BL Series electronic balances: BL-320S, BL-620S, BL-3200S (S series) and BL-220H, BL-320H(L), BL-2200H, BL-3200H(L) (H series). Windbreak is supplied only with BL-220H and BL-320H(L). AC adapter is an option. Verified (legal-for-trade) balances are excluded from menu/calibration functions marked with a note.

Section-by-Section Technical Breakdown

pp. 14-15
Components and Their Names

Identifies the balance body, pan, pan supporter, pan shaft, level gauge, power jack, three level screws, protection cover, optional AC adapter, data I/O connector, and windbreak main/lid (BL-220H and BL-320H(L) only). Also describes the display indicators and keypad functions.

  • Included: balance body, pan, pan supporter, protection cover, instruction manual, inspection certificate; AC adapter optional.
  • Windbreak main and lid supplied only with BL-220H and BL-320H(L).
  • Display indicators: calibration mark, communication mark, analog indicator, auto print mark, standby mark, unit mark.
  • Keys: POWER/BRK, CAL/MENU, TARE, UNIT, PRINT.
pp. 16-17
Installation

Covers power voltage check, AC adapter requirements, installation site restrictions, protection cover fitting, leveling with the level screws and level gauge, pan assembly, power-on self-check, and entry to standby/measurement state.

  • Use only the AC adapter authorized by Shimadzu and check that the supply voltage matches the adapter rating.
  • Use a 3P grounded plug or connect the grounding wire to the rear screw for 2P plug.
  • Avoid corrosive/flammable gas, dust, wind, vibration, electromagnetic waves, magnetic fields, direct sunlight, sudden temperature change, and extreme temperature/humidity.
  • Level by centering the air bubble in the red circle using the level screws; adjust right-back screw to touch the floor last.
  • After power-on self-check, press POWER to light all displays for one minute, then zero is shown.
pp. 18
Warm-Up

Specifies warm-up time and standby practice for accurate measurement.

  • Apply power for one hour or more before accurate measurement.
  • Keep the standby mark lit by pressing POWER without disconnecting the AC adapter when not in use.
  • Disconnect the AC adapter if the balance is not used for one month or more.
pp. 20
Measurement Procedure

Describes entering measurement mode, taring, loading samples, reading the stability mark, and overload indication.

  • Press POWER; standby mark goes off and all displays light; check no segment is missing.
  • Load tare, press TARE after stability mark lights, confirm zero, then load sample.
  • If total mass exceeds weighing capacity, an overload indication is displayed.
  • Stability mark lights when the displayed value falls within the stability band.
pp. 21-24
Menu Selection

Explains the three-level menu structure for selecting measurement mode (standard, sample pouring, high stability), zero tracking, stability detection band, unit registration, auto print, analog display, and input/output format.

  • Standard mode is suitable for most use; stability mark lights when display stays within ±1 unit for a fixed period.
  • High-stability mode is recommended in poor measurement environments; sample mode for high-speed or small-amount pouring.
  • Zero tracking should be ON for normal weighing and OFF when measuring weight changes over time or slowly adding liquid/powder.
  • Input/output format menu sets baud rate (300–9600), delimiter (C/R, L/F, C/R+L/F, comma), parity (even/odd/none), and stop bit (1/2).
pp. 25-26
Calibration

Covers setting the calibration weight value within the specified range and performing span calibration, including required warm-up, leveling, zeroing, and weight placement steps.

  • Calibration weight value can be set freely within the specified range per model.
  • Calibration requires warm-up, level check, zeroing, and stable display before and after loading the weight.
  • Calibration is not applicable to a verified balance approved as a legal measuring instrument.
  • If a weight outside the specified range is entered, an error is displayed and the balance returns to weight display.
pp. 27
Registration, Cancel, and Change of Unit

Describes registering up to three units from 14 available units, canceling registration, and changing the displayed unit.

  • Registerable units: g, kg, ct, pcs, %, oz, ozt, dwt, GN, Hong-kong tail, Singapore tail, Taiwan tail, Malaysia tail, Japanese monme.
  • % and pcs cannot be registered simultaneously.
  • Registering a fourth unit deletes the oldest registration.
  • Changing to % or pcs requires a reference value setting.
pp. 28
% Setting

Describes percent display setup using a reference sample set to 100%, including minimum displayed value dependence on reference sample weight.

  • Register % unit, tare, load reference sample, press UNIT until % is displayed, then press PRINT after stability.
  • Minimum displayed value changes with reference sample weight (REF.) according to model type H or S.
  • If % conversion is not possible, an error is displayed and the balance returns to weight display.
pp. 29
PCS (No. of pieces) Setting

Describes piece-counting setup using standard sample quantities of 10, 20, 50, or 100 pieces.

