LC-Mikros Instruction Manual
Technical Reference Summary
Covered Models & Document Scope
This manual (228-92837) covers the Nexera Mikros LC-Mikros high-pressure gradient solvent delivery module for Shimadzu HPLC systems, including standard component parts, accessories, optional parts (FCV-11AL/FCV-11ALS reservoir switching valves, DGU-403/DGU-405 degassing units, CBM-40lite controller, maintenance kit, tool kit), operation, troubleshooting, hardware validation, maintenance, and technical specifications.
Section-by-Section Technical Breakdown
The LC-Mikros is a high-pressure gradient solvent delivery module for micro to semi-micro flow rates. It provides stable gradient delivery from 1 to 500 µL/min and high-pressure operation up to 80 MPa, with automatic compressibility measurement/compensation and sample injection pressure drop recovery when used with SIL-40 series autosamplers.
- Flow rate range: 1 to 500 µL/min
- Maximum pressure: 80 MPa
- Automatic compressibility measurement and compensation
- Sample injection pressure drop recovery function (requires SIL-40 series autosampler)
- Synchronized pump control at analysis start for improved retention time stability (requires SIL-40 series autosampler)
Lists the standard parts, accessories, and optional parts with part numbers. The instrument is installed by Shimadzu service personnel. Optional parts include reservoir switching valves, degassing units, a system controller, and maintenance/tool kits.
- Standard accessories include syringe, optical cable, clamp, spiral tube wrap, pipe clip, drain parts, silicone tubing (I.D. 7 mm, O.D. 10 mm, 1 m), remote cables (2 pcs).
- Optional reservoir switching valve FCV-11AL (228-45048-58) for three flow lines; FCV-11ALS (228-45049-58) for one solvent delivery module.
- Optional degassing units DGU-403 (228-65088-58) and DGU-405 (228-65089-58).
- CBM-40lite (228-45011-58) can be installed inside the instrument; dedicated cable 228-70425-42 required.
- Maintenance kit for LC-Mikros: 228-70424-41; Tool kit: 228-70420-41.
Describes the front cover, internal components behind the front cover, left side and base panel, back panel connectors, and the operation panel including display and keypad. Key components include primary/secondary pump heads, dynamic inlet valves, line filters, seal rinsing pump, check valves, pressure sensors, purge valve, and degassing unit.
- Operation panel: press anywhere on screen except display to show operation keys.
- Power button: press and hold at least 3 seconds to turn ON; to turn OFF, press and hold at least 3 seconds, then press again at least 1 second.
- Status indicator: green = analysis ready, red = error, orange = sleep.
- Back panel connectors: [REMOTE], External Input/Output terminals, [DGU PRESS IN], [DGU/SV], [PC], main power switch, power cord connector.
- Display panel shows flow A/B (µL/min), pressure A/B, PUMP, CONC, RUN, PURGE indicators.
Covers preparing mobile phase and seal rinse solution, filling the seal rinse line, and checking pressure limits before operation. Seal rinse solution is 10% 2-propanol; consumption is 3 mL/hour when pumping and 5 mL when power is turned on. Maximum and minimum pressure limits must be set appropriately.
- Seal rinse solution: 10% 2-propanol in reservoir bottle.
- Seal rinse solution consumption: 3 mL/hour during pumping; 5 mL when power is turned on.
- When main power is turned ON, seal rinsing pump runs automatically for approximately 3 minutes.
- Purge line selection: 1 = pump A and B lines, 2 = only pump A line, 3 = only pump B line.
- Purging stops automatically 6 minutes after it begins; duration can be changed with [PURGE TIME] auxiliary function.
- Maximum allowable pressure of instrument is 80 MPa; set P.MAX according to component with lowest maximum pressure limit.
- P.MIN function is not activated for first 5 minutes of pumping; setting P.MIN to 0 disables automatic stop and alarm.
Describes screen types (monitor, auxiliary functions, VP functions), monitoring screens, auxiliary function parameters, VP functions, connection with system controller or workstation, and external input/output terminals. The operation panel has arrow key input mode and numerical key input mode, switched by pressing [Enter].
