Induztr
Engineering Digest Document Viewer Related Models

SplitFlow 80 Neo — product leaflet

Technical Reference Summary

Covered Models & Document Scope

Pfeiffer Vacuum SplitFlow 80 Neo turbomolecular vacuum pump datasheet/brochure. Two high-vacuum flange variants: DN 63 CF-F and DN 63 ISO-K. Accessories/options listed: electrical adapter, mains cable, connecting cable, CIP adapter, power supply kit and display unit, pressure sensor, mounting kit, safety valve, venting accessories, power failure venting unit, vibration damper.

Section-by-Section Technical Breakdown

pp. 1
Product Features

Highlights high performance via an extended thermal operating range giving maximum compression and high gas throughput. Compact design and any-orientation mounting make it suitable for portable analytical applications and easy integration into small instruments.

  • Maximum compression and high gas throughput due to extended thermal operating range
  • Mounting in any orientation
  • Compact footprint for integration into small analytical instruments and vacuum systems
  • Suitable for portable analytical applications
pp. 1
Accessories, Spare Parts and Options

Lists available accessories and options for the SplitFlow 80 Neo, covering electrical connectivity, integration hardware and vacuum-side components.

  • Electrical adapter, mains cable, connecting cable
  • CIP adapter
  • Power supply kit and display unit
  • Pressure sensor
  • Mounting kit
  • Safety valve
  • Venting accessories and power failure venting unit
  • Vibration damper
pp. 2-3
Performance Data

Technical specifications: pumping speeds and compression ratios for N₂, H₂, He, Ar, interstage pumping, gas throughputs, fore-vacuum limits, run-up time, noise level, weight and flange options.

  • Max pumping speed: N₂ 67 l/s, H₂ 48 l/s, He 58 l/s, Ar 66 l/s
  • Interstage pumping speed: 2.8 l/s (He and N₂)
  • Compression ratio: N₂ 1·10¹¹, H₂ 1.4·10⁵, He 1.3·10⁷, Ar 1·10¹¹
  • Fore-vacuum/interstage pump compression ratio: 1·10⁵–5·10⁵ (N₂ and He)
  • Gas throughput: N₂ 1.3, H₂ 15.3, He 2.7, Ar 0.54 hPa·l/s
  • Fore-vacuum max: 22 hPa (N₂, Ar), 14 hPa (H₂)
  • Run-up time 75 s; noise 48 dB(A) per EN ISO 2151
  • Weight approx. 1.7–3.4 kg depending on HV flange size
  • HV flange: DN 63 CF-F or DN 63 ISO-K

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication Applies To
Maximum pumping speed for N₂ is 67 l/s (H₂ 48 l/s, He 58 l/s, Ar 66 l/s) p.2 ↗ Size the backing pump and conductance accordingly for main gas loads; N₂-base processes get the highest effective speed SplitFlow 80 Neo (both flange variants)
Interstage pumping speed is 2.8 l/s for both He and N₂ p.2 ↗ Interstage port can be used for differential/tandem applications with limited throughput (~2.8 l/s) SplitFlow 80 Neo
Compression ratio for N₂ is 1·10¹¹ and for H₂ only 1.4·10⁵ p.2 ↗ Ultimate pressure for hydrogen-bearing loads depends strongly on fore-vacuum quality; hydrogen requires a good backing pump SplitFlow 80 Neo
Fore-vacuum/interstage compression ratio is 1·10⁵–5·10⁵ (N₂ and He) p.2 ↗ Tandem operation pressure window; select backing stage within this range SplitFlow 80 Neo
Gas throughput N₂ 1.3 hPa·l/s; H₂ 15.3 hPa·l/s; He 2.7 hPa·l/s; Ar 0.54 hPa·l/s p.2 ↗ Limits continuous gas load sizing; hydrogen throughput is an order of magnitude above nitrogen SplitFlow 80 Neo
Maximum fore-vacuum pressure: 22 hPa for N₂ and Ar, 14 hPa for H₂ p.3 ↗ Backing pump must maintain ≤22 hPa (≤14 hPa for hydrogen) to avoid overload; tighter limit for H₂ SplitFlow 80 Neo
Run-up time is 75 s p.3 ↗ Plan process start sequences and interlock delays around ~75 s to full speed SplitFlow 80 Neo
Sound pressure level is 48 dB(A) per EN ISO 2151 p.3 ↗ Low-noise operation suitable for lab/portable analytical environments SplitFlow 80 Neo
Weight approx. 1.7–3.4 kg depending on HV flange size p.3 ↗ Light enough for portable/mobile mounting; account for flange choice in mass budgets SplitFlow 80 Neo
HV flange options: DN 63 CF-F and DN 63 ISO-K p.3 ↗ CF-F variant for UHV/high-leak-tightness requirements; ISO-K for standard vacuum connections SplitFlow 80 Neo variants
Pump can be mounted in any orientation p.1 ↗ Enables compact, unconventional integration in small analytical instruments without orientation constraints SplitFlow 80 Neo
A power failure venting unit is available as accessory p.1 ↗ Protects the process/chamber on power loss; include it for unattended operation SplitFlow 80 Neo