Technical Reference Summary

Covered Models & Document Scope

This document (715001557 Rev. D) covers the installation of system leak sensors in Waters ACQUITY instruments. It applies to all ACQUITY components that accept a leak sensor; the exact sensor location varies by module and is documented in the relevant module user guide.

Section-by-Section Technical Breakdown

pp. 1
Installing System Leak Sensors

This procedure describes how to install leak sensors in Waters ACQUITY instruments. The sensor fits into a receptacle in the component's drip tray, and the cable connector plugs into a connector on the component bulkhead. Installation requires powering down all ACQUITY components, installing sensors one-by-one, then powering up with the sample organizer last. The leak sensor must be enabled in the software after installation.

  • Leak sensor locations vary by module; refer to the relevant user guide for module-specific location.
  • Sensor mounts in the drip tray receptacle and snaps into position.
  • Ribbon cable assembly plugs into the bulkhead connector mounted on the chassis.
  • Remove power from all ACQUITY components before installation.
  • Open all system component doors to locate bulkhead connectors.
  • Install sensors one-by-one by pushing the sensor into its receptacle, then plugging the ribbon cable assembly into the bulkhead connector.
  • After installation, turn the computer on, log onto Windows, power up ACQUITY components (sample organizer last if used), log onto the data system, and enable the leak sensor in the software.
  • For software enabling instructions, see Leak Sensor Maintenance Instructions, 71500082506.

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication Applies To
Leak sensor mounts in a receptacle in the component's drip tray and snaps into position. p.1 ↗ The sensor must be physically seated in the drip tray to detect leaks. ACQUITY instruments
The sensor cable connector plugs into a bulkhead connector mounted on the chassis. p.1 ↗ The ribbon cable assembly must be routed to and plugged into the bulkhead connector. ACQUITY instruments
Power must be removed from all ACQUITY components before installing leak sensors. p.1 ↗ Ensure all modules are powered down before beginning installation. ACQUITY instruments
All system component doors must be opened to locate the connectors on each bulkhead. p.1 ↗ Access to bulkhead connectors requires opening component doors. ACQUITY instruments
Install leak sensors one-by-one by pushing the sensor into its receptacle, then plugging the ribbon cable assembly into the bulkhead connector. p.1 ↗ Follow the sequence to ensure each sensor is properly connected. ACQUITY instruments
After installation, turn the computer on and log onto Windows. p.1 ↗ The computer must be started before powering up the ACQUITY components. ACQUITY instruments
Power up the ACQUITY components, turning the sample organizer on last, if one is used. p.1 ↗ The sample organizer must be the last component powered on. ACQUITY instruments with sample organizer
Log onto your data system and enable the leak sensor in the software. p.1 ↗ The leak sensor will not function until enabled in the software. ACQUITY instruments
For instructions on enabling the leak sensor in software, see Leak Sensor Maintenance Instructions, 71500082506. p.1 ↗ Refer to document 71500082506 for software enabling details. ACQUITY instruments
Leak sensor locations vary by module; refer to the relevant user guide in the Waters Support Library for the specific module. p.1 ↗ Module-specific documentation must be consulted for exact sensor location. ACQUITY instruments