Multi-Balance Collect PC connection simple manual
Technical Reference Summary
Covered Models & Document Scope
This manual (BA6N-0025, Oct. 2020) covers the Multi-Balance Collect PC software and its connection to Shimadzu balances. Supported balance families: AP, AT-R, MOC63u (USB cable set S321-71730-41); AT, BL, ELB (USB-serial adapter S321-62520 + I/O-RS converter cable S321-71260-01); TW/TX/TXB (USB-Serial conversion kit S321-62520-45); AU, UW/UX/UP, BW-K/BX-K (USB-Serial conversion kit S321-62520-01). MOC120H is noted as usable only with Balance keys. Up to 4 balances can be connected simultaneously.
Section-by-Section Technical Breakdown
The balance communication settings must be set identical to the software communication specifications. The required values are baud rate 1200, parity None, data bit 8, stop bit 1, handshake OFF, delimiter CR. The 'Linked to Windows' setting must be disabled.
- Baud rate (communication speed): 1200
- Parity check: None
- Data bit: 8
- Stop bit: 1
- Handshake: OFF
- Delimiter: CR
- Disable 'Linked to Windows' if set
- Balance and software communication settings must match exactly
Shimadzu genuine cables are required for each balance model; non-genuine cables may cause incorrect data output. The table lists the cable set or kit part numbers per model family.
- AP, AT-R, MOC63u: USB cable set P/N S321-71730-41
- AT, BL, ELB: USB-serial adapter P/N S321-62520 and I/O-RS converter cable P/N S321-71260-01 (both required)
- TW/TX/TXB: USB-Serial conversion kit P/N S321-62520-45
- AU, UW/UX/UP, BW-K/BX-K: USB-Serial conversion kit P/N S321-62520-01
- MOC120H can only be used with Balance keys
With the PC connected to the Internet, the USB driver installs automatically when the balance cable is connected to a PC USB port. The COM port number is verified in Windows Device Manager under 'Ports (COM and LPT)' as 'USB Serial Port (COMX)'. If automatic installation fails, manual installation and driver update procedures are provided.
- Automatic installation requires Internet connection
- Verify installation in Device Manager: Ports (COM and LPT) → USB Serial Port (COMX)
- COM port number is required for software configuration
- After driver installation, communication works without Internet
- Manual driver download: AP, AT-R, MOC63u from https://www.an.shimadzu.co.jp/balance/products/driver.htm
- Manual driver download: AU, AT, UW/UX/UP, TW/TX/TXB, BL, ELB, BW-K/BX-K from https://www.buffalo.jp/support/download/detail/?dl_contents_id=62142
- Driver folder names: AP/AT-R/MOC63u → 'cdm21228'; other series → 'XP32bit_2000'
- If 'USB Serial Port' shows '!' and/or '?' symbol, repeat driver update from step 4-2
The software is downloaded free from the Shimadzu website and decompressed from 'Multi-Balance_Collect.zip'. Installation may be blocked by PC security settings or other software; check with IT staff.
- Download from https://www.shimadzu.com/an/products/balances/balances-software/mbc/mbc_dl.html
- Decompress 'Multi-Balance_Collect.zip'
- If installation fails, check PC security settings and other software interference
Connect each balance one at a time and note the COM port number assigned in Device Manager. Start 'Multi-Balance_Collect.xltm' in Excel, open Detailed Settings, and configure the number of connected balances (up to 4) and per-balance settings on tabs A–D. Communication specifications must match the balance settings.
- Connect balances one at a time to note COM port numbers
- Use a USB hub if not enough USB ports
- Detailed Settings tabs A–D correspond to Balance A–D
- Communication specification defaults: 1200 baud, None parity, 8 data bits, 1 stop bit, Handshake OFF, Delimiter CR
- Clicking Reset restores all communication items to default values
- Only connectable COM port numbers are displayed in the pulldown
Detailed Settings include Stability Information ON/OFF, Trimming (removes units and brackets from type-approval balances), PC date/time synchronization (max 4 balances, limited model support), Menu Lock (AP Series only), and Output Interval for continuous collection (0:01 to 99:59).
- Stability Information: records 'S' (stable) or 'U' (unstable)
- Trimming: OFF by default; ON removes units and '[]' brackets; data cannot be used for transactions or proof/certification on type-approval balances
- Time Synchronization: synchronizes PC and balance dates/times; not available on series other than AT/AT-R, TW/TX/TXB, BL, ELB; seconds not displayed for UW/UX/UP, BW-K/BX-K, MOC63u
- Menu Lock: AP Series only; locks keys other than PRINT, O/T, ION, ↑, ←, →
- Output Interval: 0 min 1 s to 99 min 59 s; default 0 min 1 s; cannot be set to 0 min 0 s; only enabled for continuous data collection
- Data collection interval may not be achievable depending on number of balances and baud rate
After saving the Excel file, a Communication test screen appears. Press the PRINT key on Balance A; if connection information, COM port number, and display value match the connected balance, communication is correct. The test screen also appears automatically when OK is clicked on Detailed Settings or when a saved file is opened.
- Press PRINT key on Balance A to test
- Verify connection information (Balance A–D), COM port number, and display value
- If data not received, re-verify COM port and communication settings per sections 6-1 and 6-3 item 2
- Test can be skipped by clicking OK if connection and settings unchanged
Before data collection, set balance automatic calibration OFF, GLP print OFF, and configure the PC to never enter sleep mode. Data can be output by clicking individual Output buttons, clicking Start for continuous collection at set intervals, or pressing PRINT on the balance.
