Induztr
Engineering Digest Document Viewer Related Models

HiQuad QMA 1-300 — product leaflet

Technical Reference Summary

Covered Models & Document Scope

Pfeiffer Vacuum HiQuad Neo quadrupole mass spectrometer family, covering two analyzer variants: QMA 430 (mass range 1–300 u) and QMA 400 (mass range 1–512 u). Includes accessories/spares list (electronics, ion source, filament, wiring, multiplier, analyzer, cable, vacuum chamber, upgrade kit from previous generation).

Section-by-Section Technical Breakdown

pp. 1
Product Highlights

The HiQuad Neo is presented as a flexible, modular quadrupole mass spectrometer that is backwards-compatible with earlier generation hardware. Operation is via Pfeiffer MassSpec software, and system integration is enabled through an OPC-UA interface. High measuring speed and maximum sensitivity are highlighted as key performance traits.

  • Modular, backwards-compatible design
  • Operation via Pfeiffer MassSpec software
  • System integration via OPC-UA interface
  • Fast measuring speed, maximum sensitivity
pp. 1
Accessories, Spare Parts and Options

The document lists the available spare parts and options for the HiQuad Neo, including spare electronics, ion source, filament, multiplier, analyzer and cables. A vacuum chamber and an upgrade kit for converting old-generation units to the new generation are also offered.

  • Spare electronics
  • Spare ion source
  • Spare filament
  • Spare wiring
  • Spare multiplier
  • Spare analyzer
  • Spare cable
  • Vacuum chamber
  • Upgrade kit from old to new generation
pp. 2
Technical Data / Dimensional Drawing

A side-by-side technical data table with dimensional drawing compares the QMA 430 and QMA 400 analyzers. Key parameters include mass range, detection limit, Ar sensitivity, partial pressure ratio, measurement speed, rod system material, HF generator, preamplifier, temperature limits, dimensions and flange connection.

  • QMA 430: mass range 1–300 u; QMA 400: 1–512 u
  • Detection limit: <2·10⁻¹⁵ hPa (mbar) QMA 430; <1·10⁻¹⁵ hPa (mbar) QMA 400
  • Ar sensitivity: 2·10⁻⁴ A/hPa (QMA 430); 5·10⁻⁴ A/hPa (QMA 400)
  • Max. operating pressure: 1·10⁻⁴ / 1·10⁻⁵ hPa (both)
  • Partial pressure ratio with SEV: <1 ppb (QMA 430); <0.5 ppb (QMA 400)
  • Min. measurement speed: 125 µs/u – 60 s/u (both)
  • Rod system: stainless steel, 8 mm diameter (QMA 430); Mo, 8 mm diameter (QMA 400)
  • HF generator: QMH 800-3 (QMA 430); QMH 800-5 (QMA 400)
  • Electrometer preamplifier: EP 822 (both)
  • Analyzer operating temperature: 150 °C; bake-out temperature: 400 °C (both)
  • Dimensions (L×W): 309 × (432–454) mm (both)
  • Connecting flange: DN 63 CF-F (both)

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication Applies To
QMA 430 covers a mass range of 1–300 u; QMA 400 covers 1–512 u. p.2 ↗ Select QMA 400 when masses above 300 u must be resolved; QMA 430 suffices for lighter gas analysis. QMA 430 / QMA 400
Detection limit is <2·10⁻¹⁵ hPa (mbar) for QMA 430 and <1·10⁻¹⁵ hPa (mbar) for QMA 400. p.2 ↗ Trace-gas detection below the ppm level is feasible; QMA 400 offers roughly 2× better detection limit. QMA 430 / QMA 400
Argon sensitivity is 2·10⁻⁴ A/hPa (QMA 430) and 5·10⁻⁴ A/hPa (QMA 400). p.2 ↗ Sensitivity difference of 2.5× should be factored into signal-to-noise and minimum detectable partial pressure calculations. QMA 430 / QMA 400
Maximum operating pressure is 1·10⁻⁴ / 1·10⁻⁵ hPa (mbar) for both analyzers. p.2 ↗ The analyzer must not be exposed to pressures above ~1·10⁻⁴ hPa; appropriate pumping and pressure interlocks are required to protect the filament/SEV. whole series
Partial pressure ratio with SEV (secondary electron multiplier) is <1 ppb (QMA 430) and <0.5 ppb (QMA 400). p.2 ↗ Using the SEV option enables sub-ppb detection of impurities; for maximum sensitivity specify SEV-equipped configurations. QMA 430 / QMA 400
Minimum measurement speed spans 125 µs/u up to 60 s/u for both models. p.2 ↗ Supports both fast process monitoring (125 µs/u) and high-sensitivity slow scans; scan settings can be tuned to process dynamics. whole series
Rod system: stainless steel, 8 mm diameter (QMA 430); molybdenum, 8 mm diameter (QMA 400). p.2 ↗ Mo rods on the QMA 400 typically offer better performance/corrosion resistance; rod diameter influences resolution and mounting clearances. QMA 430 / QMA 400
HF generator model is QMH 800-3 for QMA 430 and QMH 800-5 for QMA 400; both use EP 822 electrometer preamplifier. p.2 ↗ HF generator must be matched to the analyzer model when ordering spares or replacing electronics; EP 822 is common across both, simplifying spares inventory. QMA 430 / QMA 400
Analyzer operating temperature is 150 °C and bake-out temperature is 400 °C for both models. p.2 ↗ System bake-out up to 400 °C is permissible with the analyzer installed, but continuous analyzer operation must be limited to 150 °C; plan heating zones accordingly. whole series
Dimensions are 309 mm length × 432–454 mm width; connecting flange is DN 63 CF-F for both models. p.2 ↗ Vacuum chamber port must provide a DN 63 CF-F (ConFlat) flange; the 432–454 mm width range must be checked against available space in the analysis setup. whole series
Integration is via OPC-UA interface and operation via Pfeiffer MassSpec software. p.1 ↗ Plant control systems (SCADA/DCS) can integrate the RGA natively via OPC-UA without custom drivers; plan IT/network interfaces accordingly. whole series
An upgrade kit from the old to the new (HiQuad Neo) generation is available. p.1 ↗ Existing legacy HiQuad installations can be modernized without replacing the full analyzer/vacuum hardware. whole series
Modular, backwards-compatible design with dedicated spares (electronics, ion source, filament, multiplier, analyzer, cables, wiring). p.1 ↗ Individual modules can be replaced in the field; stocking filament, ion source and multiplier spares minimizes downtime in high-duty vacuum monitoring applications. whole series