Overview of controller features, safety notes, and specifications
Procedurepp. 22-24
Mounting, electrical connections, and setup instructions
Procedurepp. 25-27
Operating procedures and front panel controls
Warningpp. 28-30
List of error codes and troubleshooting guidance
Specpp. 208-213
Detailed technical specifications and dimensions
Procedurepp. 214-216
Fuse replacement and holder assembly instructions
Specpp. 223-225
P1 purge/vent connector pinout and wiring
Specpp. 226-228
Electrical characteristics of I/O signals
Procedurepp. 229-231
Controlled venting operation and parameters
Specpp. 232-234
Serial communication protocol and commands
Overviewpp. 238-240
Display window indicators and their meanings
Procedurepp. 244-246
Soft start function configuration and use
Procedurepp. 250-252
How to operate the controller via front panel
Specpp. 253-255
Measurement parameters and readouts
Procedurepp. 259-261
Pump configuration menu options
Procedurepp. 262-264
I/O configuration menu options
pp. 268-270
Example configurations and usage scenarios
p. 255table, 3 rows
p. 256table, 2 rows
p. 202table, 6 rows
p. 212table, 19 rows
p. 242table, 9 rows
p. 201table, 5 rows
p. 240table, 7 rows
p. 252table, 7 rows
p. 281table, 6 rows
p. 205table, 4 rows
Table
STATUS
Hz
HIGH SPEED ADJ
TO
Hz
MESSAGE
X X X
X X X X
Y Y Y Y
CHANGE OK
X X X
X X X X
Y Y Y Y
OUT OF LIMITS
X X X
X X X X
Y Y Y Y
CHANGE ABORTED
First level of the configuration menu
STATUS
HIGH SPEED ADJ
XXX Hz
XXXX TO YYYY Hz
XXXX Hz
Figure 3 Controller X3501-64003 Rear Panel
Number
Description
1
Controller power entry module consisting of mains fuses, mains socket and EMC filter.
2
RS-232 / RS-485 communication port connector.
3
Logic input/output signals connector plus programmable analog output (mating connector supplied with Interlock link).
4
Pump cable.
5
Vent Valve and Purge Valve control output connector.
6
External Fan output connector.
Tab. 2
Parameter
Specification
Input: Voltage
100 ÷ 240 Vac (±10 %), 1-phase
Frequency
50 to 60 Hz
Power
800 VA maximum
Output for pump: Power during pump ramp-up
450 W maximum
Power normal (pump water cooled)
450 W maximum
Power normal (pump air cooled)
300 W maximum
Operating temperature
5 °C to +45 °C
Storage temperature
-20 °C to +70 °C
Fuse: Mains
2 x T10 AH (slow blow) 250 V 5x20 fuses
Serial communication (T-Plus kit)
RS232 cable with a 9-pin D type male connector and a 9-pin D type female connector, and T-Plus software (optional)
Power cable
With European or NEMA plug 3 meters long (optional)
Pump cable
5 meters long fixed cable
In compliance with norms
EN 61010-1 (2001) EN 61326-1 (2006) Emission level : Class B for Residential Application Immunity level : for Industrial Application
Max altitude
2000 m
Protection category
IP 20
Internal use only
Weight (both models)
6.3 kg (13.9 lbs)
Installation category
II
Pollution degree
2
Window Meanings
N.
Read/Write
Data Type
Description
Admitted Values
310 to 399
Reserved to Agilent service
400
R
A
CRC EPROM (QE)
QE8XXXX (where “XXXX” are variable)
402
R
A
CRC Param. (PA)
PA8XXXX (where “XXXX” are variables)
404
R
A
CRC parameter structure
XXXX
406
R
A
Program Listing code & revision
XXXX
407
R
A
Parameter Listing code & revision
XXXX
500 to 502
Reserved to Agilent service
503
R/W
N
RS 485 address
0 to 31 (default = 0)
504
R/W
L
Serial type selection
0 = RS 232 1 = RS 485 (default = 0)
Controls, Indicators and Connectors
№
Description
1
Keyboard push-button for START, STOP, RESET mode selection. It is active only when the front panel operation has been selected. By pressing once the starting phase begins; if pressed again it stops the pump. If the pump has been stopped automatically by a fault, this push-button must be pressed once to reset the controller and a second time to restart the pump.
2
Keyboard push-button to recall on the display the cycle number, cycle time and pump life and serial number.
3
Keyboard push-button to recall on the display the pump current, pump temperature, pump power, rotational speed. It is always active, regardless of the selected operating mode.
4
Keyboard push-button for HIGH/LOW SPEED mode selection. It is active only when the front panel operation has been selected. Pressed repeatedly, toggles between HIGH SPEED and LOW SPEED.
0 to 65535 Corresponding to 0 to 13107 sec (Default = 15)
143
R/W
N
External Fan Configuration
0 = Always ON 1 = Automatic 2 = Serial (default = 2)
144
R/W
L
External Fan activation
0 = OFF 1 = ON (default = 0)
145
R/W
L
Purge Valve open/close (N.C.)
