ATP 2300 M — product leaflet
Technical Reference Summary
Covered Models & Document Scope
Pfeiffer Vacuum ATP 2300 M magnetically levitated (magnetic bearing) turbomolecular vacuum pump. Covers a single model with flange variants (various inlet flange types; dimensions depend on flange variant) and options/accessories: installation kit, cooling water valve, purge gas valve, handheld remote controller, rack-mounted controller, inlet protection screen, heater unit, valve cable and coil, various user interfaces, on-board control options (Profibus, Remote). Remote controller type: Magpower; recommended backing pumps: ADP/ADS.
Section-by-Section Technical Breakdown
The ATP 2300 M is a turbomolecular vacuum pump with fully magnetic bearing technology, making it maintenance-free with no wearing parts. It is aimed at vibration-sensitive applications such as semiconductor manufacturing and analytical instrumentation due to low noise and vibration. It is compatible with a variety of backing pumps and offers excellent compression ratio for light gases.
- Maintenance-free magnetic bearing technology — no wearing parts, long maintenance intervals
- Low noise and vibration, suited to vibration-sensitive semiconductor and analytical applications
- Flexible compatibility with various backing pumps; excellent light-gas compression
- Accessories/options: installation kit, cooling water valve, purge gas valve, handheld remote controller, rack-mounted controller, inlet protection screen, heater unit, valve cable and coil, various inlet flange types and user interfaces
Detailed vacuum performance figures per gas: pumping speeds of roughly 1400–2050 l/s depending on gas, very high N₂/Ar compression ratios (>1·10⁸), ultimate pressure below 4·10⁻⁹ mbar, and maximum forevacuum/inlet vacuum values per gas. Gas throughput at final speed ranges from 900 to >3000 sccm.
- Pumping speed: N₂ 1450–1950 l/s, Ar 1400–1900 l/s, He 1700–2050 l/s, H₂ 1600–1700 l/s
- Compression ratio (ISO 5302): N₂ and Ar > 1·10⁸; He > 1·10⁵; H₂ > 7·10³
- Ultimate pressure < 4·10⁻⁹ mbar
- Max. forevacuum: N₂ 2.1 mbar, Ar 1.7 mbar, He 1.5 mbar, H₂ 1.9 mbar
- Max. inlet vacuum: N₂ 0.04 mbar, Ar 0.011 mbar, He 0.2 mbar, H₂ 0.7 mbar
- Gas throughput at final speed: N₂ 2200 sccm, Ar 900 sccm, He and H₂ > 3000 sccm
- Recommended purge flow rate: 50 sccm
The pump runs at 31000 rpm nominal with variable speed from 15000 rpm up to nominal, run-up to 90% final speed in under 8 minutes. Controller is powered from single-phase 200–240 V 50/60 Hz; stand-by power is 130 W with up to 1000 W during start-up. Remote controller type is Magpower, with Profibus and Remote on-board control options.
- Rotational speed 31000 rpm; variable from 15000 rpm to nominal speed
- Run-up time to 90% final speed: < 8 min (with pressure exhaust < 0.1 mbar)
- Controller power supply: 200–240 V, 50/60 Hz, single-phase
- Max. power consumption for start-up: 1000 W; nominal (stand-by) power: 130 W
- Remote controller: Magpower; on-board control options: Profibus, Remote
- Recommended backing pump: ADP/ADS
Water-cooled operation with recommended 60 l/h cooling water flow at 15–40 °C, and baking up to 120 °C is permitted. Mechanical data includes dimensions of Ø315 × 278 mm (flange-dependent), 52 kg pump weight, 8.7 kg controller weight, DN 200/DN 250 inlet and DN 40 ISO-KF outlet flanges, plus a stainless steel inlet screen fitted as factory default.
- Max. baking temperature: 120 °C
- Recommended cooling water flow rate: 60 l/h; cooling water temperature 15–40 °C
- Gas inlet / outlet: DN 200, DN 250 / DN 40 ISO-KF; N₂ purge flange DN 16 ISO-KF
- Dimensions Ø x H: 315 x 278 mm (height depending on flange variant)
- Weight: pump ~52 kg, controller ~8.7 kg
- Stainless steel inlet screen fitted as factory setting
The pump achieves < 46 dB(A) sound pressure at 1 m (ISO 2151) at nominal speed and vibration below 0.01 μm, confirming suitability for vibration-sensitive processes. A maximum permissible radial magnetic field of < 12 mT at 10 cm distance is specified.