- Three main screens: monitor screen, auxiliary functions (FUNCTION) screen, VP function screen.
- Auxiliary functions include parameter settings, control settings, system settings, utility settings.
- VP functions include product information, maintenance information, validation support, calibration support.
- External input/output terminals and remote cable connection described.
- Connection to system controller or workstation via [REMOTE] connector.
Provides tables of symptoms, probable causes, and corrective actions for common problems such as power failure, no liquid pumping, unstable pumping, flow rate deviation, retention time irreproducibility, pressure issues, ghost peaks, and baseline drift. Also lists error messages with type (Warning, Alarm, Fatal), cause, and action.
- Symptom table covers power, pumping, pulsation, flow rate, retention time, pressure, ghost peaks, baseline drift.
- Error messages include WARNING TOUCH KEY UNSTABLE, WARN: DEGAS PRS, WARN: MEAS COMP A/B, WARN: LOST PARAM, WARN: BACKUP LOST, WARN: NO PAIRING, ERROR P-MAX, ERROR P-MIN, ERROR A-HOME POS, ERROR B-HOME POS, ERROR LEAK DETECT, ERROR EXTERNAL, ROM FAILURE, RAM FAILURE, ERROR FAN, ERROR LEAK SENS, ERROR UNUSABLE PANEL.
- Warning: instrument does not stop; pressing [CE] clears message.
- Alarm: instrument stops; pressing [CE] clears message and stops alarm sound.
- Fatal: instrument stops; pressing [CE] clears alarm sound but not message; power button restarts instrument.
Covers hardware validation overview, types, periodic validation, daily inspection, inspection at maintenance, precautions, and validation of the pump including ROM/RAM self diagnosis, firmware version check, flow rate accuracy test, and leak sensor test.
- Validation types: periodic validation, daily inspection, inspection at maintenance.
- Pump validation includes ROM, RAM self diagnosis, firmware version check, flow rate accuracy test, leak sensor test.
- Precautions at inspection described.
Describes periodic inspection and maintenance, replacement of plunger seal, cleaning and inspecting (replacing) plungers, replacing rinse seal, cleaning dynamic inlet valves and check valves, cleaning/replacing dynamic inlet valve cartridge ASSY in ultrasonic bath, replacing check valves, replacing line filter, cleaning/inspecting (replacing) suction filters, replacing dust filter, exterior cleaning, and adjusting leak sensor sensitivity.
- Periodic inspection and maintenance list provided.
- Plunger seal replacement procedure includes removing pump head, replacing seal, installing pump head, and check after replacement.
- Cleaning dynamic inlet valves and check valves with 2-propanol.
- Dynamic inlet valve cartridge ASSY can be cleaned in ultrasonic bath or replaced.
- Line filter and suction filters can be cleaned in ultrasonic bath or replaced.
- Leak sensor sensitivity adjustment described.
Provides installation site requirements, wiring, plumbing, specifications for LC-Mikros and high-pressure gradient system, and lists of consumable parts and maintenance kit.
- Installation site, wiring, and plumbing requirements.
- LC-Mikros specifications and high-pressure gradient system specifications.
- Consumable parts and maintenance kit part numbers.
