TPS-mini Turbomolecular Pumping System Instruction Manual

Contents

  1. Overview pp. 1-6
    Product overview, symbols, and general safety information
  2. Warning pp. 7-18
    Safety warnings and precautions for turbomolecular pumps
  3. Procedure pp. 19-30
    Preparation, mounting, and connection procedures
  4. Procedure pp. 31-42
    Switching on, use, and stopping the TPS-mini
  5. Procedure pp. 43-54
    Maintenance and servicing instructions
  6. Procedure pp. 55-60
    Storage conditions and disposal guidance
  7. Spec pp. 268-312
    Controller description, flange connections, interconnections, and specifications
  8. Spec pp. 298-312
    RS 232/RS 485 communication, protocol, and window meanings
  9. p. 273 table, 35 rows
  10. p. 310 table, 16 rows
  11. p. 274 table, 3 rows
  12. p. 289 table, 3 rows
  13. p. 307 table, 9 rows
  14. p. 312 table, 10 rows
  15. p. 259 table, 2 rows
  16. p. 272 table, 6 rows
  17. p. 288 table, 13 rows

Tab. 3

ParameterSpecification
Pumping speed (l/s) (With inlet screen)
With CF 4.5” or ISO 63:
N248 l/s
He44 l/s
H232 l/s
With KF 40 NW:
N237 l/s
He29 l/s
H223 l/s
Base pressure*5 x 10-8 mbar (3.75 x 10-8 Torr)
Inlet flangeCF 4.5" O.D.; ISO63; KF 40 NW
Rotational speed70000 RPM
Start-up time< 2 minutes
Operating positionany
Operating ambient temperature+ 5 °C to + 35 °C
Bakeout temperature120 °C at inlet flange max. (CF flange)
80 °C at inlet flange max. (ISO flange)
Noise level56.4 dB (A) at 1 meter
Lubricantpermanent lubrication
Power supply (Line type):
Input voltage:100 – 240 Vac
Input frequency:50 – 60 Hz
Max input power:220 VA
Stand-by power:10 W
Protection fuse2.5 A
Compliance with:EN 55011 (Class A Group 1)
EN 61326-1
EN 61010-1
EN 1012-2
EN 12100
Indoor Use Only
Max Altitude 2000m
Installation categoryII
Pollution degree2
Power cableWith European, USA or UK plug 2 meters long

*Base pressure

Window-Meanings

N.Read/WriteData TypeDescriptionAdmitted Values
205RNPump statusStop = 0 Waiting intlk = 1 Starting = 2 Auto-tuning = 3 Braking = 4 Normal = 5 Fail = 6
206RNError codeBit description: see the following figure
211RNController Heatsink Temperature (°C)0 to 110
216RNController Air Temperature (°C).0 to 70
224RAPressure readingFormat = X.X E XX
226RNRotation Frequency (rpm)
300RNCycle time in minutes (zeroed by the reset command)0 to 999999
301RNCycle number (zeroed by the reset command)0 to 9999
302RNPump life in hours (zeroed by the reset command)0 to 999999
310 to 347Reserved to Agilent service
400RACRC EPROM (QE)QE8XXXX (where “XXXX” are variable)
401RABoot CRC (BL)BL2XXXX (where “XXXX” are variable)
402RACRC Param. (PA)PA8XXXX (where “XXXX” are variable)
404RACRC Parameter structure"XXXX"
406RAProgram Listing Code"XXXXXXXXXX"
407RAParameter Listing Code"XXXXXXXXXX"

Technical Specification

CharacteristicValue
Serial communication (kit)RS232 cable with a 9-pin D type male connector and a 9-pin D type female connector, and software (optional)
Storage temperature- 20 °C to + 70 °C
Weight kg (lbs)KF: 7.9 kg (17.4 lbs) ISO: 7.5 kg (16.5 lbs) CFF: 8.8 kg (19.4 lbs)

*According to standard DIN 28 428, the base pressure is that measured in a leak-free test dome, 48 hours after the completion of test dome bake-out, with a Turbopump fitted with a ConFlat flange.

J1 – IN-OUT

PIN N.SIGNAL NAMEINPUT/OUTPUT
14PROGRAMMABLE ANALOG SIGNAL (+)OUT
15GROUND
PROGRAMMABLE ANALOG SIGNAL (-)OUT

Tab. 8

N.Read/WriteData TypeDescriptionAdmitted Values
000R/WLStart/Stop (in remote/ mode the window is a read only)Start = 1 Stop = 0
001R/WLLow Speed Activation (in remote/ mode the window is a read only)No = 0 Yes = 1 (default = 0)
008R/WLRemote or Serial configurationRemote = 1 Serial = 0 (default = 1)
100R/WLSoft Start (write only in Stop condition)YES = 1 NO = 0
101R/WNR1 Set Point type0 = Frequency 1 = Power 2 = Time 3 = Normal 4 =Pressure (default = 0)
102R/WNR1 Set Point valve (expressed in Hz, W or s)(default = 867)
103R/WNSet Point delay: time between the pump start and the set point check (seconds)0 to 999999 (default = 0)
104R/WLSet Point signal activation type: the signal can be "high level active" or "low level active"0 = high level active 1 = low level active (default = 0)
105R/WNSet point hysteresis (in % of value)0 to 100 (default = 2)

Tab. 9

DescriptionPart number
Mains cable NEMA plug [3m]9699958
Mains cable EU plug [3m]9699957
Serial cable and T-Plus Software9699883
DIY Full Range Gauge Kit / 74 FS KF 409699190
DIY Full Range Gauge Kit / 74 FS ISO 639699192
DIY Full Range Gauge Kit / 74 FS CFF 4.5”9699193
FRG-700 3 m cable9699960
FRG-700 5 m cable9699961
Inlet screen DN409699309
Inlet screen DN639699300