- Sound pressure level (ISO 2151, 1 m, nominal speed): < 46 dB(A)
- Vibration level at nominal speed: < 0.01 μm
- Permissible radial magnetic field at 10 cm: < 12 mT
Key Integration, Operating & Safety Requirements
| Specification / Condition | Engineering Implication | Applies To |
|---|---|---|
| Pumping speed at nominal speed: N₂ 1450–1950 l/s, Ar 1400–1900 l/s, He 1700–2050 l/s, H₂ 1600–1700 l/s. p.2300 ↗ | Size backing pump and process throughput for gas-specific speeds; He/H₂ performance remains high, so light-gas applications do not derate significantly. | ATP 2300 M |
| Compression ratio (ISO 5302): N₂ and Ar > 1·10⁸, He > 1·10⁵, H₂ > 7·10³. p.2300 ↗ | Very high N₂/Ar compression allows simple, small backing pumps; for H₂ applications the backing pump's base pressure dominates the ultimate pressure. | ATP 2300 M |
| Ultimate pressure < 4·10⁻⁹ mbar. p.2300 ↗ | Suitable for UHV-class applications; system base pressure will be limited by seals/gaskets below this value. | ATP 2300 M |
| Max. forevacuum: N₂ 2.1 mbar, Ar 1.7 mbar, He 1.5 mbar, H₂ 1.9 mbar. p.2300 ↗ | Backing pump must maintain foreline pressure below these limits at the relevant gas load; switch-over/crossover settings should be chosen with margin. | ATP 2300 M |
| Max. inlet vacuum (max. tolerable inlet pressure): N₂ 0.04 mbar, Ar 0.011 mbar, He 0.2 mbar, H₂ 0.7 mbar. p.2300 ↗ | Do not admit process gas above these inlet pressures at full speed to avoid motor overloading; ramp gas loads gradually. | ATP 2300 M |
| Gas throughput at final rotational speed: N₂ 2200 sccm, Ar 900 sccm, He > 3000 sccm, H₂ > 3000 sccm. p.2300 ↗ | Defines maximum continuous gas load per gas during operation at full speed. | ATP 2300 M |
| Recommended purge flow rate: 50 sccm via DN 16 ISO-KF purge flange. p.2300 ↗ | Set purge gas (N₂) mass flow to 50 sccm to protect bearings from process gas/condensation; provision purge plumbing accordingly. | ATP 2300 M |
| Rotational speed 31000 rpm, variable from 15000 rpm up to nominal. p.2300 ↗ | Speed can be reduced to 15000 rpm for soft processes or reduced vibration/throughput control via the controller. | ATP 2300 M |
| Run-up time to 90% of final rotational speed is < 8 min with exhaust pressure < 0.1 mbar. p.2300 ↗ | Process start sequences should allow ~8 minutes before full-speed operation; crossover valve timing must match this. | ATP 2300 M |
| Controller power supply: 200–240 V, single-phase, 50/60 Hz; start-up power up to 1000 W, nominal (stand-by) power 130 W. p.2300 ↗ | Provide single-phase 200–240 V supply rated for ≥1000 W inrush/start-up; steady-state load is low at 130 W. | ATP 2300 M (Magpower controller) |
| Max. baking temperature 120 °C. p.2300 ↗ | Bakeout cycles must be limited to 120 °C (heater unit available as accessory); do not exceed for chamber bake with pump attached. | ATP 2300 M |
| Recommended cooling water flow rate 60 l/h at 15–40 °C. p.2300 ↗ | Water cooling circuit must deliver 60 l/h within 15–40 °C; optional cooling water valve accessory available for automatic control. | ATP 2300 M |
| Magnetic bearing technology — no wearing parts, maintenance-free with long maintenance intervals. p.2300 ↗ | No lubrication or bearing replacement schedule needed; simplifies maintenance planning versus oil-lubricated turbopumps. | ATP 2300 M |
| Sound pressure level < 46 dB(A) at 1 m (ISO 2151) and vibration < 0.01 μm at nominal speed. p.2300 ↗ | Can be mounted close to vibration-sensitive tools (metrology, lithography) without dedicated vibration isolation or acoustic enclosures. | ATP 2300 M |
| Gas inlet DN 200 or DN 250, outlet DN 40 ISO-KF, N₂ purge flange DN 16 ISO-KF. p.2300 ↗ | Match chamber/inlet flange variant to vessel (DN 200 vs DN 250, dimensions vary with flange); DN 40 ISO-KF foreline sized for recommended ADP/ADS backing pumps. | ATP 2300 M |
| Weight ~52 kg (pump) and ~8.7 kg (controller); dimensions Ø315 x 278 mm (height flange-dependent). p.2300 ↗ | Verify mounting frame/load capacity and clearance; DN 250 inlet variant increases height. | ATP 2300 M |
| Permissible radial magnetic field < 12 mT at 10 cm. p.2300 ↗ | Keep sources of strong magnetic fields (magnets, transformers) at least 10 cm away or shield them; field above 12 mT can disturb the magnetic bearing system. | ATP 2300 M |
| On-board control options: Profibus and Remote; remote controller type Magpower (handheld and rack-mounted controller available). p.2300 ↗ | Integrate into PLC/DCS via Profibus or use Magpower remote controller; select ordering option (user interface variant) at time of order. | ATP 2300 M |
| Stainless steel inlet screen included as factory setting; heater unit and inlet protection screen available as accessories. p.2300 ↗ | Particle protection is standard; order heater unit or alternative screens/flanges only if process requires them. | ATP 2300 M |