Induztr
Engineering Digest Document Viewer Related Models

AMI 1000 — product leaflet

Technical Reference Summary

Covered Models & Document Scope

Pfeiffer Vacuum AMI 1000 leak detector for non-destructive leak testing (lab and inline use) of non-porous pharmaceutical packages (vials, syringes, blister, IV bags). Single model; options: customized test chambers, calibration gas supply, venting gas, customer laptop via Ethernet/WiFi, remote data access.

Section-by-Section Technical Breakdown

pp. 1
Features and Accessories/Options

The AMI 1000 is a scalable leak detector for non-destructive testing in lab and inline applications, delivering deterministic, quantifiable, repeatable results per USP <1207>. It is a turnkey solution with customized test chambers adaptable to vials, syringes, blister, IV bags and other non-porous packages. Accessories and options include calibration gas supply, venting gas, Ethernet/WiFi laptop connection and remote data access.

  • Scalable: lab use and inline testing
  • Deterministic results per USP <1207> standard
  • Customized test chambers for various non-porous package types
  • Options: calibration gas supply, venting gas, customer laptop via Ethernet/WiFi
  • Remote access to data
pp. 2
Technical Specifications

Key performance and utility requirements: sensitivity down to 0.4 µm / 2·10⁻⁵ mbar·l/s, CDA supply at 6.3–10 bar rel. per ISO 8573-1 quality 1.3.1, power supply 90–250 V AC with ~1200 W consumption. Operating environment is limited to 20–25 °C and 30–65 % humidity. Software is 21 CFR Part 11 compliant with PDF GMP test/calibration reports and local or LDAP domain authentication.

  • Sensitivity: down to 0.4 µm, 2·10⁻⁵ mbar·l/s
  • CDA quality: 1.3.1 per ISO 8573-1; pressure 91–145 psig / 6.3–10 bar rel.
  • Typical CDA consumption: 2 Nl/cycle
  • Calibration gas (optional): 91–145 psig / 6.3–10 bar rel.
  • Venting gas (optional): 0–22 psig / 0–1.5 bar rel.
  • Power supply: 90–250 V AC; typical consumption 1200 W
  • Operating temperature 20–25 °C; humidity 30–65 %
  • User interface: 10" multi-touch full HD color screen
  • Software: 21 CFR Part 11 compliance, PDF GMP reports, local or LDAP authentication, optional remote data access
pp. 3
Interfaces, Weight and Dimensions

The unit runs Windows 10 with a 10" multi-touch display and offers network and peripheral connectivity: 1x LAN (RJ45), WiFi AP, 2x USB 3.0, and 1x HDMI. Physical data: ~130 kg weight, dimensions 913 x 558 x 1071 mm (LxWxH), sound pressure level <53 dB(A).

  • Operating system: Windows 10
  • Network: 1x LAN (RJ45), WiFi access point
  • 2x USB 3.0 (external), 1x HDMI
  • Sound pressure level: <53 dB(A)
  • Weight: approx. 130 kg
  • Dimensions: 913 x 558 x 1071 mm

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication Applies To
Detection sensitivity down to 0.4 µm / 2·10⁻⁵ mbar·l/s p.2 ↗ Suitable for verifying package integrity at sub-micron defect sizes per USP <1207>; selection must confirm this meets the required leak-rate limit of the application AMI 1000
CDA (clean dry air) supply is required for operation at 91–145 psig / 6.3–10 bar rel. p.2 ↗ Plant air infrastructure must deliver regulated compressed air within this range before commissioning AMI 1000
CDA quality must be 1.3.1 according to ISO 8573-1 p.2 ↗ Air treatment (particles, water, oil classes 1/3/1) is mandatory; standard shop air will not suffice AMI 1000
Typical CDA consumption is 2 Nl per cycle p.2 ↗ Allows sizing of compressed air supply and estimating running cost per test cycle AMI 1000
Optional calibration gas supply requires 91–145 psig / 6.3–10 bar rel. p.2 ↗ If calibration gas option is ordered, provide a second regulated gas line at the same pressure range as CDA AMI 1000 with calibration gas option
Optional venting gas pressure range 0–22 psig / 0–1.5 bar rel. p.2 ↗ Venting line must be regulated to a much lower pressure than CDA; plan a separate low-pressure supply if venting option is used AMI 1000 with venting gas option
Power supply 90–250 V AC, typical consumption 1200 W p.2 ↗ Single-phase wide-range input; provision ≥1200 W circuit capacity (recommend appropriate breaker/headroom) AMI 1000
Operating temperature limited to 20–25 °C with 30–65 % humidity p.2 ↗ Requires climate-controlled room; not usable in standard industrial environments without HVAC — critical for inline installation planning AMI 1000
Software is 21 CFR Part 11 compliant with PDF GMP test and calibration reports and local or LDAP domain authentication p.2 ↗ Suitable for regulated pharma environments; plan authentication mode (local vs domain) during IT integration AMI 1000
Windows 10 operating system with 10" multi-touch full HD screen p.2 ↗ Consider Windows 10 lifecycle/patching policy in validation and IT planning AMI 1000
Interfaces: 1x LAN (RJ45), WiFi AP, 2x USB 3.0 (external), 1x HDMI p.3 ↗ Network data export/remote access via RJ45 or WiFi; USB for peripherals/backup, HDMI for external display AMI 1000
Weight approx. 130 kg; dimensions 913 x 558 x 1071 mm (LxWxH) p.3 ↗ Plan handling equipment (two-person lift or hoist) and verify bench/floor loading and footprint in layout AMI 1000
Sound pressure level <53 dB(A) p.3 ↗ Acceptable for lab/office-adjacent areas without additional noise mitigation AMI 1000
Test chambers are customized; adaptable to vials, syringes, blister, IV bags and other non-porous packages p.1 ↗ Chamber must be specified per package format at ordering time — not a universal off-the-shelf fixture AMI 1000

Related models