- Set balance automatic calibration to OFF (MBC cannot collect during calibration)
- Set balance GLP print to OFF (GLP print increases output time and prevents interval collection)
- Set PC power options so it never sleeps (sleep interrupts USB power)
- Output methods: individual Output button, Start button for continuous collection, or PRINT key on balance
- MOC63u output method differs; refer to MOC63u instruction manual
- Save a copy of the Excel file before data collection to reuse settings
Key Integration, Operating & Safety Requirements
| Specification / Condition | Engineering Implication | Applies To |
|---|---|---|
| Communication specifications: baud rate 1200, parity None, data bit 8, stop bit 1, handshake OFF, delimiter CR. p.3 ↗ | Balance and software must be set to these exact values for data output to work. | All supported balance models |
| The 'Linked to Windows' balance communication setting must be disabled. p.3 ↗ | If left enabled, data may not be output correctly. | All supported balance models |
| AP, AT-R, MOC63u require USB cable set P/N S321-71730-41. p.3 ↗ | Use only this genuine cable set for these models to ensure correct data output. | AP, AT-R, MOC63u |
| AT, BL, ELB require both USB-serial adapter P/N S321-62520 and I/O-RS converter cable P/N S321-71260-01. p.3 ↗ | Both parts are mandatory; using only one will not establish communication. | AT, BL, ELB Series |
| TW/TX/TXB require USB-Serial conversion kit P/N S321-62520-45. p.3 ↗ | Use this specific kit for TW/TX/TXB models. | TW/TX/TXB Series |
| AU, UW/UX/UP, BW-K/BX-K require USB-Serial conversion kit P/N S321-62520-01. p.3 ↗ | Use this specific kit for these models. | AU, UW/UX/UP, BW-K/BX-K Series |
| MOC120H can only be used with Balance keys. p.3 ↗ | MOC120H is not supported by Multi-Balance Collect cable connection. | MOC120H |
| Up to 4 balances can be connected and synchronized for date/time. p.10 ↗ | Plan USB port availability; use a USB hub if needed. | All supported balance models |
| Time Synchronization function is not available on series other than AT/AT-R, TW/TX/TXB, BL, and ELB. p.14 ↗ | For other series, set date/time manually on each balance. | All except AT/AT-R, TW/TX/TXB, BL, ELB |
| Seconds are not displayed for UW/UX/UP, BW-K/BX-K Series, and MOC63u. p.14 ↗ | Time synchronization accuracy is limited to minutes for these models. | UW/UX/UP, BW-K/BX-K, MOC63u |
| Menu Lock function is only available on the AP Series. p.14 ↗ | Leave Menu Lock OFF for all non-AP series. | AP Series only |
| Output Interval can be set from 0 minutes 1 second to 99 minutes 59 seconds; default is 0 minutes 1 second; 0 minutes 0 seconds cannot be set. p.15 ↗ | Minimum continuous collection interval is 1 second. | All supported balance models |
| Trimming is OFF by default; when ON, units and '[]' brackets are removed from collected values. p.14 ↗ | Enables easier graphing and calculation, but data cannot be used for transactions or proof/certification on type-approval balances. | Type approval model balances |
| Balance automatic calibration must be set to OFF before data collection. p.17 ↗ | MBC data collection cannot be carried out during calibration. | All supported balance models |
| Balance GLP print setting must be set to OFF before data collection. p.17 ↗ | GLP print increases output time and prevents collection at set intervals. | All supported balance models |
| PC must be configured to never enter sleep mode. p.17 ↗ | Sleep mode interrupts USB port power and stops data collection. | All supported balance models |
| Driver folder names: AP/AT-R/MOC63u use 'cdm21228'; other series use 'XP32bit_2000'. p.7 ↗ | When manually updating the driver, browse to the correct folder for the balance model. | AP, AT-R, MOC63u vs. other series |
| USB driver automatic installation requires the PC to be connected to the Internet. p.5 ↗ | After installation, communication works without Internet. | All supported balance models |
Faults & Troubleshooting
| Code / Symptom | Possible Cause | Remedy | Note |
|---|---|---|---|
| USB Serial Port (with a ! and/or ? symbol) p.6 ↗ | USB driver has not been installed correctly. | Reinstall the driver according to procedures 4-1 and 4-2; repeat from step 4-2 if symbol persists. | Section 4 / p.6; Section 4-2 / p.8 |
| USB HS Serial Converter (with a ! and/or ? symbol) p.6 ↗ | USB driver has not been installed correctly for models other than AP, AT-R Series, and MOC63u. | Reinstall the driver according to procedures 4-1 and 4-2. | Section 4 / p.6 |
| Measurement screen (Figure 8) is not displayed, and a message is displayed in Excel (Figure 7) p.10 ↗ | Excel security message blocking content. | Click 'Enable contents' (A) outlined in red to close the message and display the Measurement screen. | Section 6-2 / p.10 |
| Data is not being received correctly during communication test p.16 ↗ | COM port number or communication settings are incorrect. | Refer to 6-1 and 6-3 item 2), verify COM port number and communication settings, and reconfigure if necessary. | Section 7 / p.16 |
| Data cannot be collected at the set intervals p.17 ↗ | Balance GLP print setting is ON, increasing output time. | Set the balance GLP print setting to OFF. | Section 9 / p.17 |
| Data cannot be collected with MBC p.17 ↗ | PC entered sleep mode, interrupting power supply to USB ports. | Configure PC power options so it does not enter sleep mode (set 'PC goes to sleep after' to 'Never'). | Section 9 / p.17 |
| MBC data collection cannot be carried out during calibration p.17 ↗ | Balance automatic calibration is ON. | Set the balance automatic calibration setting to OFF. | Section 9 / p.17 |