PHD 4 Capillary Leak

Technical Reference Summary

Covered Models & Document Scope

Agilent PHD-4 Capillary Leak, model no. 9693540, calibration accessory for the Agilent PHD-4 leak detector with standard-length probe

Technical Data

Parameter Value Condition
Part number 9693540 p.1 ↗ PHD-4 Capillary Leak
Reservoir volume 110 cc p.1 ↗ stainless steel reservoir
Pressure gauge range 160 psig p.1 ↗
Helium concentration of the order of 100 ppm p.1 ↗ reservoir pressure 100 PSI, 7 bar
Helium recharging valve connection Gc 1/4 p.2 ↗
O-ring Parker 2-120 BUNA (2 pcs.) p.1 ↗ supplied in nylon bag
Stabilization time at least 5 minutes p.2 ↗ before calibration
Zero reading wait time usually about 10 sec p.2 ↗ N2 purge until 0.0 PPM

Operating Limits

  • Pressure gauge maximum: 11 bar (160 psig) p.1 ↗
  • Calibration curve validity: PHD-4 devices with standard-length probe only p.1 ↗

Ordering Data

Ordering Code Part No. Variant / Code Applies To Note
PHD-4 Capillary Leak p.1 ↗ 9693540 Calibration accessory for PHD-4 leak detector