Induztr
Engineering Digest Document Viewer Related Models

XGS-600 Gauge Controller

Technical Reference Summary

Covered Models & Document Scope

Covers the XGS-600 vacuum gauge controller with six gauge board slots. Supported board types: HFIG (Hot Filament Ion Gauge), IMG (Inverted Magnetron Gauge), Convection (TC/ConvecTorr), and Analog Input boards. Field-installable board part numbers and rack-mount kits are listed. Backwards compatible with SenTorr and MultiGauge instruments.

Section-by-Section Technical Breakdown

pp. 41
Board Configuration Rules

Defines which sensor boards can be installed in which of the six XGS-600 slots. Slots 1-4 are long high-profile slots accepting any board including HFIG; slots 5-6 are short low-profile slots accepting only analog, convection, or IMG boards with specific restrictions.

  • Slots are numbered left to right as viewed from the front.
  • Slots 1-4 are Long, High Profile and can accommodate any board; HFIG boards can only be installed in slots 1-4.
  • Slots 5 and 6 are Short, Low Profile and accept only analog, convection, or IMG boards.
  • An IMG board may NOT be installed in slot 5 if any other board is installed in slot 4.
  • An IMG board may NOT be installed in slot 6 if any other board is installed in slot 5.
pp. 15-18
Installation and Connection

Covers unpacking, ferrite bead installation on the AC mains cable, cable connections per board type, and required ventilation clearance. Includes accessory and power cord part numbers.

  • Install the included clamp-on ferrite bead onto the AC power cable near the chassis with the locking tab fully engaged.
  • Keep power off while connecting gauge and I/O cables.
  • Allow approximately 3/4 inch clearance at each side of the enclosure for proper air flow; do not block air vents.
  • When installing an HFIG cable, connect the pigtails from the ground cable to the ground stud.
  • The Analog Board is identified by Dsub connector pin contacts versus the Convection Board socket contacts.
pp. 19-22
Gauge Selection and Compatibility

Provides a gauge selection chart by vacuum regime and controller card type, plus cross-reference tables for competitor BA gauges and compatible active gauges. Helps select the correct sensor model setting for non-Agilent gauges.

  • Rough vacuum (ATM to 10-3 Torr): ConvecTorr, 536, 531, CT-100, PVG Series, PCG Series, CDG500.
  • High vacuum (10-3 to 10-9 Torr): 572, 563, 571, MBA-100, MBA-200/MBA2, IMG-100, FRG700/702, FRG720/730.
  • Ultra-high vacuum (10-9 to 10-12 Torr): IMG-300, UHV-24, UHV-24p.
  • BA cross-reference maps competitor gauge models to Agilent sensor model settings 572, 571, UHV24, 563, or 564.
  • Compatible Agilent active gauges include PVG500/502, PCG750/752, FRG700/702, FRG720/730, CDG500, and CT-100.
pp. 23-40
Operating Instructions

Describes screen navigation, main screen layout, soft keys, gauge control options, degas operation, calibration procedures, and setpoint configuration. Includes Auto-On indicator behavior and user label rules.

  • Main screen shows installed sensors, real-time pressure readings, setpoint status, and soft keys SETPTS, SETUP, LARGER.
  • Hot filament gauge control options: OFF, FIL1 ON, FIL2 ON, DEGAS 1 ON, DEGAS 2 ON, DEGAS OFF.
  • Degas requires emission on and pressure below 1E-5 Torr; degas times out after approximately 25 minutes.
  • Only two hot filament gauges can degas simultaneously.
  • Calibration uses Agilent Smart Cal to automatically determine vacuum or atmosphere calibration; allow 15 minutes stabilization after first power up.
  • Setpoint ON pressure must be lower than OFF pressure or ERR appears; once valid, ERR changes to AUTO.
  • Setpoint values are not restricted by gauge measurement range; do not input a value < 1E-4 for any gauge whose measurement range is > 1E-4.
pp. 41-49
Service and Troubleshooting

Provides error code definitions, troubleshooting tables for symptoms and causes, and guidance on board replacement and configuration. Includes tips on using reference gauges and simulators for fault isolation.

  • Error codes include BD COM, GRIDLO, HITEMP, NOFIL1, NOFIL2, Open, P>MAX, No CBL, Fault, and Under.
  • Troubleshooting table covers symptoms such as hot filament gauge hunting around 1E-5 Torr, IMG reading -11 Torr range, HFIG reading UHV when pressure is higher, setpoint ERR, duplicate user label, and communication failures.
  • Reference ionization B/A gauges sealed at approximately 5E-6 Torr and a Convection gauge simulator are available for troubleshooting but are not NIST traceable.
  • A BDCOMM message that persists requires cycling AC mains power to recover.
  • Display blanking due to EMI recovers automatically via screen refresh every 20 minutes or by cycling AC power.
pp. 67-82
Serial Commands and Backwards Compatibility

Appendix B and C describe ASCII and BCD serial command sets, including pressure data format and error code transmission in MG/SenTorr format. Appendix C also details emulation of the MultiGauge 4-channel TC card using two convection cards.

