Performance Characteristics of the Agilent 1260 Infinity II LC System for Routine Analysis

Technical Reference Summary

Covered Models & Document Scope

Agilent 1260 Infinity II LC System (quaternary system) technical overview: quaternary pump G7111B, vialsampler G7129A with integrated column compartment G7130A, variable wavelength detector G7114B; application examples with InfinityLab Poroshell 120 EC-C18 columns

Technical Data

Parameter Value Condition
Maximum operating pressure 600 bar p.1 ↗ 1260 Infinity II quaternary pump
Flow rate range up to 5 mL/min p.1 ↗ 2.1, 3, and 4.6 mm id columns
Column compartment temperature up to 80 °C p.2 ↗ integrated column compartment, two LC columns
Sample cooler temperature down to 4 °C p.2 ↗ vialsampler
Detector noise <±2.5 μAU p.2 ↗ VWD
Baseline drift <1 × 10–4 AU/h p.2 ↗ VWD
Maximum data rate 120 Hz p.2 ↗ VWD
Injection loop volume 100 µL p.2 ↗ vialsampler G7129A
Metering device volume 100 µL p.2 ↗ vialsampler G7129A
Detector flow cell path length 10 mm p.2 ↗ VWD standard cell
Injection volume 5 µL p.3 ↗ isocratic sample, Poroshell 120 EC-C18
Flow rate 1.0 mL/min p.3 ↗ isocratic sample, Poroshell 120 EC-C18
Stop time 15 min p.3 ↗ isocratic sample, Poroshell 120 EC-C18
Detection wavelength 254 nm p.3 ↗ UV, 40 Hz
Post time 1 min p.3 ↗ RRLC checkout sample

Operating Limits

  • Maximum operating pressure: 600 bar p.1 ↗
  • Maximum column compartment temperature: 80 °C p.2 ↗
  • Minimum sample cooler temperature: 4 °C p.2 ↗
  • Maximum detector data rate: 120 Hz p.2 ↗