Technical Reference Summary

Covered Models & Document Scope

Pfeiffer Vacuum COMBI RU vacuum pump unit series (booster + oil-lubricated rotary vane backing pump combinations). 16 variants: COMBI RU 1002, 1052, 1102, 1152, 1454, 1604, 1654, 1754, 2004, 2204, 2304, 2954, 3404, 3554, 4704. Combinations of HiLobe boosters (1302, 2104, 2704, 4504, 6204) with R5 backing pumps (RD 0240 A, RD 0360 A, RA 0400 C, RA 0630 C).

Section-by-Section Technical Breakdown

pp. 1
Product Features

The COMBI RU is positioned as a reliable, high-performance, flexible vacuum pump unit. Reliability comes from process monitoring via integrated sensors and an intelligent booster user interface; performance from high pumping speed, fast pump-down and energy efficiency. Flexibility comes from multiple design options and booster/backing-pump combinations.

  • High uptime and increased safety through process monitoring via integrated sensors
  • Intelligent user interface of vacuum booster
  • High pumping speed, fast pump-down times, high energy efficiency
  • Various combinations of oil-lubricated rotary vane backing pumps and vacuum boosters available
pp. 1
Accessories, Spare Parts and Options

The document lists available accessories and options for the COMBI RU units, covering control, monitoring, separation and mounting options. These allow configuration to specific process and installation requirements.

  • Control unit for booster and backing pump
  • Operating fluid level monitoring
  • Operating fluid temperature switch
  • KAS condensate separator
  • SAS dust separator
  • Gas-ballast valve closed or with hand valve
  • Feet options: height-adjustable machine foot, castor wheels, floor mounting
pp. 2-3
Technical Data (All Variants)

Performance data tables for all 16 COMBI RU variants, specifying pumping speed at 1 hPa, ultimate pressure (with and without gas ballast), power consumption, supply voltage, noise level, weight and flange connections. Pumping speeds range from 1028 to 4689 m³/h; all units run on 400 V AC (±10%), 50 Hz.

  • Pumping speed at 1 hPa: 1028 m³/h (RU 1002) up to 4689 m³/h (RU 4704)
  • Ultimate pressure: 7·10⁻³ hPa (mbar) to 1·10⁻² hPa (mbar); with gas ballast 1·10⁻² to 3·10⁻² hPa (mbar)
  • Max. power consumption: 6.5 kW (RU 1002) to 22.5 kW (RU 4704)
  • Input voltage all variants: 400 V AC (±10%), 50 Hz
  • Noise level (ISO 2151, at 1 hPa): ≤80 dB(A) for RD-based units, ≤85 dB(A) for most RA 0630 C based units
  • Weight: 415 kg (RU 1002) up to 1200 kg (RU 2304/3554/4704)
  • Gas inlet/outlet: DN 100 or DN 160 ISO-F/PN 16 inlet; DN 63 or DN 100 ISO-K outlet
  • Booster range: HiLobe 1302 to HiLobe 6204; backing pumps: R5 RD 0240 A, RD 0360 A, RA 0400 C, RA 0630 C

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication Applies To
All COMBI RU variants require 400 V AC (±10%), 50 Hz supply. p.2 ↗ Single supply standard simplifies electrical design; a 400 V ±10% tolerance (360–440 V) must be guaranteed on site. whole series
Best ultimate pressure without gas ballast is 7·10⁻³ hPa (mbar), achieved by HiLobe 1302-based units (RU 1002, 1052, 1102, 1152). p.3 ↗ For lowest-pressure applications, choose a HiLobe 1302 based variant; larger HiLobe combos bottom out at 1·10⁻² hPa. RU 1002, 1052, 1102, 1152
Gas ballast degrades ultimate pressure by roughly 2–3× (e.g. RU 1002: 7·10⁻³ → 2·10⁻² hPa; RU 2004: 1·10⁻² → 3·10⁻² hPa). p.2 ↗ When condensable vapors require gas ballast operation, spec system base pressure accordingly. whole series
Max pumping speed is 4689 m³/h at 1 hPa for COMBI RU 4704 (HiLobe 6204 + R5 RA 0630 C). p.2 ↗ Upper bound of series throughput for sizing process pump-down times. COMBI RU 4704
Max power consumption ranges 6.5–22.5 kW; the RA 0630 C-based units draw 16–22.5 kW vs 6.5–9 kW for the smaller RD-based units. p.2 ↗ Electrical supply, breakers and cabling must be sized per variant; RD-based units fit smaller feeds. whole series
Noise levels per ISO 2151 at 1 hPa are ≤80 dB(A) for R5 RD-backed units and ≤85 dB(A) for most R5 RA 0630 C-backed units. p.2 ↗ RA 0630 C-based units may require workplace noise precautions or acoustic measures. whole series (per backing pump type)
Weights range from 415 kg (RU 1002) to 1200 kg (RU 2304, 3554, 4704); most RA 0630 C-based units are ~1000–1200 kg. p.2 ↗ Floor loading and lifting equipment must be planned; the heaviest variants need ~1.2 t rigging capacity. whole series
Gas inlet flanges are DN 100 ISO-F/PN 16 (smaller variants) or DN 160 ISO-F/PN 16 (RU 2004, 2204, 2954, 3404, 2304, 3554, 4704); outlets are DN 63 or DN 100 ISO-K. p.2 ↗ Piping and foreline components must match the specific variant's flange sizes. per variant
Booster options are HiLobe 1302, 2104, 2704, 4504, 6204; backing pumps are R5 RD 0240 A, R5 RD 0360 A, R5 RA 0400 C, R5 RA 0630 C. p.2 ↗ Only these validated booster/backing-pump combinations define valid ordering choices within the series. whole series
Sensors for operating fluid level monitoring and an operating fluid temperature switch are available as options. p.1 ↗ Recommended for unattended or safety-critical operation to protect against oil loss/overheating. whole series
KAS condensate separator and SAS dust separator are offered as accessories. p.1 ↗ Specify these for wet or particle-laden processes to protect the pumps. whole series
Gas-ballast valve can be ordered closed or with hand valve. p.1 ↗ Choose the hand-valve option when periodic vapor handling is needed without permanent gas ballast penalty. whole series
Mounting options: height-adjustable machine foot, castor wheels, or floor mounting. p.1 ↗ Castor option enables mobile deployment; floor mounting for fixed, vibration-critical installations. whole series
COMBI RU 1152 pairs HiLobe 1302 with R5 RA 0630 C, giving 1204 m³/h at 17.5 kW and 7·10⁻³ hPa ultimate pressure. p.3 ↗ Unusual high-power/small-booster combination for applications needing low ultimate pressure plus large backing pump capacity. COMBI RU 1152