  • Pressure information is transmitted as x.xxxExx using three bytes: two mantissa bytes and one exponent byte in two's complement.
  • Example: 7.600E+2 is transmitted as {76h} {00h} {02h}; 2.145E-7 as {21h} {45h} {F9h}; OFF as {00h} {00h} {00h}.
  • Error codes are transmitted in MG/SenTorr format; NO FIL1 is reported as E05 = {0Eh} {00h} {05h}.
  • Two convection cards in physically adjacent slots emulate a 4-channel TC card for backwards compatibility.
  • BCD base address mapping depends on XGS-600 slot number.

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication Applies To
Setpoint relay maximum rating is 30 VDC at < 2 A. p.13 ↗ Do not exceed this rating when connecting external loads to the relay outputs; use appropriate suppression for inductive loads. whole series
The XGS-600 processes a signal in less than 20 milliseconds. p.11 ↗ Suitable for fast-response control and interlock applications requiring sub-20 ms update. whole series
HFIG boards can only be installed in slots 1 through 4; slots 5 and 6 accept only analog, convection, or IMG boards. p.41 ↗ Plan board layout before installation; HFIG in slots 5 or 6 is not supported. whole series
An IMG board may NOT be installed in slot 5 if any other board is installed in slot 4, and may NOT be installed in slot 6 if any other board is installed in slot 5. p.41 ↗ Avoid adjacent IMG and other board combinations in slots 4-6 to prevent configuration conflicts. whole series
Allow approximately 3/4 inch clearance at each side of the enclosure for proper air flow. p.15 ↗ Ensure rack or bench installation provides adequate ventilation to prevent HITEMP errors. whole series
The included clamp-on ferrite bead must be installed onto the AC power cord near the chassis with the locking tab fully engaged. p.8 ↗ Required for compliance with FCC Part 15 and EU EMI requirements; failure to install may cause radiated emissions non-compliance. whole series
When using glass BA gauges, supplemental shielding via a grounded metal cage or screen may be required to meet radiated emission limits. p.8 ↗ Install grounded metal cage around glass BA gauge while allowing air flow to reduce emissions and protect against breakage. whole series with glass BA gauges
Degas requires emission on and pressure below 1E-5 Torr; degas times out after approximately 25 minutes. p.27 ↗ Pump down before degas; the controller automatically reverts to FIL1 or FIL2 after timeout. HFIG boards
Only two hot filament gauges can degas at the same time. p.28 ↗ Schedule degas operations sequentially when more than two HFIG gauges are installed. HFIG boards
Setpoint ON pressure must be lower than OFF pressure; otherwise ERR appears in the rightmost column. p.30 ↗ Enter valid ON/OFF pressure pairs to enable setpoint operation; ERR changes to AUTO once valid. whole series
Setpoint values are not restricted by the gauge measurement range; do not input a value < 1E-4 for any gauge whose measurement range is > 1E-4. p.30 ↗ Avoid setting unrealistic setpoints outside the sensor's usable range to prevent unreliable switching. whole series
For setpoint delays of 0.0 or 0.1 seconds, the XGS-600 uses a less accurate pressure measurement to obtain the fastest response. p.48 ↗ Expect slightly different trigger/release pressures when using very short setpoint delays. whole series
Convection gauge calibration at atmosphere depends on orientation; install the gauge sensor in final orientation before calibrating. p.28 ↗ Calibrate after final mechanical installation to avoid offset errors. Convection boards
Allow 15 minutes stabilization after first power up before performing calibration. p.28 ↗ Schedule calibration after warm-up to ensure accurate readings. whole series
Vacuum calibration defaults to 1E-4 Torr and atmosphere calibration defaults to 760 Torr (or equivalents in mbar and Pa). p.28 ↗ Regional atmospheric pressure can be modified in the System Setup screen. whole series
User Labels are limited to 5 characters and may not begin with HFIG, CNV, IMG, or AUX. p.32 ↗ Choose unique labels that do not conflict with system-assigned sensor IDs. whole series
The XGS-600 refreshes the entire screen approximately every 5 seconds; a blank followed by refill is normal. p.48 ↗ Do not interpret periodic screen blanking as a fault; no action required. whole series
A persistent BDCOMM message requires cycling AC mains power OFF and ON to recover. p.49 ↗ If BDCOMM remains after power cycle, the unit may require repair. whole series
Pressure data is transmitted in BCD format as x.xxxExx using three bytes: two mantissa bytes and one exponent byte in two's complement. p.78 ↗ When implementing serial communication, decode mantissa and exponent accordingly; example 7.600E+2 = {76h}{00h}{02h}. whole series with BCD serial option
Two convection cards installed in physically adjacent slots emulate a MultiGauge 4-channel TC card for backwards compatibility. p.78 ↗ When replacing a MultiGauge system, install convection cards in adjacent slots to preserve channel addressing. whole series with convection boards

