1 mbar, Ar 1.7 mbar, He 1.5 mbar, H₂ 1.9 mbar
  • Max. inlet vacuum: N₂ 0.04 mbar, Ar 0.011 mbar, He 0.2 mbar, H₂ 0.7 mbar
  • Gas throughput at final speed: N₂ 2200 sccm, Ar 900 sccm, He and H₂ > 3000 sccm
  • Recommended purge flow rate: 50 sccm
pp. 3
Rotational Speed, Power Supply and Controller

The pump runs at 31000 rpm nominal with variable speed from 15000 rpm up to nominal, run-up to 90% final speed in under 8 minutes. Controller is powered from single-phase 200–240 V 50/60 Hz; stand-by power is 130 W with up to 1000 W during start-up. Remote controller type is Magpower, with Profibus and Remote on-board control options.

  • Rotational speed 31000 rpm; variable from 15000 rpm to nominal speed
  • Run-up time to 90% final speed: < 8 min (with pressure exhaust < 0.1 mbar)
  • Controller power supply: 200–240 V, 50/60 Hz, single-phase
  • Max. power consumption for start-up: 1000 W; nominal (stand-by) power: 130 W
  • Remote controller: Magpower; on-board control options: Profibus, Remote
  • Recommended backing pump: ADP/ADS
pp. 3
Cooling, Baking, Connections and Installation Data

Water-cooled operation with recommended 60 l/h cooling water flow at 15–40 °C, and baking up to 120 °C is permitted. Mechanical data includes dimensions of Ø315 × 278 mm (flange-dependent), 52 kg pump weight, 8.7 kg controller weight, DN 200/DN 250 inlet and DN 40 ISO-KF outlet flanges, plus a stainless steel inlet screen fitted as factory default.

  • Max. baking temperature: 120 °C
  • Recommended cooling water flow rate: 60 l/h; cooling water temperature 15–40 °C
  • Gas inlet / outlet: DN 200, DN 250 / DN 40 ISO-KF; N₂ purge flange DN 16 ISO-KF
  • Dimensions Ø x H: 315 x 278 mm (height depending on flange variant)
  • Weight: pump ~52 kg, controller ~8.7 kg
  • Stainless steel inlet screen fitted as factory setting
pp. 1,3
Acoustics, Vibration and Environmental Limits

The pump achieves < 46 dB(A) sound pressure at 1 m (ISO 2151) at nominal speed and vibration below 0.01 μm, confirming suitability for vibration-sensitive processes. A maximum permissible radial magnetic field of < 12 mT at 10 cm distance is specified.

  • Sound pressure level (ISO 2151, 1 m, nominal speed): < 46 dB(A)
  • Vibration level at nominal speed: < 0.01 μm
  • Permissible radial magnetic field at 10 cm: < 12 mT

