Quick Reference Guide: Gas Chromatography Detectors

Technical Reference Summary

Covered Models & Document Scope

Overview of Agilent GC detectors: FID, NPD (N), NPD (P), µECD, TCD, FPD (S), FPD (P), NCD, SCD. Covers description, linear dynamic range, destructiveness and typical analytes for each detector.

Technical Data

Parameter Value Condition
Linear dynamic range >107 (±10%) p.1 ↗ FID
Destructive Yes p.1 ↗ FID
Linear dynamic range 105 (±10%) p.1 ↗ NPD (N)
Destructive No p.1 ↗ NPD (N)
Linear dynamic range >5 × 104 p.1 ↗ µECD
Linear dynamic range >104 p.2 ↗ NCD
Destructive Yes p.2 ↗ NCD
Linear dynamic range >104 p.2 ↗ SCD
Destructive Yes p.2 ↗ SCD
Relative sensitivity 10–15 g g p.2 ↗ NPD (P), 1 ppt
Relative sensitivity 10–12 g g p.2 ↗ NPD (N), 1 ppb
Relative sensitivity 10–9 g g p.2 ↗ µECD, 1 ppm
Relative sensitivity 10–6 g g p.2 ↗ FID, 0.1%
Relative sensitivity 10–3 g g p.2 ↗ TCD, %

Operating Limits

  • Detection limits and ranges depend on compound structure and analytical conditions: See GC data sheets for actual specifications p.2 ↗