Faults & Troubleshooting

Code / Symptom Possible Cause Remedy Note
BD COM p.46 ↗ Motherboard and sensor board stopped communicating 1. Cycle power to unit 2. Return unit for repair Table 4-1 General Error Codes
GRIDLO p.46 ↗ Pressure is too high causing a glow discharge; HFIG board failure 1. Check BA gauge for grid to filament short or grid to housing short 2. Replace gauge HFIG bds only; Table 4-1
HITEMP p.46 ↗ Fan failure; XGS ventilation openings blocked; Internal temperature above 75° C 1. Allow unit to cool 2. Check that fan is operating 3. Ensure that there is adequate ventilation around the unit HFIG and IMG bds only; Table 4-1
NOFIL1 p.46 ↗ Filament failure; Ion gauge cable not connected 1. Use second filament 2. Replace BA gauge 3. Check that cable is installed HFIG bds only; Table 4-1
NOFIL2 p.46 ↗ Filament failure; Ion gauge cable not connected 1. Replace BA gauge 2. Check internal sensor board connections 3. Check that cable is plugged in HFIG bds only; Table 4-1
Open p.46 ↗ Cable not connected; Sensor failure 1. Check cable 2. Replace sensor CNV bds only; Table 4-1
P>MAX p.46 ↗ System vented from vacuum with gauges on; Attempted to turn gauges on with vacuum near overpressure limit 1. Turn gauges off before venting system or use AUTO-ON 2. Wait for vacuum to improve before turning on ion gauges HFIG bds only; Table 4-1
No CBL p.46 ↗ Cable not connected to XGS-600 1. Check cable Analog boards only; Table 4-1
Fault p.47 ↗ Gauge defective; Cable defective; Gauge drifted 1. Check cable 2. Check active gauge Replace if necessary Analog boards only; active gauge voltage < 0.5 V; Table 4-1
Under p.47 ↗ CDG has drifted 1. Press CAL 2. Readjust zero at CDG Analog boards only; CDG voltage negative or below previous VAC CAL; Table 4-1
Hot filament gauge hunts or cycles around 1E-5Torr (1E-3Pa) p.47 ↗ Auto emission adjustment occurs at this pressure. Increased filament power causes a rise in pressure. Condition will clear as system pressure decreases. If taking pressure measurements in this range, reduce emission current setting for the gauge to 1.0 mA or less to turn off feature Table 4-2 Troubleshooting
IMG displays readings in the –11 Torr range when the pressure is known to be higher p.47 ↗ IMG cable not plugged in; IMG has not started; IMG took a long time to start 1. Check that cable is attached 2. Sometimes IMG’s are hard to start when turned on at high vacuum 3. Raise pressure 4. Clean or replace IMG Table 4-2 Troubleshooting
HFIG bd reads UHV when pressure is know to be higher p.47 ↗ Emission current not established due to worn out Th-Ir filament; Collector cable unplugged; HFIG board failure 1. Check collector connection at gauge (glass BA) and at XGS-600 2. Replace gauge or filament 3. Return unit for repair/replace Table 4-2 Troubleshooting
ERR appears in rightmost column of setpoint screen p.48 ↗ Setpoint OFF pressure is < than ON pressure Correct pressure entries so the setpoint ON pressure is less than the OFF pressure Table 4-2 Troubleshooting
Duplicate message appears when trying to enter a User Label p.48 ↗ Label is the same as another User Label Create a different entry Table 4-2 Troubleshooting
User Label not accepted p.48 ↗ User input violates User Label rules See “Sensor Setup Screen” on page 31 for User Label rules Table 4-2 Troubleshooting
Cannot turn HFIG or IMG on using the keypad p.48 ↗ Auto-ON programmed for that gauge; The keypad is locked out 1. This is the correct response 2. If manual control is desired, turn off AUTO-ON 3. Unlock keypad Table 4-2 Troubleshooting
Cannot turn on Degas p.48 ↗ Pressure is above 1E-5T; Attempted to turn on Degas2 while running Fil1; Attempted to degas third gauge 1. Pump down before degas 2. Change operation to Fil2 before attempting to degas that filament 3. XGS-600 will only degas two gauges at a time 4. Finish a degas before attempting to start a third Table 4-2 Troubleshooting
The front panel display indicates the setpoint triggers/releases at a slightly different pressure than what is programmed p.48 ↗ Time delay = 0.0, 0.1; XGS-600 uses a less accurate pressure measurement for setpoint delays of <= 0.1 secs to obtain the fastest response No action specified Table 4-2 Troubleshooting
Changes to baud rate, comm. mode, units, keypad lockout, or ATM value do not take effect p.48 ↗ Did not exit System Setup screen after making changes After making changes, exit the screen for the changes to take effect Table 4-2 Troubleshooting
Unit does not respond to RS232 or RS485 commands p.48 ↗ XGS600 in wrong serial mode or incorrect baud rate; Incorrect wiring; Incorrect terminal settings 1. Change serial comm to RS232 or RS485, select proper baud rate, set RS485 address in General Setup screen 2. Check wiring 3. Check that terminal has correct # of data, start/stop bits, etc. Table 4-2 Troubleshooting
Display goes blank followed by a refill of the screen. Operation is otherwise normal. p.48 ↗ Screen refreshes approximately every 5 seconds; XGS-600 refreshes the entire screen periodically to update. No action required. Table 4-2 Troubleshooting
One or more ion gauge channels (HFIG and IMG) appear to restart, sometimes a BDCOMM message appears first p.49 ↗ A major EMI disturbance or other anomaly has caused a sensor board watchdog timeout This is how the XGS-600 recovers from hard and soft errors. No action required. Table 4-2 Troubleshooting
BDCOMM message appears on one or more channels and stays p.49 ↗ A major EMI disturbance or other anomaly has caused a non-recoverable software error Cycle AC (mains) power to the unit OFF and ON again to recover Table 4-2 Troubleshooting
IMG reads 10 mT p.49 ↗ Pressure is greater than 10 mT Wait for pressure to drop into operating range or use Auto-On feature to control IMG Table 4-2 Troubleshooting
Display goes blank p.49 ↗ A major EMI has caused a display or software error Cycle AC (mains) power to the unit OFF and ON again to recover, or wait for the automatic screen refresh (every 20 minutes) Table 4-2 Troubleshooting
Convection or TC readings fluctuate when first at HI VAC but then quiet down p.49 ↗ This is normal behavior of the Convectorr and TC gauges when doing a pumpdown. None Table 4-2 Troubleshooting
Convection or TC readings fluctuate in excess of 1 mT continuously when at Hi VAC p.49 ↗ Possible defective gauge. Replace gauge Table 4-2 Troubleshooting
Convection or TC readings unstable at atmosphere when measuring Helium p.49 ↗ This is normal performance for thermal loss gauges in helium Refer to “Helium Measurement Capability” on page 36 for performance limits when measuring helium Table 4-2 Troubleshooting