Tab. 1

FLANGEFIXING DEVICEN.FIXING TORQUE
ISO 63M10 clamps422 Nm
M8 claws411 Nm

Tab. 2

CONTROL NAMEDESCRIPTION
J1 IN-OUTInput-output connector to remote control the TPS-mini
J2 SERIALSerial input-output connector to control the TPS-mini via an RS 232 or RS 485 connection
STATUSBlue LED to show the system operating condition
P1 VENTConnector to control the optional vent valve
J5 GAUGEConnector to control external EyeSys Gauge
ON-OFF red switch with integrated light (not shown in the picture)It is located on the other side of the system (with respect to the illustrated control panel). It is the power supply switch. The integrated red light shows the power supply status: it lights when the power supply is applied to the TPS-mini

Tab. 4

PIN N.SIGNAL NAMEINPUT/OUTPUT
1START/STOP (+)IN
2START/STOP (-)IN
3INTERLOCK (+)IN
4INTERLOCK (-)IN
5SPEED SETTING (+)IN
6SPEED SETTING (-)IN
7SOFT START (+)IN
8SOFT START (-)IN
9+24 Vdc Imax = 30 mAOUT
10NORMAL OUTPUT (relay) Vmax = 125V Imax = 200 mAOUT
11PROGRAMMABLE SET POINTOUT
12NORMAL OUTPUTOUT
13FAULT OUTPUTOUT

Engineering Facts

14 facts · every claim linked to a page
Installation

The TPS-mini cannot be mounted by its base, so a separate support or flange connection must be designed.

Prevents a mounting design that relies on bolting the base to a frame.

Source
“Le TPS-mini ne peut être fixé par sa base.”
Installation

The inlet flange must be connected to a fixed counter-flange able to withstand 50 Nm torque around its axis.

Defines the mechanical strength required of the chamber-side flange and support.

Source
“Fix the TPS-mini in a stable position connecting the inlet flange of the turbopump to a fixed counter-flange capable of withstanding a torque of 50 Nm around its axis.”
Installation

ISO 63 inlet flanges require 4 M10 clamps at 22 Nm or 4 M8 claws at 11 Nm.

Gives the exact clamp count and tightening torque for leak-free ISO mounting.

Source
“ISO 63 M10 clamps 4 22 Nm M8 claws 4 11 Nm”
Installation

A minimum free space of 20 cm around the system is required for proper air circulation.

Sets the clearance needed in the vacuum system enclosure for cooling.

Source
“Assure a free space all around the system at minimum of 20 cm to allow proper air circulation.”
Installation

The mains cable must be three-wire (Ph+N+Earth) with at least 0.75 mm2 (AWG18) conductor section.

Specifies the minimum supply cable gauge and earthing requirement.

Source
“The wire section has to be at least 0.75 mm2 (AWG18).”
Operation

During bakeout the inlet flange must not exceed 120 °C with ConFlat flanges or 80 °C with ISO flanges.

Limits bakeout temperature depending on flange type to avoid pump damage.

Source
“While heating the vacuum chamber, the temperature of the inlet flange must not exceed 120 °C with ConFlat flanging and 80 °C with ISO flanging.”
Operation

The pump must never be run with the inlet flange open to atmosphere or unblanked.

Prevents rotor overspeed and contamination damage during commissioning.

Source
“Never use the turbopump when the inlet flange is not connected to the vacuum chamber or is not blanked.”
Operation

The blue STATUS LED flashing at about 400 ms indicates ramp, braking, stop or waiting-for-interlock; about 200 ms indicates an error.

Allows field diagnosis of operating state from the LED without a controller.

Source
“slowly flashing (period of about 400 ms): the system is in ramp, or in braking, or in Stop, or in “Waiting for interlock” status;”
Operation

For emergency stop, remove the supply cable from the mains plug or move the external switch to OFF.

Defines the emergency shutdown method to be included in the site safety procedure.

Source
“To immediately stop the TPS-mini in an emergency condition it is necessary to remove the supply cable from the mains plug or moving the external switch to OFF position.”
Maintenance

Before any work on the system, disconnect mains, vent the pump, wait for rotor stop and for surface temperature below 50 °C.

Sets the safe lockout and cool-down conditions for service work.

Source
“wait until the rotor has stopped turning and wait until the surface temperature of the pump falls below 50 °C.”
Maintenance

The TPS-mini requires no maintenance and any work must be done by authorized personnel.

Confirms no scheduled servicing is needed but restricts interventions to qualified staff.

Source
“The TPS-mini does not require any maintenance. Any work performed on the system must be carried out by authorized personnel.”
Selection

The pump is designed for high throughput of N2, Ar and lighter gases; heavier gases require contacting Agilent support.

Determines whether the pump suits the process gas before selection.

Source
“The pump is designed to pump high throughput of N2, Ar and lighter gas. Should you need to pump gases heavier than Ar, please contact Agilent technical support for informations.”
Selection

The system is not suitable for pumping liquids or solid particles.

Excludes applications with condensable liquids or particulate carryover.

Source
“It is not suitable for pumping liquids or solid particles.”
Selection

The shelf life is 12 months from shipping date; if exceeded, the pump must be returned to the factory.

Affects stock rotation and acceptance of pumps stored long before installation.

Source
“the shelf life of a turbomolecular pump is 12 months from the shipping date.”