Key Integration, Operating & Safety Requirements
| Specification / Condition | Engineering Implication | Applies To |
|---|---|---|
| Flow rate range is 1 to 500 µL/min. p.9 ↗ | Select this module for micro to semi-micro LC applications within this flow range. | LC-Mikros |
| Maximum pressure is 80 MPa. p.9 ↗ | Do not exceed 80 MPa; when connecting to other equipment, set P.MAX according to the component with the lowest maximum pressure limit. | LC-Mikros |
| Seal rinse solution is 10% 2-propanol; consumption is 3 mL/hour when pumping and 5 mL when power is turned on. p.11 ↗ | Prepare sufficient seal rinse solution and monitor consumption for long runs. | LC-Mikros |
| When main power is turned ON, the seal rinsing pump automatically runs for approximately 3 minutes. p.12 ↗ | Allow about 3 minutes after power-on for the seal rinse line to fill before starting operation. | LC-Mikros |
| Purging stops automatically 6 minutes after it begins; duration can be changed using [PURGE TIME] auxiliary function. p.10 ↗ | Default purge duration is 6 minutes; adjust [PURGE TIME] if longer purging is needed. | LC-Mikros |
| P.MIN function is not activated for first 5 minutes of pumping. p.56 ↗ | Do not rely on P.MIN to detect low pressure during the first 5 minutes after pumping starts. | LC-Mikros |
| Setting P.MIN to 0 disables automatic stop and alarm on pressure drop. p.15 ↗ | Use P.MIN = 0 only when pressure-drop protection is not desired. | LC-Mikros |
| If pressure exceeds P.MAX, pumping automatically stops and ERROR P-MAX is displayed; if [SYSTEM PROTECT] is activated, pumping continues at half flow rate. p.55 ↗ | Set P.MAX appropriately; activate SYSTEM PROTECT if continued operation at reduced flow is preferred over stop. | LC-Mikros |
| For mobile phase mixtures containing acids such as trifluoroacetic acid or formic acid at high pressures (generally above 70 MPa), if leakage occurs between pump head and seal holder even after replacing plunger seal, replace plunger seal, backup ring, and pump head. p.51 ↗ | At >70 MPa with acid-containing mobile phases, plan for pump head replacement if seal replacement alone does not stop leakage. | LC-Mikros |
| To clean flow lines upstream of column for ghost peaks: deliver 17% phosphoric acid solution at 0.5 mL/min for 30 minutes, then purified water at 0.5 mL/min for 30 minutes. p.53 ↗ | Use this procedure to remove contamination causing ghost peaks. | LC-Mikros |
| Power button: press and hold at least 3 seconds to turn ON; to turn OFF, press and hold at least 3 seconds, then press again at least 1 second. p.8 ↗ | Follow this sequence to avoid improper shutdown. | LC-Mikros |
| Status indicator: green = analysis ready, red = error, orange = sleep. p.8 ↗ | Use indicator color for quick status check. | LC-Mikros |
| Silicone tubing supplied: I.D. 7 mm, O.D. 10 mm, 1 m length. p.10 ↗ | Use specified tubing dimensions for drain connections. | LC-Mikros |
| Optional reservoir switching valve FCV-11AL (228-45048-58) supports three flow lines; FCV-11ALS (228-45049-58) supports one solvent delivery module. p.11 ↗ | Select FCV-11AL for up to three modules, FCV-11ALS for single-module automatic switching. | LC-Mikros with optional FCV valves |
| CBM-40lite (228-45011-58) can be installed inside the instrument; dedicated cable 228-70425-42 is required. p.11 ↗ | Order the dedicated cable when installing CBM-40lite internally. | LC-Mikros with optional CBM-40lite |
| Sample injection pressure drop recovery and synchronized pump control at analysis start are available only when using the instrument with SIL-40 series autosampler. p.9 ↗ | These performance features require SIL-40 series autosampler compatibility. | LC-Mikros with SIL-40 series autosampler |
Faults & Troubleshooting
| Code / Symptom | Possible Cause | Remedy | Note |
|---|---|---|---|
| Even if the main power switch is turned ON, power is not applied. p.57 ↗ | Power plug is disconnected. | Connect the plug correctly. | |
| Even if the main power switch is turned ON, power is not applied. p.57 ↗ | Power cord is disconnected internally. | Replace with a new power cord of the same type. | |