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication Applies To
Pumping speed at nominal speed: N₂ 1450–1950 l/s, Ar 1400–1900 l/s, He 1700–2050 l/s, H₂ 1600–1700 l/s. p.2300 ↗ Size backing pump and process throughput for gas-specific speeds; He/H₂ performance remains high, so light-gas applications do not derate significantly. ATP 2300 M
Compression ratio (ISO 5302): N₂ and Ar > 1·10⁸, He > 1·10⁵, H₂ > 7·10³. p.2300 ↗ Very high N₂/Ar compression allows simple, small backing pumps; for H₂ applications the backing pump's base pressure dominates the ultimate pressure. ATP 2300 M
Ultimate pressure < 4·10⁻⁹ mbar. p.2300 ↗ Suitable for UHV-class applications; system base pressure will be limited by seals/gaskets below this value. ATP 2300 M
Max. forevacuum: N₂ 2.1 mbar, Ar 1.7 mbar, He 1.5 mbar, H₂ 1.9 mbar. p.2300 ↗ Backing pump must maintain foreline pressure below these limits at the relevant gas load; switch-over/crossover settings should be chosen with margin. ATP 2300 M
Max. inlet vacuum (max. tolerable inlet pressure): N₂ 0.04 mbar, Ar 0.011 mbar, He 0.2 mbar, H₂ 0.7 mbar. p.2300 ↗ Do not admit process gas above these inlet pressures at full speed to avoid motor overloading; ramp gas loads gradually. ATP 2300 M
Gas throughput at final rotational speed: N₂ 2200 sccm, Ar 900 sccm, He > 3000 sccm, H₂ > 3000 sccm. p.2300 ↗ Defines maximum continuous gas load per gas during operation at full speed. ATP 2300 M
Recommended purge flow rate: 50 sccm via DN 16 ISO-KF purge flange. p.2300 ↗ Set purge gas (N₂) mass flow to 50 sccm to protect bearings from process gas/condensation; provision purge plumbing accordingly. ATP 2300 M
Rotational speed 31000 rpm, variable from 15000 rpm up to nominal. p.2300 ↗ Speed can be reduced to 15000 rpm for soft processes or reduced vibration/throughput control via the controller. ATP 2300 M
Run-up time to 90% of final rotational speed is < 8 min with exhaust pressure < 0.1 mbar. p.2300 ↗ Process start sequences should allow ~8 minutes before full-speed operation; crossover valve timing must match this. ATP 2300 M
Controller power supply: 200–240 V, single-phase, 50/60 Hz; start-up power up to 1000 W, nominal (stand-by) power 130 W. p.2300 ↗ Provide single-phase 200–240 V supply rated for ≥1000 W inrush/start-up; steady-state load is low at 130 W. ATP 2300 M (Magpower controller)
Max. baking temperature 120 °C. p.2300 ↗ Bakeout cycles must be limited to 120 °C (heater unit available as accessory); do not exceed for chamber bake with pump attached. ATP 2300 M
Recommended cooling water flow rate 60 l/h at 15–40 °C. p.2300 ↗ Water cooling circuit must deliver 60 l/h within 15–40 °C; optional cooling water valve accessory available for automatic control. ATP 2300 M
Magnetic bearing technology — no wearing parts, maintenance-free with long maintenance intervals. p.2300 ↗ No lubrication or bearing replacement schedule needed; simplifies maintenance planning versus oil-lubricated turbopumps. ATP 2300 M
Sound pressure level < 46 dB(A) at 1 m (ISO 2151) and vibration < 0.01 μm at nominal speed. p.2300 ↗ Can be mounted close to vibration-sensitive tools (metrology, lithography) without dedicated vibration isolation or acoustic enclosures. ATP 2300 M
Gas inlet DN 200 or DN 250, outlet DN 40 ISO-KF, N₂ purge flange DN 16 ISO-KF. p.2300 ↗ Match chamber/inlet flange variant to vessel (DN 200 vs DN 250, dimensions vary with flange); DN 40 ISO-KF foreline sized for recommended ADP/ADS backing pumps. ATP 2300 M
Weight ~52 kg (pump) and ~8.7 kg (controller); dimensions Ø315 x 278 mm (height flange-dependent). p.2300 ↗ Verify mounting frame/load capacity and clearance; DN 250 inlet variant increases height. ATP 2300 M
Permissible radial magnetic field < 12 mT at 10 cm. p.2300 ↗ Keep sources of strong magnetic fields (magnets, transformers) at least 10 cm away or shield them; field above 12 mT can disturb the magnetic bearing system. ATP 2300 M
On-board control options: Profibus and Remote; remote controller type Magpower (handheld and rack-mounted controller available). p.2300 ↗ Integrate into PLC/DCS via Profibus or use Magpower remote controller; select ordering option (user interface variant) at time of order. ATP 2300 M
Stainless steel inlet screen included as factory setting; heater unit and inlet protection screen available as accessories. p.2300 ↗ Particle protection is standard; order heater unit or alternative screens/flanges only if process requires them. ATP 2300 M