Instruction Manual - PC Control Software
Technical Reference Summary
Covered Models & Document Scope
SMV-301 PC Control Software (version documented 2015–2020) for operating the Automatic Mooney Viscometer SMV-301 and SMV-301RT. Test modes: Mooney test, Mooney scorch test, Stress relaxation test (SMV-301RT only), Stress relaxation cycle test (SMV-301RT only). Method file extensions: *.msmv (SMV-301), *.msmr (SMV-301RT), *.msm* (all).
Section-by-Section Technical Breakdown
Describes installing the SMV-301 software from the Setup CD-ROM, selecting an installation folder, and completing setup. Also describes uninstalling the software via Windows Control Panel and a power-management caution to prevent communication loss.
- Insert the SMV-301 Setup CD-ROM; setup starts automatically, otherwise run [SMV.vbs].
- Installation folder can be changed via [Reference] / [Browse Folder].
- Uninstall via [Start]-[Control Panel]-'Uninstall a program', select 'SMV-301'; PC must start with administrator authority.
- Set [System standby] and [Turn off hard disks] to [Never] to avoid losing communication between PC and testing machine.
Covers enabling external control on the testing machine touch panel, setting a 4-digit ID number, connecting via RS-232C cable, and registering Hardware IDs in the software. Up to four testing machines can be connected.
- On [Initial condition] screen press [Device], set External Control from [off] to [on].
- Press [ID No.] and enter any 4-digit number; ID No. cannot be changed while external control connection is established.
- Connect testing machine and PC using RS-232C cables.
- In [System Setting] dialog, set each Hardware ID connected to a testing machine; up to four testing machines.
- Main Window shows status of each testing machine for which a connection port is specified.
Defines the settable ranges and increments for test conditions and the four test modes. Includes sample/text field limits, temperature, preheating/after time, Mooney set time, scorch points, rotor speed, pass/fail limits, and result display digit settings.
- Test temperature: 200 °C max, minimum increment 0.1 °C (setting range 0–200 °C).
- Preheating time and after time: 0–99 min 59 s, minimum increment 1 s; for stress relaxation measurement preheating and after time are 0–5 minutes respectively.
- Mooney set time: 0–99 min 59 s, minimum increment 1 s, 10 points max.
- Scorch point: 0–99.9 M, minimum increment 0.1 M, 10 points max.
- Rotor speed: 0.209 rad/s (2 rpm) for SMV-301; 0.0104–2.092 rad/s (0.1–20 rpm) for SMV-301RT.
- Pass/fail: Mooney value 1–10, Scorch time 1–10.
- Sample name/remarks: 25 two-byte characters (50 one-byte); testing machine accepts up to 20 alphanumeric digits, transmission stops at first two-byte character.
- Result display digits adjustable (e.g., Mooney viscosity XXX.X M, 0–2 decimals; stress relaxation MSR X.XXX, 0–4 decimals).
Describes the Mooney viscosity test of raw or compounded rubber: sample chamber temperature control, rotor rotation after preheating, and reaction torque measurement. Up to ten Mooney viscosity detection time points can be set.
- Test temperature set in range 0–200 °C; example set value 100 °C.
- Preheating time is from upper hot plate lowering to rotor start; example 1 minute.
- After time is pressure-keeping time from test completion to upper hot plate rise; example 0 minutes.
- Testing standards require Mooney viscosity at 4 minutes after preheating, but machine allows up to 10 detection time points.
Describes the scorch test for compounded rubber, measuring scorch time from minimum viscosity to a predetermined viscosity increase (scorch point). Up to ten arbitrary scorch points can be set.
- Scorch point prescribed as 5 M in standards, but machine allows up to ten arbitrary scorch points up to 99.9 minutes.
- Peak value Vp and minimum Mooney value Vm can be measured.
- Test temperature range 0–200 °C; example 100 °C.
- Preheating time example 1 minute; after time example 0 minutes.
Covers stress relaxation measurement per ISO 289-4 or ASTM D 1646, including MSR, standard error, intercept, slope, correlation, area, and decay time/torque attenuation. Stress relaxation measurement is made during after time.
- ISO 289-4 uses data 1.6 to 5.0 seconds after rotor stops to determine MSR and standard error.
- ASTM D 1646 uses data from 1 second after rotor stops to end of after time to determine intercept (k), slope (a), correlation (r), and area (A).
- Decay tX: time for torque to drop by x% from 100% torque at rotor stop.
- Decay Xt: torque attenuation rate at t seconds after rotor stop based on 100% torque at rotor stop.
- Preheating time and after time are 0–5 minutes respectively for stress relaxation measurement.
Describes cycle tests combining Mooney tests and stress relaxation measurements in up to 10 steps, with maximum 5 stress relaxation measurements and total measurement time up to 30 minutes. Scorch test cannot be set.
- Any combination of Mooney test and stress relaxation measurement, 10 steps max; maximum stress relaxation measurements is 5.
- Maximum total time period for a cycle test is 30 minutes.
- A scorch test cannot be set.
- No stress relaxation measurement is made during after time.
- Number of cycle settable 1–10; example set value 4.
Covers selection of test mode, data display format, sensor information, minimum Mooney detection time, and data processing items such as peak value, minimum Mooney, scorch time, delta Mooney, and pass/fail limits.
- Minimum Mooney detection time initial setting is 5 seconds to avoid unstable data immediately after rotor start.
- After detecting minimum viscosity, the maximum value detected before the minimum is judged as peak value.
