Application Note
Technical Reference Summary
Covered Models & Document Scope
Nota de aplicación sobre el espectrómetro FTIR portátil Agilent 4100 ExoScan para medir el espesor de imprimación epoxi sobre aluminio; también menciona la variante FTIR relacionada 4200 FlexScan (mismos componentes ópticos, óptica/electrónica separadas por cable).
Technical Data
| Parameter | Value | Condición |
|---|---|---|
| Calibration band center | 1610 cm-1 p.3 ↗ | epoxy primer thickness calibration |
| Absorbance of thinnest sample | approximately 0.07 absorbance units p.3 ↗ | 1610 cm-1 band |
| Absorbance of thickest sample | 0.7 absorbance units p.3 ↗ | 1610 cm-1 band |
| Linear range of IR absorbance | typically 0.05 to 1 absorbance units p.3 ↗ | |
| Alternative stronger band | 1510 cm-1 p.3 ↗ | usable for thicker calibration ranges |
| Alternative weaker band | 1285 cm-1 p.3 ↗ | usable for thinner calibration ranges |
| Calibration thicknesses | 0.06, 0.09, 0.17, 0.30 and 0.39 mils p.3 ↗ | epoxy primer calibration samples |
| Optical head weight | 3 pound p.3 ↗ | 4200 FlexScan |
| Battery and electronics pack weight | 4 pound p.3 ↗ | 4200 FlexScan |
Operating Limits
- Linear range of IR absorbance: typically 0.05 to 1 absorbance units p.3 ↗