Turbo-V 1001 Rack Controller

Contents

  1. Overview pp. 16-21
    Overview of controller features, safety notes, and specifications
  2. Procedure pp. 22-24
    Mounting, electrical connections, and setup instructions
  3. Procedure pp. 25-27
    Operating procedures and front panel controls
  4. Warning pp. 28-30
    List of error codes and troubleshooting guidance
  5. Spec pp. 208-213
    Detailed technical specifications and dimensions
  6. Procedure pp. 214-216
    Fuse replacement and holder assembly instructions
  7. Spec pp. 223-225
    P1 purge/vent connector pinout and wiring
  8. Spec pp. 226-228
    Electrical characteristics of I/O signals
  9. Procedure pp. 229-231
    Controlled venting operation and parameters
  10. Spec pp. 232-234
    Serial communication protocol and commands
  11. Overview pp. 238-240
    Display window indicators and their meanings
  12. Procedure pp. 244-246
    Soft start function configuration and use
  13. Procedure pp. 250-252
    How to operate the controller via front panel
  14. Spec pp. 253-255
    Measurement parameters and readouts
  15. Procedure pp. 259-261
    Pump configuration menu options
  16. Procedure pp. 262-264
    I/O configuration menu options
  17. pp. 268-270
    Example configurations and usage scenarios
  18. p. 255 table, 3 rows
  19. p. 256 table, 2 rows
  20. p. 202 table, 6 rows
  21. p. 212 table, 19 rows
  22. p. 242 table, 9 rows
  23. p. 201 table, 5 rows
  24. p. 240 table, 7 rows
  25. p. 252 table, 7 rows
  26. p. 281 table, 6 rows
  27. p. 205 table, 4 rows

Table

STATUSHzHIGH SPEED ADJTOHzMESSAGE
X X XX X X XY Y Y YCHANGE OK
X X XX X X XY Y Y YOUT OF LIMITS
X X XX X X XY Y Y YCHANGE ABORTED

First level of the configuration menu

STATUSHIGH SPEED ADJ
XXX HzXXXX TO YYYY Hz
XXXX Hz

Figure 3 Controller X3501-64003 Rear Panel

NumberDescription
1Controller power entry module consisting of mains fuses, mains socket and EMC filter.
2RS-232 / RS-485 communication port connector.
3Logic input/output signals connector plus programmable analog output (mating connector supplied with Interlock link).
4Pump cable.
5Vent Valve and Purge Valve control output connector.
6External Fan output connector.

Tab. 2

ParameterSpecification
Input: Voltage100 ÷ 240 Vac (±10 %), 1-phase
Frequency50 to 60 Hz
Power800 VA maximum
Output for pump: Power during pump ramp-up450 W maximum
Power normal (pump water cooled)450 W maximum
Power normal (pump air cooled)300 W maximum
Operating temperature5 °C to +45 °C
Storage temperature-20 °C to +70 °C
Fuse: Mains2 x T10 AH (slow blow) 250 V 5x20 fuses
Serial communication (T-Plus kit)RS232 cable with a 9-pin D type male connector and a 9-pin D type female connector, and T-Plus software (optional)
Power cableWith European or NEMA plug 3 meters long (optional)
Pump cable5 meters long fixed cable
In compliance with normsEN 61010-1 (2001) EN 61326-1 (2006) Emission level : Class B for Residential Application Immunity level : for Industrial Application
Max altitude2000 m
Protection categoryIP 20
Internal use only
Weight (both models)6.3 kg (13.9 lbs)
Installation categoryII
Pollution degree2

Window Meanings

N.Read/WriteData TypeDescriptionAdmitted Values
310 to 399Reserved to Agilent service
400RACRC EPROM (QE)QE8XXXX (where “XXXX” are variable)
402RACRC Param. (PA)PA8XXXX (where “XXXX” are variables)
404RACRC parameter structureXXXX
406RAProgram Listing code & revisionXXXX
407RAParameter Listing code & revisionXXXX
500 to 502Reserved to Agilent service
503R/WNRS 485 address0 to 31 (default = 0)
504R/WLSerial type selection0 = RS 232 1 = RS 485 (default = 0)

Controls, Indicators and Connectors

№Description
1Keyboard push-button for START, STOP, RESET mode selection. It is active only when the front panel operation has been selected. By pressing once the starting phase begins; if pressed again it stops the pump. If the pump has been stopped automatically by a fault, this push-button must be pressed once to reset the controller and a second time to restart the pump.
2Keyboard push-button to recall on the display the cycle number, cycle time and pump life and serial number.
3Keyboard push-button to recall on the display the pump current, pump temperature, pump power, rotational speed. It is always active, regardless of the selected operating mode.
4Keyboard push-button for HIGH/LOW SPEED mode selection. It is active only when the front panel operation has been selected. Pressed repeatedly, toggles between HIGH SPEED and LOW SPEED.
5LCD back-lighted alphanumeric display: dot matrix 4 lines x 16 characters.

Figure 2 Controller X3501-64003 Front Panel

Window Meanings

N.Read/WriteData TypeDescriptionAdmitted Values
126R/WNVent valve opening delay expressed in 0.2 sec0 to 65535 Corresponding to 0 to 13107 sec (Default = 15)
143R/WNExternal Fan Configuration0 = Always ON 1 = Automatic 2 = Serial (default = 2)
144R/WLExternal Fan activation0 = OFF 1 = ON (default = 0)
145R/WLPurge Valve open/close (N.C.)0 = close 1 = open (Default =1)
147R/WNVent Valve opening (Time expressed in 0.2 s)0 to 65535 bit 0 = infinite min 0.2 s, Max 13107 s (default = 0)
155RNPower Limit Applied expressed in W
157R/WLGas Load Type0 = Ar 1 = N2 (default = 0)

Counters / Controller Data

DisplayDescription
CYCLE XXXXNumber of start/stop cycles
TIME XXXXXmTime of last cycle (minutes)
P.LIFE XXXXXXtotal operating time (hours) of the pump
MNModel Number of the controller
SNSerial Number
FWFirmware version
PAParameter Listing version

Pressing once the COUNTERS pushbutton displays CYCLE, TIME, P.LIFE; pressing twice displays MN, SN; pressing a third time displays FW, PA.