| Even if the main power switch is turned ON, power is not applied. p.57 ↗ | Power supply does not meet the specifications for the instrument. | Use the power supply that meets the specifications for the instrument. | |
| Even if (power button) is pressed, power is not applied. p.57 ↗ | The main power is turned to OFF. | Turn ON the main power switch located at the back of the instrument. | 2.4 Back (P.7) |
| No liquid is pumped. (The instrument does not run.) p.57 ↗ | [PUMP] was not pressed. "PUMP" is not highlighted on the display panel. | Press [PUMP]. "PUMP" is highlighted on the display panel, and the pump becomes operational. | |
| No liquid is pumped. (The instrument does not run.) p.57 ↗ | Flow rate is set to "0". | Set a flow rate other than 0. | |
| No liquid is pumped. (The instrument does not run.) p.57 ↗ | Error message (P.MAX, P.MIN etc.) appears. | Press [CE] to clear the message and make corrective action for errors. | |
| The instrument is running, but no liquid is pumped. p.57 ↗ | Air bubbles are generated inside the pump head. | Press [PURGE] to purge out the bubbles. | |
| The instrument is running, but no liquid is pumped. p.57 ↗ | Air bubbles are produced through the suction filter and pump inlet. | Firmly connect the suction filter fitting. | |
| Pumping is unstable, with large pump pulsation. p.57 ↗ | Air bubbles are generated inside the pump head. | Press [PURGE] to purge out the bubbles. | |
| Pumping is unstable, with large pump pulsation. p.57 ↗ | Previous mobile phase is still remaining inside the pump head. | Press [PURGE], and thoroughly purge the old mobile phase. | |
| Pumping is unstable, with large pump pulsation. p.58 ↗ | Air bubbles are generated in the suction filter tubing. | Press [PURGE], and thoroughly purge the old mobile phase. Shake the suction filter to drive out the bubbles. If the suction filter is clogged, clean it in an ultrasonic bath. Otherwise, replace it. | 7.9 Cleaning and Inspecting (Replacing) Suction Filters (P.97) |
| Pumping is unstable, with large pump pulsation. p.58 ↗ | Degas the mobile phase. | Degas the mobile phase. | |
| Pumping is unstable, with large pump pulsation. p.58 ↗ | The check valve is fouled. | Pump 2-propanol to clean the check valve. If cleaning the check valve is not effective, replace it. | 7.5 Cleaning Dynamic Inlet Valves and Check Valves (P.89); 7.6 Cleaning and Inspecting (Replacing) Dynamic Inlet Valve C |
| Pumping is unstable, with large pump pulsation. p.58 ↗ | The dynamic inlet valve cartridge is dirty. | Pump 2-propanol to clean the dynamic inlet valve cartridge. If cleaning is not effective, clean the dynamic inlet valve cartridge in an ultrasonic bath or replace it. | 7.5 Cleaning Dynamic Inlet Valves and Check Valves (P.89); 7.6 Cleaning and Inspecting (Replacing) Dynamic Inlet Valve C |
| Pumping is unstable, with large pump pulsation. p.59 ↗ | Liquid is leaking from the gap between the pump head and seal holder, or from the rinse solution outlet. | Replace the plunger seal and the backup ring. If replacing the plunger seal and backup ring do not stop the leakage, replace the plunger. If analytical conditions require pumping mobile phase mixtures containing acids such as trifluoroacetic acid or formic acid at high pressures (in general, above 70 MPa) and liquid leaks from between the pump head and seal holder even after replacing the plunger seal, then replace the plunger seal, backup ring, and pump head. | 7.2 Replacement of Plunger Seal (P.66); 7.3 Cleaning and Inspecting (Replacing) Plungers (P.80) |
| Pumping is unstable, with large pump pulsation. p.59 ↗ | Liquid is leaking from flow line connections. | Retighten the male nuts. Otherwise, replace the tubing, nuts, and ferrules. | |
| Pumping is unstable, with large pump pulsation. p.59 ↗ | Flow line is (partially) clogged. | Clean the line filter in an ultrasonic bath or replace it. Identify the clogged parts and replace it. | 7.8 Replacing the Line Filter (P.95) |
| Plunger seals wear out too quickly. p.59 ↗ | Plunger seals wear out too quickly. | Replace the plunger. | 7.3 Cleaning and Inspecting (Replacing) Plungers (P.80) |
| Flow rate is below the set value. p.59 ↗ | The check valve is not working properly. | If cleaning the check valve is not effective, replace it. | 7.5 Cleaning Dynamic Inlet Valves and Check Valves (P.89); 7.7 Replacing Check Valves (P.92) |