Technical Data

Parameter Value Condition
Setpoint relay maximum rating 30 VDC p.13 ↗ < 2 A
Signal processing time < 20 ms p.11 ↗
Maximum gauges displayed on one screen 8 gauges p.11 ↗
Maximum ion gauges per unit 5 ion gauges p.11 ↗
Maximum convection or Agilent gauges per unit 12 gauges p.11 ↗
Number of set points 8 set points p.11 ↗
Internal over-temperature threshold 75 °C p.46 ↗ HITEMP error
Degas timeout approximately 25 minutes p.27 ↗
Degas maximum pressure 1E-5 Torr p.27 ↗ pressure must be below this value
Vacuum calibration default 1E-4 Torr p.28 ↗
Atmosphere calibration default 760 Torr p.28 ↗
Convection/TC vacuum reading fluctuation after calibration 1.0E-4 to ~3.0E-4 Torr p.28 ↗ occasional reading up to 1.0E-3 Torr
IMG-100 measurement range 2 x 10-9 Torr p.20 ↗
IMG-300 measurement range 1 x 10-11 Torr p.20 ↗
UHV-24 measurement range 1 x 10-10 Torr p.20 ↗
UHV-24p measurement range 2 x 10-11 Torr p.20 ↗
572/563 measurement range 1 x 10-9 Torr p.19 ↗
IMG-300 bakeout temperature with magnet 250 °C p.20 ↗
IMG-300 bakeout temperature without magnet 450 °C p.20 ↗
UHV-24 bakeout temperature with magnet 250 °C p.20 ↗
UHV-24/UHV-24p PTFE cable bakeout temperature 250 °C p.16 ↗
Required side clearance for air flow approximately 3/4 inch p.15 ↗ each side of enclosure
Screen refresh interval approximately 5 seconds p.48 ↗
Automatic screen refresh interval after EMI blanking every 20 minutes p.49 ↗
Setpoint delay accuracy threshold 0.0, 0.1 seconds p.48 ↗ less accurate pressure measurement used for delays <= 0.1 s
User Label maximum length 5 characters p.32 ↗
Maximum set points per sensor 2 set points p.29 ↗
Maximum simultaneous degas gauges 2 hot filament gauges p.28 ↗
Reference ionization gauge sealed-off pressure approximately 5E-6 Torr p.47 ↗ 6.6E-4 Pa