OmniStar — product leaflet
Technical Reference Summary
Covered Models & Document Scope
Pfeiffer Vacuum OmniStar GSD 350 and ThermoStar GSD 350 quadrupole mass spectrometers (gas analysis versions of the GSD 350 series). Variants: mass ranges 1-100 / 1-200 / 1-300 u, standard or corrosive gas version, choice of filament material, capillary operating temperature versions (200/350 °C), and various inlet systems. The two models differ mainly in gas connection/inlet: OmniStar uses stainless steel or quartz capillary with a software-controlled inlet valve; ThermoStar uses a quartz capillary continuously open.
Section-by-Section Technical Breakdown
The document positions the OmniStar/ThermoStar GSD 350 as user-friendly, compact and lightweight portable quadrupole mass spectrometers for gas analysis. Key claims include automatic calibration and tuning, low detection limits, multi-gas analysis, and detection of small concentrations of reactive and condensable gases.
- User-friendly: easy operation, automatic calibration and tuning
- Compact and lightweight, portable
- High performance: low detection limits, multi-gas analysis
- Capable of detecting even small concentrations of reactive and condensable gases
A list of available spare parts and configurable options for the GSD 350 series. Options affect mass range, gas compatibility, filament material, calibration and the inlet system.
- Spare inlet systems, electronics, ion source, filament, analyzer
- Spare vacuum gauge and vacuum pump
- Mass ranges as an option
- Standard or corrosive gas version
- Choice of filament material
- Mass calibration option
- Various inlet systems
Side-by-side specifications for OmniStar and ThermoStar variants, including mass range, detection limit, rod system, electronics (PrismaPro), detectors, resolution, interfaces and physical/electrical data. Both models share nearly all performance parameters; the main difference is the gas inlet arrangement.
- Mass range: 1-100 / 1-200 / 1-300 u (both models)
- Gas connection: OmniStar stainless steel or quartz capillary; ThermoStar quartz capillary only
- Gas inlet: OmniStar via software-controlled inlet valve or user interface; ThermoStar continuously open
- Detection limit: < 100 ppb / < 1 ppm / < 10 ppm depending on mass range variant
- Sample gas pressure up to 1200 hPa (mbar); two-stage, segregation-free pressure reduction
- Rod system: stainless steel, 6 mm diameter, 125 mm length
- Electronics: PrismaPro; detectors: C-SEM and Faraday
- Contribution to neighboring mass: < 10 / < 20 / < 50 ppm; resolution (adjustable, at 10% peak height) 0.5-2.5 u
- Dwell time 1 ms - 16 s/u; gas flow rate 1-2 sccm
- Dimensions 615 x 358 x 274 mm; weight approx. 23-26 kg; supply voltage 100-240 V AC
- Operating temperature of capillary up to 200/350 °C depending on version
Connectivity, control and software environment for both models. Operation is via a 7" color touch screen or a web user interface, with PV MassSpec as the software platform.
- Interface: Ethernet; analog input 5x ±10 V / 16 bit; analog output 4x 0...10 V / 16 bit; digital input 4x; digital output 6x sink, optically insulated, 24 V; relay 2x
- Operation via 7" colored touch screen or Web User Interface
- Software: PV MassSpec
Key Integration, Operating & Safety Requirements
| Specification / Condition | Engineering Implication | Applies To |
|---|---|---|
| Detection limit is < 100 ppb for the 1-100 u mass range, < 1 ppm for 1-200 u, and < 10 ppm for 1-300 u p.2 ↗ | Select the smallest mass range that covers the target gases to obtain the best detection limit; trace-gas applications below 1 ppm require the 1-100 u variant | both OmniStar and ThermoStar GSD 350 |
| Sample gas pressure can be up to 1200 hPa (mbar), handled via a two-stage, segregation-free pressure reduction p.2 ↗ | The analyzer can sample near-atmospheric process gas directly without an external vacuum pumping stage between sample and inlet, provided the built-in pressure reduction is used | whole series |
| Capillary operating temperature is up to 200 °C or 350 °C depending on version p.2 ↗ | Choose the 350 °C version for hot process gases or condensable species; the capillary temperature must be set above the dew point of condensable components | whole series, ordering variant dependent |
| Rod system is stainless steel, 6 mm diameter, 125 mm length p.2 ↗ | Small-diameter rod system defines resolution/sensitivity trade-off; corrosive-gas applications require the corrosive version option since bare stainless steel is attack-prone | whole series |
| Resolution is adjustable 0.5-2.5 u at 10% peak height; contribution to neighboring mass is < 10 / < 20 / < 50 ppm p.2 ↗ | Resolution can be tuned per application; peak overlap (e.g., isobaric fragments) must be checked against the ≤10-50 ppm crosstalk figures for the respective mass range variant | whole series |
| Dwell time is adjustable from 1 ms to 16 s per u; gas flow rate is 1-2 sccm p.2 ↗ | Fast events can be monitored at 1 ms dwell times; the low sample flow (1-2 sccm) means minimal disturbance to the process gas system | whole series |
| OmniStar gas inlet is via a software-controlled inlet valve or user interface; ThermoStar inlet is continuously open p.2 ↗ | For applications requiring intermittent sampling or inlet control by the measurement routine, choose OmniStar; ThermoStar (quartz capillary only, continuously open) suits continuous high-temperature process monitoring | model selection OmniStar vs ThermoStar |
| Supply voltage is 100-240 V AC p.2 ↗ | No line-voltage variant selection needed; unit connects to standard single-phase supplies worldwide | whole series |
| Dimensions are 615 x 358 x 274 mm and weight approx. 23-26 kg p.2 ↗ | Bench/packaging design must accommodate a ~26 kg portable unit; mounting surfaces and handles must be rated accordingly | whole series |
| Mass spectrometer electronics are the PrismaPro platform with C-SEM and Faraday detectors p.2 ↗ | Detector choice (C-SEM for low-level signals, Faraday for high signals/large dynamic range) affects detection strategy; spare analyzer/ion source/filament are available as spare parts | whole series |
| I/O: Ethernet; analog input 5x ±10 V / 16 bit; analog output 4x 0...10 V / 16 bit; digital input 4x; digital output 6x sink, optically insulated, 24 V; relay 2x p.3 ↗ | PLC integration possible without additional I/O converters; digital outputs are sink-type, optically isolated at 24 V — design the receiving circuitry accordingly | whole series |
| Operation via 7" color touch screen or Web User Interface; software is PV MassSpec p.3 ↗ | Remote commissioning/monitoring can run through the Ethernet web UI; PV MassSpec license/system must be part of the plant software plan | whole series |
| Mass ranges available: 1-100 u, 1-200 u, 1-300 u; options include standard vs corrosive gas version, filament material choice, mass calibration option and various inlet systems p.1 ↗ | Ordering must specify mass range, gas version (corrosive for aggressive media), filament material, inlet system and capillary temperature version (200/350 °C) | whole series, ordering configuration |
| Automatic calibration and tuning are built in; the instrument supports analysis of reactive and condensable gases p.1 ↗ | No manual tuning expertise needed for routine operation, but corrosive-gas or condensable-species sampling should use the corresponding version/heated capillary option to avoid damage and condensation | whole series |