Tab. 8

DescriptionPart number
Turbo-V 1001 Rack ControllerX3501-64003
T-Plus communication software and serial cable kit969-9883
Turbo-V HT Rack Air Cooling kitX3501-68001
Mains cable NEMA plug 3m long969-9958
Mains cable European plug 3m long969-9957
Turbo-V ventX3501-68002

Tab. 1

MESSAGEDESCRIPTIONREPAIR ACTION
CHECK CONNECTION TO PUMPWrong connection between the pump and the controller.Check connections between controller and pump. Press the START push-button twice to start the pump.
PUMP WAITING INTERLOCKThe interlock signal of J1 connector is activated by an interruption of the link between pin 3 and 9 or between pin 4 and 15 of P1 connector, or because the external interlock signal is open.Reset the short circuit between pin 3 - 9 and between pin 4 - 15 of J1 connector, or close the external interlock signal.
FAULT: PUMP OVERTEMP.The pump upper bearing temperature exceeded 60 °C.Wait until the temperature decrease below threshold value. Press the START push-button twice to start the pump.
FAULT: CONTROLLER OVERTEMPERATUREThe controller internal temperature exceeded 65 °C.Wait until the temperature decrease below threshold value. Check that the pump cable length has been set correctly. Press the START push-button twice to start the pump.

Engineering Facts

14 facts · every claim linked to a page
Installation

The controller must be powered from a properly grounded 3-wire socket to avoid electric shock and meet CE requirements.

Grounding and correct cord/plug are mandatory for safe installation and CE compliance.

Source
“Use this power cord and plug in conjunction with a properly grounded power socket to avoid electrical shock and to satisfy CE requirements.”
Safety

The input power cable must be disconnected before any servicing because high voltage inside can be lethal.

Defines the lockout step required before maintenance work.

Source
“Before servicing the unit, disconnect the input power cable.”
Installation

The controller must be placed so that free air flows through its ventilation holes and must not be used in dusty, explosive, or wet environments.

Cooling clearance and environment restrictions determine where the unit can be mounted.

Source
“it must be positioned so that free air can flow through the holes. Do not install or use the controller in an environment exposed to atmospheric agents (rain, snow, ice), dust, aggressive gases, or in explosive environments or those with a high fire risk.”
Operation

Operating ambient must stay within 0 to +45 °C and 0–95 % RH without condensation.

Sets the allowable site conditions for reliable operation.

Source
“temperature: from 0 °C to +45 °C;”
Operation

The pump must never be run with its inlet open to atmosphere; it must be connected to the system or blanked off.

Prevents equipment damage and personnel injury during commissioning.

Source
“Never operate the Turbo-V pump if the pump inlet is not connected to the system or blanked off.”
Maintenance

The controller itself requires no maintenance, but any work on the system must be done by authorized personnel.

Clarifies service scope and who is allowed to intervene.

Source
“The Turbo-V 1001 Rack Controller does not require any maintenance. Any work performed on the system must be carried out by authorized personnel.”
Operation

After an automatic fault stop, START must be pressed once to reset and a second time to restart the pump.

Defines the correct recovery sequence after a fault trip.

Source
“If the pump has been stopped automatically by a fault, this push-button must be pressed once to reset the controller and a second time to restart the pump.”
Operation

The configuration menu is entered by holding push-buttons 2 and 3 together for at least 2 seconds.

Needed to set parameters during commissioning.

Source
“Push-buttons 2 and 3, if pressed together for at least 2 seconds allow to access the controller configuration menu.”
Installation

Remote I/O and serial cables longer than 3 m must be shielded.

Cable shielding is required for EMC compliance on longer runs.

Source
“All connecting cables for remote I/O and serial connections, if longer than 3 m, must be of the shielded type.”
Installation

Serial and I/O port cables must be shielded and no longer than 30 m, with the shield bonded to the connector case at least on the controller side.

Sets the maximum cable length and shielding rule for signal wiring.

Source
“A shielded cable of 30 m maximum length has to be utilized for both serial and I/O port connections.”
Operation

The purge and vent valves are normally closed, so they stay closed on power failure.

Determines fail-safe behavior of the venting system on power loss.

Source
“The two valves are Normally Closed (N.C.) so if a power fail occurs, the valves will remain closed.”
Operation

Valve operating mode can only be changed while the pump is stopped.

Prevents configuration changes that could disturb a running pump.

Source
“The valve operating mode can be changed only with the pump in STOP status.”
Operation

Improper venting can damage the pump; the Auto Pump Speed mode should be used.

Guides correct venting configuration to avoid pump damage.

Source
“The vent could damage the pump. Please, use the “Auto Pump Speed” mode or refer to Agilent personnel.”
Operation

During controlled venting the vent valve modulates opening so pump deceleration stays between 3 and 5 Hz/s, adjusted every 10 seconds.

Explains the deceleration limit that protects the pump during venting.

Source
“the pump deceleration quickness is maintained in the range between 3 and 5 Hz/s. The timing is adjusted every 10 seconds.”