Induztr
Engineering Digest Document Viewer Related Models

E2 — product leaflet

Technical Reference Summary

Covered Models & Document Scope

Pfeiffer Vacuum E2 and VE2 leak detectors (micro-flow leak testing). Two variants: E2 (pressure range 1–11 bar abs.) and VE2 (vacuum, down to 138 mbar abs.). Accessories/options include LeakTek software, Adaptive Test tools, Leak Rx, expansion tanks, pressure regulators, calibration orifices (ECD, SE, ECS), remote exhaust valve, oil-free vacuum packages, remote pendant, mass extraction UUT filter, quick connectors.

Section-by-Section Technical Breakdown

pp. 1
Features

Highlights the Micro-Flow technology with patented temperature, pressure and flow sensors, delivering short cycle times and high sensitivity and repeatability. The units are described as flexible for a wide range of applications with optional stainless steel housing and LAN remote interface.

  • Patented Micro-Flow technology with patented temperature, pressure and flow sensors
  • Short cycle times, best-in-class sensitivity, high repeatability
  • Optional stainless steel housing
  • Remote interface through LAN
pp. 1
Accessories, Spare Parts and Options

Lists available accessories, options and spare parts, including software tools, calibration orifices, vacuum generation packages and pneumatic accessories.

  • LeakTek software and Adaptive Test utility tools; Leak Rx
  • Expansion tank small/medium/large; precision pressure regulator small/medium volume
  • Verification orifice ECD, SE (ECD-X-PSE·X-P), ECS air calibration
  • Glass orifice calibrated at barometric inlet, vacuum outlet on primary standard
  • Remote exhaust valve with filter
  • Oil-free vacuum generation and control packages (ca. 0.3–0.8 bar abs. and ca. 0.07–0.15 bar abs.); rough vacuum package
  • Remote pendant with on/off button
  • Mass extraction UUT filter; quick connector with plug
pp. 2
Technical Specifications (E2 vs VE2)

Comparison table of the E2 and VE2 models covering sensitivity, pressure range, power, environment, interfaces, I/O, weight and dimensions.

  • Sensitivity: E2 0.01 cc/min and higher; VE2 1·10⁻⁴ sccs
  • Pressure range: E2 14.5–165 psia (1–11 bar abs.); VE2 vacuum to 138 mbar abs. up to barometric
  • Power supply: 110/120 / 220/240 V AC (50/60 Hz)
  • Operating temperature 15–45 °C; humidity 30–80 %
  • User interface: RJ45 Ethernet, RS232; software LeakTek with local or LDAP domain authentication; Windows 10 remote laptop OS
  • Interfaces: USB, wired Ethernet; 1x LAN (RJ45)
  • Weight ~10 kg; dimensions 305 x 305 x 305 mm
  • Pneumatic connections: 1/4" Swagelok test ports, 3/8" Swagelok vacuum
  • Digital I/O: opto-isolated 5–30 V DC (inputs: start, stop, test type, remote pressure switch; outputs: pass, fail, test type, test on, clamp, custom)
  • Analog output: single channel 0–5 V DC pressure control

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication Applies To
E2 sensitivity is 0.01 cc/min and higher p.2 ↗ Use E2 for fine leak testing where leak rates down to 0.01 cc/min must be detected E2
VE2 sensitivity is 1·10⁻⁴ sccs p.2 ↗ VE2 suitable for ultra-fine vacuum leak testing applications VE2
E2 pressure range 1–11 bar abs. (14.5–165 psia) p.2 ↗ Ensures test part and fixtures are rated for up to 11 bar absolute overpressure E2
VE2 pressure range from vacuum 138 mbar abs. to barometric pressure p.2 ↗ VE2 is the choice when testing requires vacuum conditions down to ~138 mbar absolute VE2
Power supply 110/120 or 220/240 V AC, 50/60 Hz p.2 ↗ Works on standard industrial mains worldwide without transformers; verify supply voltage variant at installation whole series
Operating temperature 15–45 °C, humidity 30–80 % p.2 ↗ Installation environment must be conditioned within these limits for specified accuracy/repeatability whole series
Digital I/O is opto-isolated, 5–30 V DC p.2 ↗ PLC handshake signals must stay within 5–30 V DC; inputs cover start/stop/test type/remote pressure switch, outputs pass/fail/test type/test on/clamp/custom whole series
Analog output is single channel 0–5 V DC for pressure control p.2 ↗ Downstream pressure monitoring/control hardware must accept 0–5 V DC signal range whole series
Pneumatic connections: 1/4" Swagelok for test ports, 3/8" Swagelok for vacuum p.2 ↗ Plumbing/tubing kits must be specified with matching Swagelok fitting sizes whole series
Weight approx. 10 kg, dimensions 305 x 305 x 305 mm p.2 ↗ Compact footprint allows bench or integration mounting with standard handling whole series
Network connection 1x LAN (RJ45); also USB and RS232 serial interface p.2 ↗ Remote operation/integration via Ethernet; RS232 and USB available for legacy tooling whole series
Software is LeakTek with local or LDAP domain authentication on Windows 10 remote laptop p.2 ↗ IT integration possible via domain authentication; requires Windows 10 host laptop whole series
Optional oil-free vacuum generation and control packages at ca. 0.3–0.8 bar abs. and ca. 0.07–0.15 bar abs., plus rough vacuum package p.1 ↗ Select package matching required VE2 vacuum level; oil-free design prevents contamination of test parts option for VE2 use
Calibration options include verification orifice ECD/SE (ECD-X-PSE·X-P), ECS air calibration, and glass orifice calibrated at barometric inlet/vacuum outlet on primary standard p.1 ↗ Plan periodic calibration/verification using these reference orifices for traceable leak-rate checks whole series
Optional stainless steel housing and LAN remote interface p.1 ↗ Stainless housing option for harsh/cleanroom environments; LAN enables remote monitoring whole series (options)

Related models