  • Standard piece counts: 10, 20, 50, or 100 pcs; accuracy improves with larger piece count.
  • Register PCS unit, tare, load standard sample, select PCS display, and press PRINT.
  • If reference sample weight is less than readability × number of set pieces, PCS setting is not made and an error is displayed.
pp. 30
Performance Checks

Defines repeatability and eccentric error checks to verify conformance to specifications.

  • Perform checks in a room without sudden temperature changes after sufficient warm-up.
  • Repeatability: load/unload 10 times near capacity; standard deviation must be less than 1.5 times the specified value.
  • Eccentric error limits: BL-320H(L)/3200H(L) 7 counts; BL H series 4 counts; BL S series 2 counts.
  • Eccentric test uses a sample of approximately 1/4 capacity moved to five pan positions.
pp. 31
Maintenance

Provides cleaning instructions and cautions for the balance and pan.

  • Wipe dirt with a soft cloth and a small amount of mild detergent.
  • Do not use organic solvents or chemical dusters; they may damage painted surfaces and the display panel.
  • Use the protection cover in dusty or dirty environments.
  • The pan can be washed with water; dry it sufficiently before mounting.
pp. 33
Specifications

Lists capacity, readability, standard deviation, linearity, calibration weight, pan size, dimensions, weight, sensitivity stability, environmental limits, and power ratings for all models.

  • S series capacities: 320 g, 620 g, 3200 g; readabilities 0.01 g, 0.01 g, 0.1 g.
  • H series capacities: 220 g, 320 g, 2200 g, 3200 g; readabilities 0.001 g or 0.01 g.
  • Ambient temperature/humidity: 5–40 °C, 20–85% RH (no condensation).
  • Rated power: DC12 V, 1.0 A; AC adapter primary 100–240 V, 400 mA, 50/60 Hz.
  • Pollution Degree 2, Overvoltage Category II, altitude up to 2000 m, indoor use only.
pp. 34
Parts List

Lists optional accessories, peripheral devices, calibration weights, cables, and maintenance parts with part numbers.

  • Printers: EP-100 and EP-110 (impact dot).
  • Calibration weights available in OIML F1/F2 classes with part numbers S321-55846, S321-55860, S321-55861, S321-55849, S321-55863.
  • I/O RS cable S321-75885-01; USB serial adapter S321-62520.
  • Maintenance parts include pans, pan supporters, guard, guard cover, protection cover, windbreak main and lid.
pp. 35-37
Peripheral Devices

Covers connecting printers and personal computers, input/output format, and command codes.

  • Disconnect balance and printer AC adapters before connecting to the data I/O connector.
  • Manual printing: PRINT key outputs displayed value.
  • Autoprint: in g display, when within zero ±3 counts and a sample over 20 counts is loaded and stable, the balance prints automatically.
  • PC connection uses USB serial adapter and I/O-RS converter cable; USB driver may require prior installation without Internet access.
  • Output data format is ASCII (JIS); baud rate, parity, and delimiter set via menu.