| Flow rate is below the set value. p.59 ↗ | The dynamic inlet valve does not work properly. | If cleaning is not effective, clean the dynamic inlet valve cartridge in an ultrasonic bath or replace it. | |
| Flow rate is below the set value. p.59 ↗ | The suction filter is clogged. | Clean the suction filter in an ultrasonic bath. Otherwise, replace it. | 7.9 Cleaning and Inspecting (Replacing) Suction Filters (P.97) |
| Flow rate is below the set value. p.60 ↗ | Liquid is leaking from the gap between the pump head and seal holder, or from the rinse solution outlet. | Replace the plunger seal and the backup ring. If replacing the plunger seal and backup ring do not stop the leakage, replace the plunger. If analytical conditions require pumping mobile phase mixtures containing acids such as trifluoroacetic acid or formic acid at high pressures (in general, above 70 MPa) and liquid leaks from between the pump head and seal holder even after replacing the plunger seal, then replace the plunger seal, backup ring, and pump head. | 7.2 Replacement of Plunger Seal (P.66); 7.3 Cleaning and Inspecting (Replacing) Plungers (P.80) |
| Flow rate is below the set value. p.60 ↗ | Liquid is leaking from flow line connections. | Retighten the male nuts. Otherwise, replace the tubing, nuts, and ferrules. | |
| Retention time cannot be reproduced. p.60 ↗ | The check valve is not working properly. | Pump 2-propanol to clean the check valve. If cleaning the check valve is not effective, replace it. | 7.5 Cleaning Dynamic Inlet Valves and Check Valves (P.89); 7.7 Replacing Check Valves (P.92) |
| Retention time cannot be reproduced. p.60 ↗ | The dynamic inlet valve does not work properly. | Pump 2-propanol to clean the dynamic inlet valve cartridge. If cleaning is not effective, clean the dynamic inlet valve cartridge in an ultrasonic bath or replace it. | 7.9 Cleaning and Inspecting (Replacing) Suction Filters (P.97); 7.7 Replacing Check Valves (P.92) |
| In a high-pressure gradient system, the 2 pump units show different pressures. p.60 ↗ | Zero points of the pressure sensors of 2 pump units are not the same. | Use [ZERO ADJUST] auxiliary function to execute zero adjustment of the pressure sensors. | 4.3.5 System Settings Group (P.27) |
| In a high-pressure gradient system, the 2 pump units show different pressures. p.60 ↗ | Line filter of one of the pump units is clogged. | Inspect or replace the line filter. | 7.8 Replacing the Line Filter (P.95) |
| In a high-pressure gradient system, the 2 pump units show different pressures. p.60 ↗ | Flow line(s) is/are clogged at the junction of the 2 pump unit flow lines. | Identify the clogged parts and replace it. | |
| The pressure does not rise. p.60 ↗ | Liquid is leaking from flow line connections. | Retighten the male nuts. Otherwise, replace the tubing, nuts, and ferrules. | |
| Pressure rises too much. (Remove the column to check.) p.60 ↗ | The line filter is clogged. | Inspect or replace the line filter. | |
| Pressure rises too much. (Remove the column to check.) p.60 ↗ | Flow lines are clogged. | Identify the clogged parts and replace it. | |
| Pressure rises too much. (Remove the column to check.) p.61 ↗ | Inner diameter of tubings is too small. | Use tubings with the specified inner diameter. | 7.8 Replacing the Line Filter (P.95) |
| Ghost peaks appear. p.61 ↗ | Flow lines, line filter or suction filter are fouled. | Follow the procedure below to rinse the flow lines upstream of the column. (1) Deliver 17% phosphoric acid solution through the flow lines at 0.5 mL/min for 30 minutes. (2) Deliver purified water through the flow lines at 0.5 mL/min for 30 minutes. | |
| Baseline drifts. p.61 ↗ | |||
| WARNING TOUCH KEY UNSTABLE p.62 ↗ | As the key recognition of the operation unit is unstable, the key operation is temporarily disabled. | This is displayed when a large intensity is added to the operation unit, in which no operation is accepted for a specific time (30 seconds at maximum). Leave the hand from the operation unit, and wait for a while. | Warning |
