Automated Plasma Processing for Quantitative, Targeted LC/MS Analysis
Technical Reference Summary
Covered Models & Document Scope
Nota de aplicación sobre el procesamiento automatizado de plasma para el análisis cuantitativo dirigido de proteínas mediante LC/MS utilizando la plataforma Agilent Bravo Automated Liquid Handling Platform; compara la digestión automatizada con tripsina en placas de 96 pocillos con la digestión manual en tubos de microcentrífuga.
Technical Data
| Parameter | Value | Condición |
|---|---|---|
| Samples processed per day | up to 96 samples/day p.1 ↗ | Bravo Platform, automated |
| Shaking speed | 800 rpm p.2 ↗ | digestion |
| Digestion time | 30 minutes p.2 ↗ | 37 °C |
| Digestion temperature | 37 °C p.2 ↗ | |
| TCEP concentration | 25 mM p.2 ↗ | reduction |
| IAA concentration | 80 mM p.2 ↗ | alkylation |
| Centrifugation speed | 5,000 x g p.3 ↗ | 5 minutes |
| Reconstitution solvent | 3 % acetonitrile, 5 % acetic acid p.3 ↗ | 3 % ACN, 5 % HOAc |
| Injection volume | 1 μL p.3 ↗ | HPLC-Chip |
| Median recovery (PAR) | 50 % p.3 ↗ | manual (tubes) |
| Median precision (CV) | 5.9 % p.3 ↗ | manual (tubes) |
Operating Limits
- maximum samples per day: 96 samples/day (automated) vs 24 samples/day (manual) p.1 ↗
- digestion temperature: 37 °C p.2 ↗
- centrifugation: 5,000 x g for 5 minutes p.3 ↗