Calibration system Basic — product leaflet
Technical Reference Summary
Covered Models & Document Scope
Pfeiffer Vacuum 'Calibration system Basic' – a standardized, manually operated vacuum calibration system compliant with ISO/TS 3567. Variants/options include reference gauges (CMR or CCT type, DakkS-certified), OmniControl, TPG 366 indicator/controller for 6 gauges, various connection flanges, and customer-specific flange/pump configurations or a larger recipient on request.
Section-by-Section Technical Breakdown
The Calibration system Basic is a standardized calibration system designed for user-friendly manual operation with an integrated control unit. It complies with the requirements of ISO/TS 3567 and can be adapted via individual flange connections and DakkS-certified reference gauges.
- Standardized, user-friendly calibration system
- Operated by hand and via integrated control unit (manual actuator)
- Compliant with ISO/TS 3567
- Individual flange connections to fit customer requirements
- DakkS-certified reference gauges available
A wide range of accessories and options is available for the calibration system, including reference gauges and control/indicator units. Individual configurations can be requested from the manufacturer.
- Wide range of spare parts
- Reference gauges (CMR or CCT type)
- OmniControl
- TPG 366 indicator and controller for 6 gauges
- Various connection flanges
- Individual flange and/or pump configurations on request
- Larger recipient on request
The system achieves an ultimate pressure below 1·10⁻⁷ hPa and covers a measurement range from 1·10⁻⁴ to 1·10³ hPa. It is a manually actuated stainless steel unit weighing approx. 22.1 kg with a defined set of ISO-KF connection flanges and 115/230 V AC supply.
- Ultimate pressure: < 1·10⁻⁷ hPa (mbar)
- Measurement range: 1·10⁻⁴ – 1·10³ hPa (mbar)
- Pumping speed backing pump at 50 Hz: 0.9 m³/h
- Pumping speed for N₂: 67 l/s
- Input voltages: 115/230 V AC (50/60 Hz)
- Power consumption: 230 VA
- Actuator / mode of operation: Manual
- Cable length: 3 m
- Material: Stainless steel, glass bead blasted
- Precision: depending on used reference gauges
- Weight: approx. 22.1 kg
- Connecting flange (in): DN 16 ISO-KF
- Connection for gauge: 2x DN 40 ISO-KF, 3x DN 16 ISO-KF, 2x DN 25 ISO-KF / 6x DN 40 ISO-KF
- Gas inlet: DN 16 ISO-KF
- Dimensional drawing provided on page 2
Key Integration, Operating & Safety Requirements
| Specification / Condition | Engineering Implication | Applies To |
|---|---|---|
| Ultimate pressure of the calibration system is < 1·10⁻⁷ hPa (mbar). p.2 ↗ | Sufficient vacuum base pressure for calibrating gauges down to the 1·10⁻⁴ hPa lower measurement limit. | Calibration system Basic (whole system) |
| Measurement range is 1·10⁻⁴ to 1·10³ hPa (mbar). p.2 ↗ | Defines the pressure window in which gauges can be calibrated; reference gauges must cover this range. | Calibration system Basic |
| Pumping speed for N₂ is 67 l/s. p.2 ↗ | Sets pumpdown time constants when evacuating recipients; relevant for cycle-time planning. | Calibration system Basic |
| Backing pump pumping speed at 50 Hz is 0.9 m³/h. p.2 ↗ | Fore-vacuum capacity is fixed; verify suitability if larger recipients are requested. | Calibration system Basic |
| Input voltages are 115/230 V AC at 50/60 Hz with 230 VA power consumption. p.2 ↗ | Operable on both US and European mains supplies without transformers; dimension supply circuit for 230 VA. | Calibration system Basic |
| Actuator and mode of operation are manual, with a 3 m cable length. p.2 ↗ | No automated actuation in the Basic version; consider OmniControl option if automation is required. | Calibration system Basic |
| Housing material is stainless steel, glass bead blasted, weight approx. 22.1 kg. p.2 ↗ | Hydrocarbon-free clean vacuum-compatible construction; ~22 kg handling/mounting load must be supported. | Calibration system Basic |
| Gauge connections are 2x DN 40 ISO-KF, 3x DN 16 ISO-KF, 2x DN 25 ISO-KF, or alternatively 6x DN 40 ISO-KF. p.2 ↗ | Up to six gauges can be mounted depending on flange configuration, matching the TPG 366 6-gauge controller option. | Calibration system Basic (variant-dependent) |
| Gas inlet and inlet connecting flange are both DN 16 ISO-KF. p.2 ↗ | Gas admission hardware must use small-KF (DN 16 ISO-KF) fittings; plan leak/valve interfaces accordingly. | Calibration system Basic |
| Calibration precision depends on the reference gauges used (CMR or CCT type, DakkS-certified). p.1 ↗ | System accuracy is not intrinsic; select DakkS-certified reference gauges to meet metrological requirements. | Calibration system Basic with CMR/CCT reference gauges |
| System complies with ISO/TS 3567 requirements. p.1 ↗ | Vacuum gauge calibrations performed on this system can be claimed as ISO/TS 3567-conformant. | Calibration system Basic |
| Individual flange and/or pump configurations and a larger recipient are available on request. p.1 ↗ | Non-standard gauge counts or volumes can be ordered as custom configurations rather than modifying the standard unit. | Calibration system Basic (options) |