- Pass/fail judgment can be enabled by checkbox with upper and lower limits.
- Data processing items include peak value, minimum Mooney, scorch time, delta Mooney, ASTM intercept/slope/correlation/area, decay Xt, decay tX.
Describes the four-step test procedure from power-on through preparation, conducting tests, and saving results. Also covers operations during a test, report printing, and saving in text or PDF format.
- Step 1: Turn on power; Step 2: Prepare for tests; Step 3: Conduct tests; Step 4: Save test results.
- Operations that can be performed during a test are described.
- Result output selection guide and report printing are provided.
- Results can be saved in text format and PDF format.
Covers reanalysis of stored test results, overlaying multiple results, recalculation via Method Wizard, printing, and statistical functions including histograms, control charts, and process capability index.
- Reanalysis can be performed with different analysis settings.
- Multiple sets of test results can be overlaid.
- Method Wizard allows changing various conditions at once (recalculation).
- Statistics include histograms, control charts, and process capability index.
- SPC/SQC data files can be opened.
Describes toggling show/hide states of chart windows and result window, resizing windows, and changing their positions.
- Toggle show/hide states of chart windows and result window.
- Resize chart windows and result window.
- Change positions of chart windows and result window.
Key Integration, Operating & Safety Requirements
| Specification / Condition | Engineering Implication | Applies To |
|---|---|---|
| Test temperature setting range is 0 to 200 °C with minimum increment 0.1 °C. p.39 ↗ | Covers standard Mooney test temperatures up to 200 °C; setpoints must be entered to 0.1 °C resolution. | Whole series (SMV-301 and SMV-301RT) |
| Preheating time and after time range 0–99 min 59 s with minimum increment 1 s; for stress relaxation measurement preheating and after time are 0–5 minutes respectively. p.39 ↗ | Stress relaxation tests are limited to 5-minute preheat and 5-minute after-time windows. | Whole series; stress relaxation limits apply to SMV-301RT |
| Mooney set time range 0–99 min 59 s, minimum increment 1 s, up to 10 points. p.39 ↗ | Up to ten Mooney viscosity detection time points can be programmed per test. | Whole series |
| Scorch point range 0–99.9 M, minimum increment 0.1 M, up to 10 points. p.39 ↗ | Scorch time can be evaluated against up to ten arbitrary viscosity-rise thresholds. | Whole series |
| Rotor speed is fixed at 0.209 rad/s (2 rpm) for SMV-301; SMV-301RT allows 0.0104–2.092 rad/s (0.1–20 rpm). p.39 ↗ | Only SMV-301RT supports variable rotor speed for stress relaxation and cycle tests. | SMV-301 vs SMV-301RT |
| Pass/fail limits: Mooney value 1–10; Scorch time 1–10. p.39 ↗ | Pass/fail judgment can be configured within these numeric ranges. | Whole series |
| Sample name and remarks limited to 25 two-byte characters (50 one-byte); testing machine accepts up to 20 alphanumeric digits and transmission stops at first two-byte character. p.39 ↗ | Use ASCII-only text for fields transmitted to the testing machine to avoid truncation. | Whole series |
| Method file extensions: *.msmv for SMV-301, *.msmr for SMV-301RT, *.msm* for all. p.36 ↗ | Method files must match the testing machine type; the dialog defaults to the currently selected machine type. | Whole series |
| Up to four testing machines can be connected and registered by Hardware ID. p.25 ↗ | Multi-machine setups require unique Hardware ID registration in [System Setting]. | Whole series |
| External control ID No. is any 4-digit number and cannot be changed while connection with external control is established. p.24 ↗ | Disconnect before changing the ID No. on the testing machine. | Whole series |
| PC power settings [System standby] and [Turn off hard disks] must be set to [Never]. p.22 ↗ | Prevents loss of RS-232C communication between PC and testing machine. | Whole series |
| Minimum Mooney detection time initial setting is 5 seconds. p.49 ↗ | Avoids false minimum-viscosity detection from torque fluctuations immediately after rotor start. | Whole series |
| Stress relaxation cycle test: maximum 10 steps, maximum 5 stress relaxation measurements, total measurement time up to 30 minutes; scorch test cannot be set. p.45 ↗ | Cycle test programming must respect the 30-minute total limit and cannot include scorch segments. | SMV-301RT only |
| ISO 289-4 stress relaxation uses data 1.6 to 5.0 seconds after rotor stops; ASTM D 1646 uses data from 1 second after rotor stops to end of after time. p.43 ↗ | After time must be set long enough to cover the ASTM evaluation window. | SMV-301RT only |
| Method file names cannot contain / \ : * ? " < > | characters. p.38 ↗ | Use only valid Windows filename characters when saving methods. | Whole series |
| Default method save folder after installation is (installation drive):\SMV\Methods. p.38 ↗ | Method files are stored under the SMV installation directory by default. | Whole series |
Faults & Troubleshooting
| Code / Symptom | Possible Cause | Remedy | Note |
|---|---|---|---|
| Connection may be lost during communication between the PC and testing machine p.22 ↗ | Depending on the [System standby] and [Turn off hard disks] settings on the PC | Select [Start] - [Setting] - [Control Panel] - [Power Options]. Set the [System standby] and [Turn off hard disks] settings to [Never]. | CAUTION on p.22 |
| The setup program will start automatically. If the setup program does not start p.18 ↗ | Setup program did not auto-run from the SMV-301 Setup CD-ROM | Select [SMV.vbs] on the CD-ROM, and start the program. | How to install the SMV-301 software |