| WARN: DEGAS PRS p.62 ↗ | Vacuum pressure of degassing unit connected to this instrument or the external degassing unit has been abnormal for a certain time. | Check the vacuum pressure of degassing unit. If the vacuum pressure is "NG" (error), turn the power OFF once and turn it ON again. If the vacuum pressure is "OK", it is of no matter, for the degassing unit is returning to run automatically. | Warning; Showing the vacuum pressure of internal degassing unit [uDEGAS PRESS] (P.18) |
| WARN: MEAS COMP A p.63 ↗ | The compressibility cannot be measured. | Check whether the mobile phase is empty, and perform purging again. If the warning message is still displayed, replace the primary side cartridge ASSY or the plunger seal. | Warning |
| WARN: MEAS COMP B p.63 ↗ | The compressibility cannot be measured. | Check whether the mobile phase is empty, and perform purging again. If the warning message is still displayed, replace the primary side cartridge ASSY or the plunger seal. | Warning |
| WARN: LOST PARAM p.63 ↗ | Parameters may have been lost. | Parameter resetting is required. Contact your Shimadzu representative. | Warning |
| WARN: BACKUP LOST p.63 ↗ | The preliminary data of the instrument is either damaged or lost. | The instrument can be used as it is, but the backup function of the instrument is failed, and the data recovery function when the data is lost is not usable. Contact your Shimadzu representative. | Warning |
| WARN: NO PAIRING p.63 ↗ | Pairing settings have not been configured between the instrument and operation panel. | Operations can be performed even if this message is displayed, but a malfunction has occurred with the instrument data backup function, so the recovery function is not available whenever data is lost. Contact your Shimadzu representative. | Warning |
| ERROR P-MAX p.63 ↗ | Pump discharge pressure has exceeded the set max. pressure limit, [P.MAX]. Pumping will automatically stop due to the limiter function. If [SYSTEM PROTECT] auxiliary function has been activated, pumping will continue at half flow rate. | Inspect the clogging in the flow line. If no clogging is found, set the appropriate [P.MAX] value or decrease the flow rate. | Alarm; Setting the maximum pressure limit during purging [P.MAX] (P.23); Setting the system protection [SYSTEM PROTECT] |
| ERROR P-MIN p.64 ↗ | Pump discharge pressure has fallen below the set min. pressure limit, [P.MIN]. Pumping will automatically stop due to the limiter function. | Inspect the leakage in the flow line. If no leakage is found, set the appropriate [P.MIN] value. Note that [P.MIN] function is not activated for first 5 minute of pumping. | Alarm; Setting the minimum pressure limit during purging [P.MIN] (P.23) |
| ERROR A-HOME POS p.64 ↗ | The motor home position cannot be detected, the motor does not run, or the motor is slipping. | Turn the power OFF and ON again, and press [PUMP]. If this message still appears, turn OFF the main power and contact your Shimadzu representative. | Alarm |
| ERROR B-HOME POS p.64 ↗ | The motor home position cannot be detected, the motor does not run, or the motor is slipping. | Turn the power OFF and ON again, and press [PUMP]. If this message still appears, turn OFF the main power and contact your Shimadzu representative. | Alarm |
| ERROR LEAK DETECT p.64 ↗ | The leak sensor has detected leakage. | Inspect the leakage in the flow line. Wipe away any liquid around the leak sensor. | Alarm |
| ERROR EXTERNAL p.64 ↗ | An error signal was sent from the external device connected to the external input/output terminal [ERR]. | Inspect the external device, and eliminate the cause of the error. | Alarm |
| ROM FAILURE p.64 ↗ | ROM has an error. | Turn OFF the main power and contact your Shimadzu representative. | Fatal |
| RAM FAILURE p.64 ↗ | RAM has an error. | Turn OFF the main power and contact your Shimadzu representative. | Fatal |
| ERROR FAN p.64 ↗ | An error has occurred with cooling fan operation. | Turn OFF the main power and contact your Shimadzu representative. | Fatal |
| ERROR LEAK SENS p.64 ↗ | There is an error with the value detected by the leak sensor. | Turn OFF the main power and contact your Shimadzu representative. | Fatal |
| ERROR UNUSABLE PANEL p.64 ↗ | An error has been detected in the communication between the instrument and the operation panel. | Turn OFF the main power and contact your Shimadzu representative. | Fatal |