ACP 90 — product leaflet
Technical Reference Summary
Covered Models & Document Scope
Pfeiffer Vacuum ACP 90 multi-stage Roots vacuum pump, dry (maintenance-free) design. Document is a 2-page datasheet covering one model; available options/accessories listed: dust filter, external silencer, isolation valve, preventive maintenance packages, corrosive gas version. Pumping speed curves given for 50 Hz and 60 Hz operation.
Section-by-Section Technical Breakdown
Highlights of the ACP 90: efficiency via an IE5 permanent magnet motor with low power consumption, high pumping speed at high pressures suited to large volumes and frequent pump-downs, high water vapor capacity, and reliability with high-quality bearings and near maintenance-free operation.
- Low power consumption with IE5 permanent magnet motor
- High pumping speed at high pressures — ideal for large volumes and frequent pump-downs
- High water vapor capacity
- Nearly maintenance-free, high-quality bearings, long life cycle
Lists optional accessories and variants available for the ACP 90, including protection and process adaptations.
- Dust filter
- External silencer
- Isolation valve
- Preventive maintenance packages
- Corrosive gas version
Complete performance and installation data: pumping speed, ultimate pressures (with/without gas ballast), power figures, supply voltages, connections, dimensions and weight for the ACP 90.
- Max. pumping speed 88 m³/h (50 Hz and 60 Hz curves provided)
- Ultimate pressure 3·10⁻² mbar without dilution; 1.6·10⁻¹ mbar with gas-ballast valve
- Max. continuous inlet pressure 1013 mbar (atmospheric)
- Max. water pumping capacity 2400 g/h (depending on configuration)
- Power consumption 1.3 kW at ultimate pressure, max. 2 kW
- Supply: 200–240 VAC 1-/2-phase; 380–480 VAC 3-phase
- Air cooled; sound pressure level <65 dB(A) at ultimate pressure
- Flanges: DN 40 ISO-KF inlet and outlet
- Dimensions 671 x 319 x 324 mm; weight 75 kg; max. leak rate <5·10⁻⁶ hPa·l/s
Key Integration, Operating & Safety Requirements
| Specification / Condition | Engineering Implication | Applies To |
|---|---|---|
| Max. pumping speed is 88 m³/h p.90 ↗ | Size the backing/foreline and pump-down time calculations around 88 m³/h; performance differs between 50 Hz and 60 Hz per the pumping speed curves. | ACP 90 |
| Ultimate pressure is 3·10⁻² mbar without dilution and 1.6·10⁻¹ mbar with gas-ballast valve open p.90 ↗ | Using gas ballast for vapor handling degrades achievable base pressure ~5x; select based on whether vapor purge or low ultimate pressure is the priority. | ACP 90 |
| Max. continuous inlet pressure is 1013 mbar p.90 ↗ | Pump can start and run from atmospheric pressure without a bypass or soft-start staging — suitable for direct pump-down of vented vessels. | ACP 90 |
| Max. water pumping capacity is 2400 g/h, depending on configuration p.90 ↗ | For condensable-heavy processes, verify the configuration supports this capacity; consider gas ballast operation (which raises ultimate pressure). | ACP 90 (configuration-dependent) |
| Power consumption is 1.3 kW at ultimate pressure, 2 kW max p.90 ↗ | Dimension supply/breaker sizing for 2 kW peak; the IE5 PM motor keeps running power low at base pressure. | ACP 90 |
| Supply voltage is 200–240 VAC (1-/2-phase) or 380–480 VAC (3-phase) p.90 ↗ | Check site power availability before ordering; both single-phase and three-phase installations are supported. | ACP 90 |
| Sound pressure level is <65 dB(A) at ultimate pressure p.90 ↗ | Acceptable for lab/close-proximity installation without mandatory acoustic enclosure; external silencer available as option for further reduction. | ACP 90 |
| Cooling method is air cooling p.90 ↗ | Ensure adequate ambient airflow/clearance at installation; no cooling water circuit required. | ACP 90 |
| Max. leak rate is <5·10⁻⁶ hPa·l/s p.90 ↗ | Tight enough for clean/high-vacuum train integration; use this as acceptance criterion for leak checking after installation. | ACP 90 |
| Inlet and outlet connections are both DN 40 ISO-KF p.90 ↗ | Simplify installation with standard KF40 clamps, centering rings and O-rings; same fitting on foreline side. | ACP 90 |
| Dimensions 671 x 319 x 324 mm (L x W x H), weight 75 kg p.90 ↗ | Plan mounting frame and lifting/handling for 75 kg; footprint fits compact skids. | ACP 90 |
| A corrosive gas version and preventive maintenance packages are available p.90 ↗ | Specify the corrosive-gas variant for aggressive process gases rather than modifying the standard pump; plan maintenance via the PM packages. | ACP 90 (option) |
| Pumping speed curves are provided separately for 50 Hz and 60 Hz supply frequencies p.90 ↗ | Use the correct frequency curve for pump-down calculations per the installation region/mains supply. | ACP 90 |