ATH 4506 M — product leaflet

Technical Reference Summary

Covered Models & Document Scope

Pfeiffer Vacuum ATH 4506 M and ATH 4506 MT turbomolecular vacuum pumps (high-temperature/high-flow variants with wide process range and corrosion resistance). Document is a 3-page datasheet/brochure covering specifications, dimensions, pumping speed curves, and available accessories/options.

Section-by-Section Technical Breakdown

pp. 1
Features, Accessories and Options

Highlights the ATH 4506 M/MT positioning as a high temperature and high flow turbomolecular pump with wide process range and high corrosion resistance. Built-in sensor provides performance diagnostics, maintenance optimization and external communications, with self-diagnostics and optimized temperature management for low cost of ownership.

  • Self-diagnostics and built-in sensor for performance diagnostics, maintenance optimization and external communication
  • Economical operation via optimized temperature management, low power and cooling water consumption
  • Accessories/options: protective seal gas purge valve, cooling water valve, handheld remote control, protective inlet filter, mounting kit
  • Inlet flange options: VG350 or VG400; exhaust flanges DN40 or DN50
  • Heating options, additional middle purge for harsh processes, rotor temperature sensor, purge presence sensor
  • Input power options: 200, 440 or 480 VAC
pp. 2
Pumping Speed and Vacuum Performance Data

Provides pumping speeds for N₂, Ar, He and H₂, ultimate pressure, and maximum fore/inlet vacuum limits, identical for M and MT variants. Gas throughput at final rotational speed differs slightly between the two variants, and the MT specifies a fixed recommended purge flow while the M allows 0–100 sccm.

  • Pumping speed N₂ 4500 l/s; Ar 4370 l/s; He 3900 l/s; H₂ 3700 l/s (both variants)
  • Ultimate pressure < 1·10⁻⁸ mbar
  • Max. forevacuum: 3.5 mbar (N₂ and Ar); max. inlet vacuum: 0.5 mbar (N₂), 0.22 mbar (Ar)
  • Gas throughput at final rotational speed N₂: 5500 sccm (MT) / 6550 sccm (M); Ar: 2800 sccm (MT) / 3500 sccm (M)
  • Recommended purge flow: 100 sccm (MT) vs 0–100 sccm (M)
  • Rotational speed 23500 rpm, variable 15000 rpm to nominal speed
  • Run-up time to 90% of final speed with exhaust pressure < 0.1 mbar: < 9 min
pp. 3
Electrical, Cooling and Mechanical Specifications

Electrical power data, heating and cooling water requirements, control interfaces and mechanical dimensions/weight. Both variants share identical electrical, cooling, mounting and interface data; heating temperature applies only to the MT.

  • Max. power consumption for start-up: 2000 W; nominal (stand-by) power: 220 W; max. current consumption: 14 A
  • Controller power supply: 230/440/480 V AC
  • Max. heating temperature 50–95 °C (MT only)
  • Cooling water: 60–180 l/h recommended flow, 15–35 °C water temperature
  • Pump-mounted controller; on-board control options: Remote, EtherCAT
  • Dimensions Ø x H: 450 x 422 mm; weight approx. 140 kg
  • Gas inlet: VG350, VG400, VF350; gas outlet: DN 40 ISO-K or DN 50 ISO

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication Applies To
Pumping speed is 4500 l/s for N₂, 4370 l/s for Ar, 3900 l/s for He and 3700 l/s for H₂, identical for ATH 4506 M and MT. p.2 ↗ Size backing pumps and process throughput calculations on gas-specific speeds; He/H₂ speeds are ~82% of the N₂ rating. ATH 4506 M and MT
Ultimate pressure < 1·10⁻⁸ mbar. p.2 ↗ Suitable for high/ultra-high vacuum process and analytical applications; system base pressure will be limited by chamber outgassing and backing pump. Both variants
Maximum forevacuum (backing pressure) is 3.5 mbar for N₂ and Ar. p.2 ↗ Backing pump must maintain ≤3.5 mbar at full gas throughput; oversizing the backing line is unnecessary but undersizing will cause rotor instability/shutdown. Both variants
Maximum inlet vacuum (inlet pressure) 0.5 mbar (N₂) / 0.22 mbar (Ar). p.2 ↗ Defines the upper inlet pressure for safe continuous operation at full speed when pumping through gases; stays within process windows during gas flooding. Both variants
Gas throughput at final rotational speed: N₂ 5500 sccm (MT) / 6550 sccm (M); Ar 2800 sccm (MT) / 3500 sccm (M). p.2 ↗ The M variant tolerates higher continuous gas load; select M for high gas-flow processes, MT for high-temperature applications. ATH 4506 M vs MT
Recommended purge flow rate is 100 sccm fixed for MT, adjustable 0–100 sccm for M. p.2 ↗ Process purge/lean gas shielding must be supplied at these rates; MT requires continuous purge, M allows purge-free operation. ATH 4506 MT (fixed purge), ATH 4506 M (0–100 sccm)
Rotational speed 23500 rpm nominal, variable down to 15000 rpm. p.2 ↗ Speed can be reduced to ~64% for throughput/temperature control without changing hardware; plan control setpoints accordingly. Both variants
Run-up time to 90% of final rotational speed is < 9 min with exhaust pressure < 0.1 mbar. p.2 ↗ Sequence pump start with backing pressure < 0.1 mbar to achieve the quoted <9 min start time; important for cycle-time planning. Both variants
Controller power supply 230/440/480 V AC; input power options 200, 440 or 480 VAC. p.1 ↗ Verify site supply matches one of these voltages before ordering; no 400 V-only version required to be handled specially. Both variants
Max. start-up power consumption 2000 W, nominal/stand-by power 220 W, max. current 14 A. p.3 ↗ Size supply cable and breaker for 2000 W / 14 A inrush at start-up; steady-state load is only 220 W. Both variants
Maximum heating temperature 50–95 °C, applicable to MT only. p.3 ↗ MT supports heated operation up to 95 °C for condensable gas handling; the M variant has no heating option in this datasheet. ATH 4506 MT
Cooling water flow 60–180 l/h at 15–35 °C. p.3 ↗ Provide a water circuit within this flow/temperature window; insufficient flow or >35 °C water degrades temperature management and cost-of-ownership claims. Both variants
Dimensions Ø450 x 422 mm (height), weight approx. 140 kg. p.3 ↗ Plan mounting structure and handling equipment for 140 kg; inlet flange face is a 450 mm diameter envelope. Both variants
Gas inlet options VG350, VG400, VF350; gas outlet DN 40 ISO-K or DN 50 ISO. p.1 ↗ Match chamber flange (VG350/VG400/VF350) and backing line (DN40 ISO-K or DN50 ISO) at ordering time; flange choice is an ordering option, not a retrofit. Both variants
On-board control options: Remote and EtherCAT; remote controller is pump-mounted. p.3 ↗ For automated tools, integrate via EtherCAT; handheld remote control is available as an accessory for manual operation. Both variants
Available options include protective seal gas purge valve, cooling water valve, rotor temperature sensor, purge presence sensor, additional middle purge for harsh processes, and protective inlet filter. p.1 ↗ For corrosive/harsh processes, order the middle purge and sensors; the purge presence sensor enables interlocking pump start on confirmed purge flow. Both variants (options)