Technical Reference Summary

Covered Models & Document Scope

Technical overview describing use of the injector program of the Agilent 1260 Infinity II Vialsampler (G7129C) and 1260 Infinity II Multisampler (G7167A) for preparation of calibration solutions by serial dilution; covers 100 µL analytical head/100 µL loop default setup and 900 µL analytical head (option #163)/900 µL loop (option #156) for the Multisampler, with sample thermostat option #101; system includes 1260 Infinity II Flexible Pump (G7104C), Multicolumn Thermostat (G7116A) and Diode Array Detector HS (G7117C).

Technical Data

Parameter Value Condition
Analytical head volume 100 µL p.2 ↗ Vialsampler G7129C, default setup
Loop volume 100 µL p.2 ↗ Vialsampler G7129C, default setup
Sample thermostat option #101 p.2 ↗ Vialsampler G7129C and Multisampler G7167A
Diluent draw volume 80 µL p.3 ↗ injector program, location D1F-E1
Wait time 0.6 minutes p.3 ↗ injector program
Column Agilent InfinityLab Poroshell 120 EC-C18 p.5 ↗
Calibration concentration levels 20, 4, 0.8, 0.16 p.2 ↗ serial dilution
Caffeine stock solution concentration 100 µg/mL p.7 ↗
Calibration function slope (Multisampler 100 µL, manual) 27.85 p.7 ↗ y=27.85*x+0.53
Calibration function slope (Multisampler 100 µL, autosampler) 27.64 p.7 ↗ y=27.64*x-0.18
Calibration function slope (Vialsampler 100 µL, manual) 27.76 p.7 ↗ y=27.76*x+0.43
Calibration function slope (Vialsampler 100 µL, autosampler) 27.65 p.7 ↗ y=27.65*x-0.73
Calibration function slope (Multisampler 900 µL, manual) 27.69 p.7 ↗ y=27.69*x+0.54
Calibration function slope (Multisampler 900 µL, autosampler) 27.53 p.7 ↗ y=27.53*x+0.02
Calibration function R² (Multisampler 100 µL, manual) 0.99999 p.7 ↗
Calibration function R² (Multisampler 100 µL, autosampler) 1.00000 p.7 ↗
Calibration function R² (Vialsampler 100 µL, manual) 0.99999 p.7 ↗
Calibration function R² (Vialsampler 100 µL, autosampler) 0.99999 p.7 ↗
Calibration function R² (Multisampler 900 µL, manual) 0.99999 p.7 ↗
Calibration function R² (Multisampler 900 µL, autosampler) 1.00000 p.7 ↗
Signal-to-noise acceptance criterion 3:1 or 2:1 p.7 ↗ LOD estimation

Operating Limits

  • LOD signal-to-noise ratio acceptance: 3:1 or 2:1 p.7 ↗