Optimizing Microplastic Characterization by LDIR: Automated versus Manual Workflows
Technical Reference Summary
Covered Models & Document Scope
Technical overview of the Agilent 8700 LDIR chemical imaging system with Clarity control software; covers the Particle Analysis (automated) workflow with default size range 20 to 500 µm and the manual analysis workflow; evaluated with NIST-traceable polystyrene latex beads of 2, 5, and 10 µm on low-E slides (Agilent part number M7300-68010) and Al-coated polyester filters (Agilent part number M7300-68011)
Technical Data
| Parameter | Value | Condition |
|---|---|---|
| Default particle size range (Particle Analysis workflow) | 20 to 500 µm p.1 ↗ | adjustable based on user methodology |
| Minimum detectable particle size, automated mode | 5 µm p.7 ↗ | 5 µm PS beads, low-E slides and filters |
| IR wavenumber used for imaging | 1,442 cm–1 p.3 ↗ | |
| Low-E IR reflective slide dimensions | 25 × 75 mm p.2 ↗ | Agilent part number M7300-68010 |
| Al-coated polyester filter diameter | 25 mm p.2 ↗ | Agilent part number M7300-68011 |
| Al-coated polyester filter pore size | 0.8 µm p.2 ↗ | Agilent part number M7300-68011 |
| Al-coated polyester filter coating | 100/0 nm p.2 ↗ | Agilent part number M7300-68011 |
| PS latex bead sizes evaluated | 2, 5, and 10 µm p.1 ↗ | NIST-traceable |
Operating Limits
- Default particle size range (Particle Analysis workflow): 20 to 500 µm p.1 ↗
- Minimum detectable particle size, automated mode, low-E slides: 2 µm p.7 ↗
- Minimum detectable particle size, automated mode, low-E slides and filters: 5 µm p.7 ↗
- Minimum detectable particle size, automated mode, low-E slides and filters: 10 µm p.7 ↗
Ordering Data
| Ordering Code | Part No. | Variant / Code | Applies To | Note |
|---|---|---|---|---|
| Low-E infrared (IR) reflective slide p.2 ↗ | M7300-68010 | 25 × 75 mm | Substrate used in the study | |
| Al-coated polyester filter p.2 ↗ | M7300-68011 | 25 mm, 100/0 nm coating, 0.8 µm pore size | Substrate used in the study |