ASM 310 — product leaflet
Technical Reference Summary
Covered Models & Document Scope
Pfeiffer Vacuum ASM 310 portable helium/hydrogen mass-spectrometer leak detector (single model). Options/accessories listed: sniffer probe with 5 or 10 m hose, helium spray gun, remote control, trolley, transport box, country-specific power cable.
Section-by-Section Technical Breakdown
The ASM 310 is presented as a portable, compact helium leak detector with retractable handle, low weight and a detachable control panel. It targets best sensitivity-to-size-ratio and maintenance/service leak-detection applications.
- Portable design: compact, retractable handle, lightweight, detachable control panel
- Targeted at maintenance applications and leak detection service providers
- Accessories/options: sniffer probe (5 or 10 m hose), helium spray gun, remote control, trolley, transport box, country-specific power cable
Key performance data: detectable gases, minimum detectable leak rates for spraying and sniffing tests, pumping speed, inlet pressure limit, power supply, start-up and response times. Based on diaphragm backing pump with 1.7 m³/h roughing capacity.
- Test methods: spraying (vacuum) and sniffing tests
- Detectable gases: ⁴He, ³He, H₂
- Min. detectable leak rate (He, spraying): 1·10⁻¹² mbar·l/s (1·10⁻¹³ Pa·m³/s)
- Min. detectable leak rate (He, sniffing): 1·10⁻⁷ mbar·l/s (1·10⁻⁸ Pa·m³/s)
- Helium pumping speed: 1.1 l/s; backing pump: diaphragm; roughing capacity 1.7 m³/h
- Power supply 90–240 V, 50/60 Hz; max power consumption 300 W; max test inlet pressure 15 mbar
- Start-up time <2 min (20 °C); response time <1 s; noise <45 dB(A); operating temperature 10–40 °C; weight approx. 21 kg
Physical envelope and vacuum interface of the unit. Dimensions are 350 × 254 × 414 mm (L×W×H) with a DN 25 ISO-KF inlet port.
- Dimensions (L×W×H): 350 × 254 × 414 mm
- Inlet port: DN 25 ISO-KF
Key Integration, Operating & Safety Requirements
| Specification / Condition | Engineering Implication | Applies To |
|---|---|---|
| Minimum detectable leak rate for helium in spraying (vacuum) test is 1·10⁻¹² mbar·l/s (1·10⁻¹³ Pa·m³/s) p.2 ↗ | Suitable for high-sensitivity vacuum leak checking; spec the unit where vacuum-mode leak rates down to 1e-12 mbar·l/s must be detected | ASM 310 |
| Minimum detectable leak rate for helium in sniffing test is 1·10⁻⁷ mbar·l/s (1·10⁻⁸ Pa·m³/s) p.2 ↗ | Five orders of magnitude worse than vacuum mode; use sniffer mode only for external/pressure-test applications where 1e-7 sensitivity suffices | ASM 310 |
| Detectable tracer gases are ⁴He, ³He and H₂ p.2 ↗ | Hydrogen can be used as tracer gas alternative to helium; ³He detection also supported | ASM 310 |
| Maximum test inlet pressure is 15 mbar p.2 ↗ | Test objects must be pre-evacuated below 15 mbar before connecting to the inlet or a roughing arrangement is needed; exceeding risks inlet contamination | ASM 310 |
| Helium pumping speed is 1.1 l/s with diaphragm backing pump of 1.7 m³/h roughing capacity p.2 ↗ | Diaphragm pump eliminates oil management; small pumping speed limits usable test volume, so larger chambers need external pre-pumping | ASM 310 |
| Power supply 90–240 V AC, 50/60 Hz, max power consumption 300 W p.2 ↗ | Universal voltage input — no voltage variant selection needed worldwide; size power circuits for ≥300 W | ASM 310 |
| Start-up time at 20 °C is under 2 minutes p.2 ↗ | Ready-to-use in <2 min, suitable for quick service/maintenance call-outs | ASM 310 |
| Response time is under 1 second p.2 ↗ | Fast signal feedback for spray-gun leak pinpointing; account for <1 s delay when moving the spray probe | ASM 310 |
| Operating temperature range is 10–40 °C p.2 ↗ | Do not use or store in service environments outside 10–40 °C; cold starts below 10 °C not specified | ASM 310 |
| Weight is approximately 21 kg; dimensions 350 × 254 × 414 mm p.2 ↗ | Single-person portable; verify lift/handling limits and storage/trolley footprint accordingly | ASM 310 |
| Inlet port is DN 25 ISO-KF p.3 ↗ | Specify ISO-KF DN 25 clamps, centering rings and adapters for test-port connections | ASM 310 |
| Communication interfaces: standard 15-pin I/O and RS-232 p.2 ↗ | Remote triggering/status via 15-pin I/O possible; PC integration limited to RS-232 (no USB/Ethernet in this document) | ASM 310 |
| Sound pressure level below 45 dB(A) p.2 ↗ | Quiet operation permits use in occupied lab/production areas without extra noise mitigation | ASM 310 |
| Sniffer probe is available with 5 m or 10 m hose length p.1 ↗ | Choose hose length based on distance from unit to sniffing point; longer hose may affect sniffer response but option covers remote locations | ASM 310 sniffer probe accessory |