Let Your Autosampler Do Your Pipetting
Technical Reference Summary
Covered Models & Document Scope
Technical overview describing use of the injector program of the Agilent 1260 Infinity II Vialsampler (G7129C) and 1260 Infinity II Multisampler (G7167A) for preparation of calibration solutions by serial dilution; covers 100 µL analytical head/100 µL loop default setup and 900 µL analytical head (option #163)/900 µL loop (option #156) for the Multisampler, with sample thermostat option #101; system includes 1260 Infinity II Flexible Pump (G7104C), Multicolumn Thermostat (G7116A) and Diode Array Detector HS (G7117C).
Technical Data
| Parameter | Value | Condition |
|---|---|---|
| Analytical head volume | 100 µL p.2 ↗ | Vialsampler G7129C, default setup |
| Loop volume | 100 µL p.2 ↗ | Vialsampler G7129C, default setup |
| Sample thermostat option | #101 p.2 ↗ | Vialsampler G7129C and Multisampler G7167A |
| Diluent draw volume | 80 µL p.3 ↗ | injector program, location D1F-E1 |
| Wait time | 0.6 minutes p.3 ↗ | injector program |
| Column | Agilent InfinityLab Poroshell 120 EC-C18 p.5 ↗ | |
| Calibration concentration levels | 20, 4, 0.8, 0.16 p.2 ↗ | serial dilution |
| Caffeine stock solution concentration | 100 µg/mL p.7 ↗ | |
| Calibration function slope (Multisampler 100 µL, manual) | 27.85 p.7 ↗ | y=27.85*x+0.53 |
| Calibration function slope (Multisampler 100 µL, autosampler) | 27.64 p.7 ↗ | y=27.64*x-0.18 |
| Calibration function slope (Vialsampler 100 µL, manual) | 27.76 p.7 ↗ | y=27.76*x+0.43 |
| Calibration function slope (Vialsampler 100 µL, autosampler) | 27.65 p.7 ↗ | y=27.65*x-0.73 |
| Calibration function slope (Multisampler 900 µL, manual) | 27.69 p.7 ↗ | y=27.69*x+0.54 |
| Calibration function slope (Multisampler 900 µL, autosampler) | 27.53 p.7 ↗ | y=27.53*x+0.02 |
| Calibration function R² (Multisampler 100 µL, manual) | 0.99999 p.7 ↗ | |
| Calibration function R² (Multisampler 100 µL, autosampler) | 1.00000 p.7 ↗ | |
| Calibration function R² (Vialsampler 100 µL, manual) | 0.99999 p.7 ↗ | |
| Calibration function R² (Vialsampler 100 µL, autosampler) | 0.99999 p.7 ↗ | |
| Calibration function R² (Multisampler 900 µL, manual) | 0.99999 p.7 ↗ | |
| Calibration function R² (Multisampler 900 µL, autosampler) | 1.00000 p.7 ↗ | |
| Signal-to-noise acceptance criterion | 3:1 or 2:1 p.7 ↗ | LOD estimation |
Operating Limits
- LOD signal-to-noise ratio acceptance: 3:1 or 2:1 p.7 ↗