c190-e092

Technical Reference Summary

Covered Models & Document Scope

Nexera™ GPC System (Gel Permeation Chromatography System), standard configuration: CBM-40 system controller, LC-40D solvent delivery unit, DGU-403 degassing unit, SIL-40C autosampler, CTO-40C column oven, RID-20A refractive index detector, LabSolutions workstation with LabSolutions GPC Software; Shim-pack GPC-800/80M/2000 series columns listed separately.

Technical Data

Parameter Value Condition
System configuration Modular p.12 ↗ Nexera GPC System standard
Mobile phase flow rate accuracy No more than 1% or 2 µL/min, whichever is greater (0.01 to 2 mL/min) p.12 ↗ Nexera GPC System standard
Mobile phase flow rate precision No more than 0.06% RSD or 0.02 min SD, whichever is greater p.12 ↗ Nexera GPC System standard
Detector cell volume 9 µL p.12 ↗ RID-20A detector
Detector cell temperature control range 30 to 60 °C p.12 ↗ RID-20A detector
Calibration curve approximation formulae Linear, 3rd-order, 3rd-order + hyperbolic curve, 5th-order, 5th-order + hyperbolic p.12 ↗ LabSolutions GPC Software
Data output method ASCII format p.12 ↗ LabSolutions GPC Software
Mobile phase containers monitored up to twelve containers p.4 ↗ 1-liter containers; up to 4 with 2- to 5-liter containers
Solvent usage savings 30 % p.7 ↗ RID-20A GPC analysis
Stabilization time After 30 Minutes p.7 ↗ Power ON, RID-20A

Operating Limits

  • Detector cell temperature control range: 30 to 60 °C p.12 ↗
  • Mobile phase flow rate range for accuracy specification: 0.01 to 2 mL/min p.12 ↗
  • Mobile phase containers monitored: up to 12 (1 L) or up to 4 (2–5 L) p.4 ↗

Ordering Data

Ordering Code Part No. Variant / Code Applies To Note
228-65502-58 p.12 ↗ 228-65502-58 System Controller CBM-40, qty 1
228-65002-58 p.12 ↗ 228-65002-58 Solvent Delivery Unit LC-40D, qty 1
228-65018-58 p.12 ↗ 228-65018-58 Degassing Unit DGU-403, qty 1
228-65101-58 p.12 ↗ 228-65101-58 Autosampler SIL-40C, qty 1
228-65202-58 p.12 ↗ 228-65202-58 Column Oven CTO-40C, qty 1
228-65306-58 p.12 ↗ 228-65306-58 Refractive Index Detector RID-20A, qty 1
223-62700-92 p.12 ↗ 223-62700-92 LC Workstation LabSolutions, qty 1
223-19189-92 p.12 ↗ 223-19189-92 GPC Software LabSolutions GPC Software, qty 1
228-20803-91 p.11 ↗ 228-20803-91 Shim-pack GPC-801 column, 300 × 8.0 mm, exclusion limit 1.5 × 10³
228-20808-91 p.11 ↗ 228-20808-91 Shim-pack GPC-805 column, 300 × 8.0 mm, exclusion limit 4 × 10⁶