Induztr
Engineering Digest Document Viewer Related Models

c391-e083

Technical Reference Summary

Covered Models & Document Scope

Shimadzu TOC-4200 online TOC analyzer; measurement ranges from 5 mgC/L FS to 50,000 mgC/L FS, optional high-sensitivity range 0 to 1 mgC/L; dilution up to 50x standard; TOC modes NPOC, TC-IC, NPOC+POC (optional), TN (optional); sampling units for single-stream, multi-stream (up to six flow lines), suspended solids, backwash strainer and sample switching; optional web-based monitoring, external I/O, RS-232C, printer, N2 carrier gas, pure water supply, purge, power terminal block, regulator tubing and air duct kits.

Technical Data

Parameter Value Condition
Measurement range 5 to 50,000 mgC/L full-scale p.3 ↗ TOC-4200
High-sensitivity measurement range 0 to 1 mgC/L p.3 ↗ TOC-4200, optional
Dilution factor up to 50x p.3 ↗ TOC-4200, standard dilution function
Number of flow lines up to six flow lines p.4 ↗ multi-stream sample switching unit
Measuring cycle 4 minutes p.6 ↗ short measuring cycle
Width 550 mm p.16 ↗ external dimensions
Depth 388 mm p.16 ↗ external dimensions
Height 214 mm p.16 ↗ external dimensions
Maintenance space over 500 mm p.16 ↗ external dimensions

Operating Limits

  • Measurement range full-scale: 5 to 50,000 mgC/L p.3 ↗
  • High-sensitivity measurement range (optional): 0 to 1 mgC/L p.3 ↗
  • Maximum dilution: 50x p.3 ↗
  • Maximum number of flow lines: 6 p.4 ↗
  • Short measuring cycle: 4 minutes p.6 ↗

Ordering Data

Ordering Code Part No. Variant / Code Applies To Note
PCB retaining fitting, terminal p.12 ↗ 638-84218 Required to expand the analog input/output
PCB retaining fitting, main p.12 ↗ 638-84217 Required to expand boards other than the analog input/output
RS-232C kit p.12 ↗ 638-66228 D-sub 9 pin (male)
Printer kit p.13 ↗ 638-54066 Contains a printer and printer interface
N2 carrier gas kit p.13 ↗ 638-41574 Required to use high-purity nitrogen gas
Automatic pure water supply kit p.13 ↗ 638-57172 Valve set to directly supply purified water
Purge kit p.13 ↗ 638-40252 Plumbing set required to purge the interior
Power terminal block kit p.13 ↗ 638-68162 Required to supply power with a terminal
Regulator tubing kit p.13 ↗ 638-42064 Pipe set to attach the regulator (AW30-0)
Air duct kit p.13 ↗ 638-41204 Contains a pipe and connectors for the duct

Related models