SI 00409

Technical Reference Summary

Covered Models & Document Scope

Application note for the determination of K, Na and Zn in albumin using flame AAS on the Agilent 240 FS double-beam AA spectrometer; covers method conditions, sample preparation and QC results, not a product ordering datasheet.

Technical Data

Parameter Value Condition
Wavelength 589.6 nm p.3 ↗ Na
Bandpass 0.2 nm p.3 ↗ Na
Fuel flow 2 L/min p.3 ↗ Na
Fuel flow 2 L/min p.3 ↗ K
Fuel flow 2 L/min p.3 ↗ Zn
Mode Emission p.3 ↗ Na
D2 correction No p.3 ↗ Na
Replicates 3 p.3 ↗ Na
Replicates 3 p.3 ↗ K
Replicates 3 p.3 ↗ Zn
QC recovery 101.8 % p.4 ↗ Na, QC after calibration