0 = close 1 = open (Default =1)
147
R/W
N
Vent Valve opening (Time expressed in 0.2 s)
0 to 65535 bit 0 = infinite min 0.2 s, Max 13107 s (default = 0)
155
R
N
Power Limit Applied expressed in W
157
R/W
L
Gas Load Type
0 = Ar 1 = N2 (default = 0)
Counters / Controller Data
Display
Description
CYCLE XXXX
Number of start/stop cycles
TIME XXXXXm
Time of last cycle (minutes)
P.LIFE XXXXXX
total operating time (hours) of the pump
MN
Model Number of the controller
SN
Serial Number
FW
Firmware version
PA
Parameter Listing version
Pressing once the COUNTERS pushbutton displays CYCLE, TIME, P.LIFE; pressing twice displays MN, SN; pressing a third time displays FW, PA.
Tab. 8
Description
Part number
Turbo-V 1001 Rack Controller
X3501-64003
T-Plus communication software and serial cable kit
969-9883
Turbo-V HT Rack Air Cooling kit
X3501-68001
Mains cable NEMA plug 3m long
969-9958
Mains cable European plug 3m long
969-9957
Turbo-V vent
X3501-68002
Tab. 1
MESSAGE
DESCRIPTION
REPAIR ACTION
CHECK CONNECTION TO PUMP
Wrong connection between the pump and the controller.
Check connections between controller and pump. Press the START push-button twice to start the pump.
PUMP WAITING INTERLOCK
The interlock signal of J1 connector is activated by an interruption of the link between pin 3 and 9 or between pin 4 and 15 of P1 connector, or because the external interlock signal is open.
Reset the short circuit between pin 3 - 9 and between pin 4 - 15 of J1 connector, or close the external interlock signal.
FAULT: PUMP OVERTEMP.
The pump upper bearing temperature exceeded 60 °C.
Wait until the temperature decrease below threshold value. Press the START push-button twice to start the pump.
FAULT: CONTROLLER OVERTEMPERATURE
The controller internal temperature exceeded 65 °C.
Wait until the temperature decrease below threshold value. Check that the pump cable length has been set correctly. Press the START push-button twice to start the pump.
Engineering Facts
14 facts · every claim linked to a page
Installation
The controller must be powered from a properly grounded 3-wire socket to avoid electric shock and meet CE requirements.
Grounding and correct cord/plug are mandatory for safe installation and CE compliance.
Source
“Use this power cord and plug in conjunction with a properly grounded power socket to avoid electrical shock and to satisfy CE requirements.”
Safety
The input power cable must be disconnected before any servicing because high voltage inside can be lethal.
Defines the lockout step required before maintenance work.
Source
“Before servicing the unit, disconnect the input power cable.”
Installation
The controller must be placed so that free air flows through its ventilation holes and must not be used in dusty, explosive, or wet environments.
Cooling clearance and environment restrictions determine where the unit can be mounted.
Source
“it must be positioned so that free air can flow through the holes. Do not install or use the controller in an environment exposed to atmospheric agents (rain, snow, ice), dust, aggressive gases, or in explosive environments or those with a high fire risk.”
Operation
Operating ambient must stay within 0 to +45 °C and 0–95 % RH without condensation.
Sets the allowable site conditions for reliable operation.
Source
“temperature: from 0 °C to +45 °C;”
Operation
The pump must never be run with its inlet open to atmosphere; it must be connected to the system or blanked off.
Prevents equipment damage and personnel injury during commissioning.
Source
“Never operate the Turbo-V pump if the pump inlet is not connected to the system or blanked off.”
Maintenance
The controller itself requires no maintenance, but any work on the system must be done by authorized personnel.
Clarifies service scope and who is allowed to intervene.
Source
“The Turbo-V 1001 Rack Controller does not require any maintenance. Any work performed on the system must be carried out by authorized personnel.”
Operation
After an automatic fault stop, START must be pressed once to reset and a second time to restart the pump.
Defines the correct recovery sequence after a fault trip.
Source
“If the pump has been stopped automatically by a fault, this push-button must be pressed once to reset the controller and a second time to restart the pump.”
Operation
The configuration menu is entered by holding push-buttons 2 and 3 together for at least 2 seconds.
Needed to set parameters during commissioning.
Source
“Push-buttons 2 and 3, if pressed together for at least 2 seconds allow to access the controller configuration menu.”
Installation
Remote I/O and serial cables longer than 3 m must be shielded.
Cable shielding is required for EMC compliance on longer runs.
Source
“All connecting cables for remote I/O and serial connections, if longer than 3 m, must be of the shielded type.”
Installation
Serial and I/O port cables must be shielded and no longer than 30 m, with the shield bonded to the connector case at least on the controller side.
Sets the maximum cable length and shielding rule for signal wiring.
Source
“A shielded cable of 30 m maximum length has to be utilized for both serial and I/O port connections.”
Operation
The purge and vent valves are normally closed, so they stay closed on power failure.
Determines fail-safe behavior of the venting system on power loss.
Source
“The two valves are Normally Closed (N.C.) so if a power fail occurs, the valves will remain closed.”
Operation
Valve operating mode can only be changed while the pump is stopped.
Prevents configuration changes that could disturb a running pump.
Source
“The valve operating mode can be changed only with the pump in STOP status.”
Operation
Improper venting can damage the pump; the Auto Pump Speed mode should be used.
Guides correct venting configuration to avoid pump damage.
Source
“The vent could damage the pump. Please, use the “Auto Pump Speed” mode or refer to Agilent personnel.”
Operation
During controlled venting the vent valve modulates opening so pump deceleration stays between 3 and 5 Hz/s, adjusted every 10 seconds.
Explains the deceleration limit that protects the pump during venting.
Source
“the pump deceleration quickness is maintained in the range between 3 and 5 Hz/s. The timing is adjusted every 10 seconds.”