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication Applies To
Warm-up time of one hour or more is required before accurate measurement. p.18 ↗ Allow sufficient power-on time before calibration or critical weighing. whole series
Ambient temperature and humidity range: 5–40 °C, 20–85% RH, no condensation. p.33 ↗ Install only within these environmental limits; condensation must be avoided. whole series
Rated electric power supply: DC12 V, 1.0 A; AC adapter primary: AC100–240 V, 400 mA, 50/60 Hz. p.33 ↗ Use only the authorized AC adapter matching the supply voltage. whole series
Pollution Degree 2, Overvoltage Category II, altitude up to 2000 m, indoor use only. p.33 ↗ Do not install outdoors or in environments exceeding these ratings. whole series
Calibration weight settable range: BL-320S >150 g and <capacity; BL-220H >100 g and <capacity; BL-620S >300 g and <capacity; BL-320H(L) >150 g and <capacity; BL-3200S >1500 g and <capacity; BL-2200H >1000 g and <capacity; BL-3200H(L) >1500 g and <capacity. p.26 ↗ Select a calibration weight within the model-specific range. per model
Eccentric error limits: BL-320H(L), BL-3200H(L): 7 counts; BL H series: 4 counts; BL S series: 2 counts. p.30 ↗ Use these limits when performing eccentricity checks. per series
Repeatability check: standard deviation must be less than 1.5 times the specified value after 10 load/unload cycles near capacity. p.30 ↗ Acceptance criterion for repeatability verification. whole series
Standard mode stability mark lights when display stays within ±1 unit (resolution value) for a fixed period. p.23 ↗ Defines stability indication in standard measurement mode. whole series
Zero tracking should be ON for normal weighing and OFF when measuring weight changes over time or slowly adding liquid/powder. p.23 ↗ Prevents zero drift from masking slow weight changes. whole series
Autoprint triggers when display is within zero ±3 counts and a sample over 20 counts of g display is loaded and stable. p.35 ↗ Defines automatic print threshold for printer operation. whole series
Baud rate options: 300, 600, 1200, 2400, 4800, 9600; delimiter options: C/R, L/F, C/R and L/F, comma; parity: even, odd, none; stop bit: 1 or 2. p.24 ↗ Communication parameters must match the connected device. whole series
Registerable units: g, kg, ct, pcs, %, oz, ozt, dwt, GN, Hong-kong tail, Singapore tail, Taiwan tail, Malaysia tail, Japanese monme; % and pcs cannot be registered simultaneously. p.27 ↗ Plan unit registration accordingly; maximum three units registered. whole series
PCS standard piece counts: 10, 20, 50, or 100 pcs; accuracy improves with larger count. p.29 ↗ Use larger sample counts for better piece-counting accuracy. whole series
If reference sample weight is less than readability × number of set pieces, PCS setting is not made. p.29 ↗ Ensure sufficient sample weight for piece counting. whole series
Sensitivity stability: S series ±10 ppm/°C; H series ±3 ppm/°C (BL-220H, BL-2200H) or ±5 ppm/°C (BL-320H(L), BL-3200H(L)). p.33 ↗ Temperature effects on sensitivity differ by model. per model
Main body size: S series approx. 171(W)×243(D)×77(H) mm (BL-3200S 78 mm); H series approx. 171(W)×243(D)×114(H) mm (BL-2200H, BL-3200H(L) 78 mm). p.33 ↗ Space planning for installation. per model
Pan size: approx. 100×100 mm for BL-320S, BL-220H, BL-320H(L); approx. 159×126 mm for BL-620S, BL-3200S, BL-2200H, BL-3200H(L). p.33 ↗ Select model based on sample size. per model
Design standard period of usage is ten years. p.7 ↗ Performance may not be maintainable beyond this period. whole series

Faults & Troubleshooting

Code / Symptom Possible Cause Remedy Note
Nothing is displayed by connecting the AC adapter to the outlet. p.32 ↗ The AC adapter is disconnected. The electrical board is turned OFF. Check AC adapter connection and power. 13. Troubleshooting
Nothing is displayed by connecting the AC adapter to the outlet. p.32 ↗ There is an internal error in the balance. Contact the nearest Shimadzu sales or service center. 13. Troubleshooting
The display fluctuates. p.32 ↗ Influence from vibration or wind. Improve the installation site. Change the measurement mode to High-stability mode. 13. Troubleshooting
The display fluctuates. p.32 ↗ Influence from electric noise or electromagnetic wave. Maintain a proper distance from the noise source. 13. Troubleshooting
The display does not change from zero even if a sample having the weight near readability is loaded. p.32 ↗ Zero tracking works. Refer to “6. Menu Selection”. 13. Troubleshooting
The display slowly changes when small amount of sample is loaded. (Normal: readability/1 sec.) p.32 ↗ The averaging processing is in High-stability mode. Change the measurement mode if necessary. 13. Troubleshooting
The display has appeared suddenly. p.32 ↗ There has been an instantaneous power failure. Press the POWER key (the balance enters weight display mode). 13. Troubleshooting
Data communication cannot be made. p.32 ↗ Setting of communication parameter is wrong. Wiring of RS-232C cable is wrong. Refer to “16.3 Input/output format”. Check wiring. 13. Troubleshooting
Error displayed during PCS or % setting. p.32 ↗ Set value exceeds the specified range. Refer to “9. % setting” and “10. PCS setting”. 13. Troubleshooting
Does not proceed to next step. (The stability mark does not light) p.32 ↗ Influence from vibration or wind. Improve the installation site. Change the measurement mode to High-stability mode. 13. Troubleshooting
Error displayed during calibration. p.32 ↗ The weight used for calibration is wrong. Check the weight and retry the calibration. 13. Troubleshooting
Error displayed during calibration. p.32 ↗ There is an internal error in the balance. Contact the nearest Shimadzu sales or service center. 13. Troubleshooting