Induztr
Engineering Digest Document Viewer Related Models

Air Cell Mode of the Agilent 9500 ICP-MS

Contents

  1. Overview p. 1
    Introduces Air Cell Mode of the Agilent 9500 ICP-MS
  2. Procedure p. 2
    Shows examples of interference reduction using Air Cell Mode
  3. Procedure p. 3
    Describes suppressing REE2+ and REE-oxide ion interferences
  4. p. 4
    Summarizes simple and powerful interference reduction approach

Engineering Facts

12 facts · every claim linked to a page
Operation

Air cell mode uses ambient laboratory air as the reaction gas, so no external gas cylinder or compressor is required for interference reduction.

Simplifies site requirements and gas infrastructure planning when selecting the instrument.

Source
“Using ambient air from the laboratory, Air cell mode eliminates the need for an external gas supply”
Installation

Atmospheric pressure alone is sufficient to feed air to the cell, so no compressor needs to be installed.

Avoids specifying and installing a compressed air supply during lab setup.

Source
“atmospheric pressure is sufficiently high to provide air without needing a compressor”
Maintenance

An Air cell gas filter removes moisture and hydrocarbons, and an automatic control valve limits air exposure to extend filter service life.

Defines a consumable filter and valve that must be planned for in the maintenance schedule.

Source
“An automatic control valve is installed before the filter to minimize exposure to ambient air, extending its service life.”
Performance

Collision cell mode with KED is less effective against doubly charged ions and can sometimes worsen those interferences, which is why Air cell mode is needed.

Guides method selection for matrices producing doubly charged REE interferences.

Source
“collision cell mode is less effective at suppressing doubly charged ion interferences (M++)”
Operation

A standard multitune method combines AHM and Air cell mode to simultaneously reduce polyatomic and doubly charged interferences across soil, seawater, and food matrices.

Allows a single preset method to cover diverse sample types without external reaction gases.

Source
“The multitune method enables efficient, simultaneous reduction of both polyatomic and doubly charged ion interferences”
Selection

High-purity oxygen or nitrous oxide can be supplied externally to the DCS instead of air when higher sensitivity or lower BECs are needed.

Affects whether optional gas plumbing should be provisioned for demanding applications.

Source
“it is also possible to use external supplies of high-purity oxygen (O2) or nitrous oxide (N2O) in the DCS instead of air”
Performance

For Ge, AHM cannot reduce M++ interferences, but Air cell mode detects GeO+ product ions via mass shift for low-level detection.

Informs method choice for germanium in light-REE-containing samples.

Source
“AHM cannot reduce the M++ interferences on 72Ge and 73Ge (high BECs)”
Performance

Air cell mode can outperform AHM for some polyatomic interferences, as shown by lower BECs for Gd detected as GdO+.

Supports selecting Air cell mode for challenging polyatomic overlaps.

Source
“Air cell mode can be more effective than AHM at reducing polyatomic ion interferences in some cases.”
Performance

For S, P, and Si, Air cell mode gives lower BECs and detection limits and higher sensitivity than AHM.

Relevant when configuring methods for difficult elements like sulfur and phosphorus.

Source
“Air cell mode enables analysis with even lower BECs and detection limits (DLs) and higher sensitivity”
Operation

For gold in tantalum-containing samples, on-mass analysis in Air cell mode works because TaO+ reacts with air to form TaOO+ while Au+ does not react.

Shows when on-mass rather than mass-shift mode should be used.

Source
“the interfering ion, TaO+, reacts efficiently with air to form TaOO+”
Performance

MoO+ and WO+ interferences on Cd and Hg react with air to form MoOO+ and WOO+, while Cd and Hg do not readily react, enabling quantification.

Guides interference management for Cd and Hg in Mo/W-rich environmental and food samples.

Source
“Both of these interferences, MoO+ (on Cd+) and WO+ (on Hg+), react with air to form MoOO+ and WOO+”
Selection

Because multitune preset methods need no external reaction cell gases, the 9500 can replace conventional single quadrupole ICP-MS systems.

Supports upgrade decisions from single quad to triple quad without adding gas infrastructure.

Source
“the 9500 offers a streamlined workflow and serves as a seamless, high-performance replacement for conventional single quadrupole ICP-MS systems”