Engineering Digest Document Viewer Related Models

C142-E050

Technical Reference Summary

Covered Models & Document Scope

ALTRACE energy-dispersive X-ray fluorescence (EDX) elemental analyzer; document also references EDX-7200 for comparison. Covers standard ALTRACE configuration with 48-sample autosampler, contamination-proof cup, vacuum measurement unit, and six automatic exchange filters.

Technical Data

Parameter Value Condition
Detection limit 1 ppm p.4 ↗ Pb in aqueous solution, 600 seconds measurement time
Detection limit 5 ppm p.4 ↗ Pb, EDX-7200 comparison, 600 seconds measurement time
Detection limit 1 ppm p.4 ↗ Cd in aqueous solution, 600 seconds measurement time
Detection limit 5 ppm p.4 ↗ Cd, EDX-7200 comparison, 600 seconds measurement time
Measurement range sub-ppm to % p.5 ↗ batch element analysis
Measurement time 600 seconds p.4 ↗ detection limit measurements
X-ray generator cooling method Air-cooled (with fan) p.11 ↗
Sample replacement 48-sample autosampler p.11 ↗
Protective function Contamination-proof cup p.11 ↗
Data processor SSD 500 GB min. p.11 ↗
Filter #1 energy range 14 to 38 keV p.5 ↗ elements Mo,Rh,Pd,Ag,Cd,Sn,Sb
Filter #2 energy range 2 to 4 keV p.5 ↗ elements S,Cl
Filter #3 energy range 5 to 10 keV p.5 ↗ elements Cr,Mn,Fe,Co,Ni
Filter #4 energy range 9 to 10 keV p.5 ↗ element Hg
Filter #5 energy range 9 to 14 keV p.5 ↗ elements As,Br,Zr,Hg,Pb,Bi
Filter #6 energy range 4 to 5 keV p.5 ↗ elements Ti,V

Operating Limits

  • Minimum detectable concentration (Pb, aqueous): 1 ppm p.4 ↗
  • Minimum detectable concentration (Cd, aqueous): 1 ppm p.4 ↗
  • Measurement range: sub-ppm to % level p.5 ↗