Reduce PFAS Background with the Agilent PFAS Analysis HPLC Conversion Kit

Technical Reference Summary

Covered Models & Document Scope

Agilent InfinityLab PFAS Analysis HPLC Conversion Kit — a ready-to-use kit for converting standard Agilent LC systems (recommended 1290 Infinity II/III) for trace-level PFAS analysis with LC/MS/MS; document also covers the InfinityLab PFC delay column and conversion options for other Agilent LC modules.

Technical Data

Parameter Value Condition
Column temperature 50 °C p.3 ↗ PFAS-optimized LC setup
Sample diluent Methanol/water (80/20 v/v) p.3 ↗ PFAS-optimized LC setup
Mobile phase A 5 mM ammonium acetate in water p.3 ↗ PFAS-optimized LC setup
Seat backflush Methanol/water (10/90 v/v), 10 seconds p.3 ↗ PFAS-optimized LC setup
Stop time 14 minutes p.3 ↗ PFAS-optimized LC setup
Post time 0.3 minutes p.3 ↗ PFAS-optimized LC setup
PFHpA background > 3,000 fg p.6 ↗ Standard LC setup
PFNA background > 500 fg p.6 ↗ Standard LC setup
Retention time shift (PFTeA) 1.7 min p.7 ↗ with InfinityLab PFC delay column
Retention time shift (PFHxS) 2.0 min p.7 ↗ with InfinityLab PFC delay column

Operating Limits

  • Recommended LC system: Agilent 1290 Infinity II/III p.8 ↗
  • Quaternary low-pressure mixing pumps: should not be considered for PFAS analyses p.8 ↗

Ordering Data

Ordering Code Part No. Variant / Code Applies To Note
High-pressure filter assembly with PEEK/stainless steel p.2 ↗ 5067-1602 Pump head adapter assembly
Multisampler tubing, cut to length, PEEK p.2 ↗ 0890-1761 Inline filter