ADM Flow Meter Operation Manual

Contents

  1. Warning p. 2
    Legal and safety notices for the instrument
  2. Overview pp. 3-5
    Product overview, features, and controls
  3. Procedure pp. 6-7
    Button functions and power-on procedure
  4. Warning pp. 8-9
    Notes, cautions, and warnings for safe use
  5. Overview p. 10
    How the flow meter measures flow
  6. Spec pp. 11-12
    Product specifications and display details
  7. Procedure pp. 13-15
    Split mode, auto-off, and mode selection
  8. Spec pp. 16-19
    Tables of cartridge and mainframe information
  9. Procedure pp. 20-22
    Flow measurement and storing results
  10. Procedure pp. 23-24
    Reading and setting a split ratio
  11. Overview p. 25
    Considerations about flow measurements
  12. Procedure pp. 26-27
    Connecting and using the USB interface
  13. Warning pp. 28-30
    Warning, error, battery, and power indications
  14. Procedure pp. 31-34
    Installing tubing and replacing batteries
  15. Procedure p. 35
    Upgrading firmware and retrieving data
  16. pp. 36-37
    Using the stand and backpressure notes
  17. pp. 38-39
    Disposal instructions and blank page
  18. p. 11 table, 8 rows
  19. p. 4 table, 16 rows
  20. p. 12 table, 3 rows
  21. p. 16 table, 6 rows
  22. p. 17 table, 7 rows

Table 1 Agilent ADM flow meter specifications

ParameterValue
Flow range0.5 to 750 mL/min, auto-ranging
Accuracy*: 0.5 to 500 mL/min±2% of reading, or ±0.2 mL/min, whichever is greater.
Accuracy*: 501 to 750 mL/min±3% of reading
Tubing Temperature Range–62 to 110 °C
Operating Temperature Range0 to 45 °C
Storage Temperature Range†–15 to 50 °C
Power(3) AA batteries (alkaline), or USB power
Display128 × 64 pixels monochrome with 16 gray scales

* When verified at ambient temperature of 20 to 24 °C at sea level. Optimum performance is in the range of 15 to 35 °C. Accuracy may decrease outside of these ranges. † Damage to the instrument may occur if it is exposed to temperatures greater than 50 °C, leading to inaccurate results.

Table

РазделСтраница
Warning indication28
Error indication28
Low battery indication29
Power Indications30
Installing and Replacing the Cartridge31
Installing the cartridge31
Swapping or replacing the cartridge with ADM flow meter cartridge31
Replacing the Batteries33
Upgrading the Firmware/Retrieving Certificate of Calibration35
Using the Flow Meter Stand36
A Note About Backpressure37
Technical Support37
Other Features37
Safety and Regulatory Certification38
Class A EMC certification38
Disposal38

Flow (mL/min) / Display Resolution (mL/min)

Flow (mL/min)Display Resolution (mL/min)
0.5 to 9.990.01
10.0 to 99.90.1
100 to 7501.0

Table 2 Error and warning code descriptions

Error/ Warning CodeTypeDescriptionTroubleshooting
01 No Cal InfoErrorNo calibration informationContact Agilent Technical Support
02 No VerifErrorFlow meter is not verifiedContact Agilent Technical Support
03 PresTemp ErrErrorPressure/ temperature sensor failureContact Agilent Technical Support
04 No CartErrorNo cartridge detectedInstall a valid cartridge.
05 No MF InfoWarningNo mainframe informationContact Agilent Technical Support
06 Auto RecoverWarningAuto-recovery modeReboot the flow meter

Table 2 Error and warning code descriptions (continued)

Error/ Warning CodeTypeDescriptionTroubleshooting
07 OverflowWarningMeasurement exceeds instrument rangeMeasure gas flow within the acceptable flow range
08 Cart ExpWarningCartridge expiredInstall a new cartridge
09 Low RTC BatWarningRTC (real time clock) backup battery lowContact Agilent Technical Support
10 Op TempErrorOut of operating temperature range condition (0 to 45 °C)Operate the flow meter within the acceptable temperature range
11 Op PresErrorOut of operating atmospheric pressure rangeOperate the flow meter within the acceptable atmospheric pressure range
12 Ver MismatchErrorFirmware version mismatch with cartridge versionUpgrade firmware to latest version
13 Invalid CartErrorUnable to identify cartridge typesContact Agilent Technical Support

Engineering Facts

13 facts · every claim linked to a page
Safety

Flows above 750 mL/min can destroy the cartridge and void the warranty, so the measured flow must be kept within the instrument range.

Prevents cartridge damage and warranty loss during flow measurement.

Source
“Flows exceeding 750 mL/min may damage the cartridge and void the warranty.”
Safety

Only a 5 V DC, 0.5 A USB source and the supplied USB cable may be used to run the device.

Ensures correct and safe USB powering of the meter.

Source
“Use only a 5 V DC, 0.5 A USB power source and the supplied USB cable to run this device.”
Safety

The meter must not be used as safety equipment.

Defines the instrument as a measurement tool, not a protective device.

Source
“Do not use the ADM flow meter as safety equipment.”
Selection

The flow range is 0.5 to 750 mL/min with auto-ranging.

Confirms the meter covers the intended gas flow range.

Source
“0.5 to 750 mL/min, auto-ranging”
Performance

Accuracy is ±2% of reading or ±0.2 mL/min (whichever is greater) from 0.5 to 500 mL/min, and ±3% of reading from 501 to 750 mL/min.

Sets the measurement uncertainty to expect at low and high flows.

Source
“±2% of reading, or ±0.2 mL/min, whichever is greater.”
Operation

The meter must always be switched on before it is connected to a flow source.

Avoids measurement errors and potential cartridge stress during connection.

Source
“Always turn the flow meter on before connecting it to a flow source.”
Operation

Readings stabilize in 1 to 2 seconds, but flows below 1 mL/min may need up to 5 seconds.

Sets the waiting time before recording a valid low-flow measurement.

Source
“Low flows (less than 1 mL/min) may take up to 5 seconds for the reading to stabilize.”
Maintenance

The cartridge calibration is valid for one year from first use, and the meter still functions after expiry with a warning.

Plans cartridge replacement and calibration cycles without downtime.

Source
“The cartridge calibration is valid for a period of one year from its first use.”
Maintenance

The cartridge is replaced by loosening two captive thumbscrews and lifting it out of the mainframe.

Describes the field-replaceable cartridge procedure for maintenance.

Source
“Loosen both captive thumbscrews completley and lift the cartridge out of the CrossLab CS Mainframe body.”
Operation

USB power does not recharge the installed batteries.

Prevents expecting battery recharge when running on USB power.

Source
“Using the USB port to supply power does not recharge the installed batteries.”
Operation

In USB power mode the Auto off function is always off and cannot be changed.

Clarifies power behavior when the meter is USB-powered.

Source
“In USB power mode, the Auto off function is always off and the option to change the Auto off setting is disabled.”
Operation

When connected to PC software, the meter buttons are disabled.

Explains why front-panel control is unavailable during data acquisition.

Source
“When the flow meter is connected to PC software, the buttons on the flow meter are disabled.”
Performance

Soap bubble meters typically read slightly higher, by nearly 4% at room temperature due to water vapor.

Explains expected discrepancy versus bubble-type flow meters.

Source
“At room temperature, water vapor can raise readings by nearly 4%.”