iMScope™ QT — Shimadzu flagship QTOFMS mass spectrometry imaging system with integrated optical microscope; combined MSI + LC–MS (LCMS-9030) configuration; document covers the iMScope QT system and its pretreatment/analysis accessories (matrix vapor deposition system, imaging data analysis software).
Technical Data
| Parameter | Value | Condition |
|---|---|---|
| Ion type (example measurement) | [M+H]+ p.7 ↗ | chlorpromazine, MSI |
| Theoretical m/z | 319.10302 m/z p.7 ↗ | chlorpromazine [M+H]+ |
| Measured m/z | 319.10317 m/z p.7 ↗ | chlorpromazine [M+H]+, iMScope QT |
| Mass accuracy difference | +0.15 mDa p.7 ↗ | chlorpromazine [M+H]+ |
| Mass accuracy (relative) | 0.458 ppm p.7 ↗ | chlorpromazine [M+H]+ |
| Composition estimation score | 98.99 p.7 ↗ | C17H19N2SCl, LabSolutions Insight Explore |
| Spatial resolution (example measurement) | 5 μm p.6 ↗ | cerebellum measurement |
| Flight tube temperature stability | ±0.1 °C p.9 ↗ | intelligent temperature control system |
| Temperature control range (plotted) | 0 to 60 °C p.9 ↗ | flight tube temperature control |
| Detector type | QTOFMS p.3 ↗ | iMScope QT |
| Integrated optical microscope | built-in p.8 ↗ | iMScope QT |
| LC–MS combination | LCMS-9030 p.7 ↗ | combined MSI + LC–MS system |
| Data analysis software | LabSolutions Insight / LabSolutions Insight Explore p.7 ↗ | quantification and composition estimation |
Operating Limits
- Flight tube temperature control range: 0 to 60 °C p.9 ↗
- Flight tube temperature stability: ±0.1 °C p.9 ↗
- Demonstrated spatial resolution: 5 μm p.6 ↗