7000E/7010C Series Triple Quadrupole GC/MS Operating Manual

Contenido

  1. Overview pp. 1-6
    Manual introduction, where to find more information
  2. Procedure pp. 7-12
    Installing and disassembling the EI HES source and interface
  3. Warning pp. 13-30
    Safety warnings, hydrogen accumulation, symbols, moving the MS
  4. Procedure pp. 31-99
    Installing capillary and Intuvo columns, gaskets, conditioning, calibration
  5. Procedure pp. 100-126
    Pumping down, vacuum pressure, analyzer temperatures, CC gas configuration
  6. Procedure pp. 127-147
    Operating the CI MS, carbon dioxide CI, flow control module, CI autotune
  7. Procedure pp. 148-216
    Vacuum system, analyzer chambers, source removal and assembly, filaments, EM horn
  8. Procedure pp. 217-237
    Switching between EI XTR, EI HES, and CI sources
  9. Procedure pp. 238-250
    CI source maintenance and filament installation
  10. p. 131 table, 12 rows
  11. p. 34 table, 8 rows
  12. p. 54 table, 7 rows
  13. p. 78 table, 8 rows
  14. p. 110 table, 3 rows
  15. p. 132 table, 4 rows
  16. p. 149 table, 12 rows
  17. p. 168 table, 4 rows
  18. p. 169 table, 14 rows
  19. p. 171 table, 8 rows

Table 11 CI tune default settings

ParameterMethane PositiveMethane NegativeIsobutane PositiveIsobutane NegativeAmmonia PositiveAmmonia Negative
Ion polarityPositiveNegativePositiveNegativePositiveNegative
Emission150 µA50 µA150 µA50 µA150 µA50 µA
Electron energy150 eV150 eV150 eV150 eV150 eV150 eV
Filament111111
Repeller3 V3 V3 V3 V3 V3 V
Ion focus130 V130 V130 V130 V130 V130 V
Entrance lens offset20 V20 V20 V20 V20 V20 V
EM volts120014001200140012001400
Shutoff valveOpenOpenOpenOpenOpenOpen
Gas selectAABBBB
Suggested flow20%40%20%40%20%40%
Source temp250 °C150 °C250 °C150 °C250 °C150 °C

Table 3 Gas flows

FeatureGas flows
High-vacuum pumpSplit flow turbo
Optimal He column gas flow, mL/min (carrier gas)1 to 2
Reagent gas flow, mL/min1 to 2
CC gas flow3 to 4
Maximum recommended gas flow, mL/min*4 to 6
Maximum gas flow, mL/min†6.5
Maximum column id0.53 mm (30 m length)
JetClean option H2 flow, 0.4 mL/min

* Total gas flow into the MS = column flow + CC gas flow + reagent gas flow (if applicable) + JetClean H2 flow (if applicable). † Expect degradation of spectral performance and sensitivity.

Table 4 Gas flows

FeatureGas flows
High-vacuum pumpSplit flow turbo
Optimal He column gas flow, mL/min (carrier gas)1 to 2
CC gas flow3 to 4
Maximum recommended gas flow, mL/min*4 to 6
Maximum gas flow, mL/min†6.5
Maximum column id0.53 mm (30 m length)
JetClean option H2 flow0.4 mL/min

* Total gas flow into the MS = column flow + CC gas flow + JetClean H2 flow (if applicable); † Expect degradation of spectral performance and sensitivity

Table 5 Gas flows

FeatureGas flows
High-vacuum pumpSplit flow turbo
Optimal He column gas flow, mL/min (carrier gas)1 to 2
Reagent gas flow, mL/min1 to 2
CC gas flow3 to 4
Maximum recommended gas flow, mL/min*4 to 6
Maximum gas flow, mL/min†6.5
Maximum column id0.53 mm (30 m length)
JetClean option H2 flow, 0.4 mL/min

* Total gas flow into the MS = column flow + CC gas flow + reagent gas flow (if applicable) + JetClean H2 flow (if applicable). † Expect degradation of spectral performance and sensitivity.

Table 8 Recommended temperature settings

ZoneEI operation
MS Source230 °C default. High limit 350 °C. Application dependent.
MS Quad 1150 °C
MS Quad 2150 °C

Table 11 CI tune default settings (continued)

ParameterMethaneIsobutaneAmmonia
Front quad temp150 °C150 °C150 °C
Rear quad temp150 °C150 °C150 °C
Interface temp280 °C280 °C280 °C
AutotuneYesYesNo

Table 14 Maintenance schedule

TaskFrequency
Check the foreline pump oil levelEvery week
Check the calibration vial(s)Every 6 months
Replace the foreline pump oil*Every 6 months
Replace dry scroll pump tip sealsEvery year
Tune the MSAs needed
Check the foreline pumpAs needed
Clean the ion sourceAs needed
Check the carrier gas trap(s) on the GC and MSAs needed
Replace the worn out partsAs needed
Lubricate side plate or vent valve O-rings†As needed
Replace GC gas suppliesAs needed
Leak check the systemAs needed

* Or as needed. † Vacuum seals other than the side plate O-ring and vent valve O-ring do not need to be lubricated. Lubricating other seals can interfere with their correct function.

Table 15 Parts list for EI HES (Figure 51)

Item numberItem descriptionPart number
1Source finger gripG7002-20008
2Filament blockG7002-20019
3Extractor lens (5)*, with 3 mm openingG7004-20061
4Ceramic insulator for extractorG7002-20064

Table 15 Parts list for EI HES (Figure 51) (continued)

Item numberItem descriptionPart number
5Entrance lens assembly, Extended, HES (1)*Gxxxx-xxxxx
6Ion focus lens (2)*G7004-20068
7Lens insulator/holderG7002-20074
8M2 x 0.4 screw x 12 mm long gold plated screwG7002-20083
9Source bodyG7002-20084
10Post extractor lens 2 (3)*G7004-20090
11Post extractor lens 1 (4)*G7004-20004
12M2 x 6 mm gold plated screwG7002-20109
13Locking ring lens insulatorG7002-20126
14High efficiency dual filamentG7002-60001
15Ring heater/sensor assemblyG7002-60043
16Source mount 1.5 mmG7002-60053
17Repeller assemblyG7002-60057
Not shownHES assemblyGxxxx-xxxxx

* The number in parenthesis is the number engraved on the lens

Table 16 Parts list for EI XTR source (Figure 52)

ItemDescriptionPart number
1SetscrewsG3870-20446
2ScrewsG3870-20021
3Source bodyG3870-20440
4Extractor lensG3870-20444
5Extractor lens insulatorG3870-20445
6FilamentsG7005-60061
7Spring washer3050-1301
7Flat washer3050-0982

Engineering Facts

14 facts · every claim linked to a page
Selection

The TQ GC/MS is a standalone capillary GC detector compatible with Agilent 8890, 9000, or 7890 GCs, so the GC platform must be chosen accordingly.

Confirms GC compatibility before selecting the MS for an existing GC.

Source
“The 7000E/7010C Series TQ GC/MS is a standalone capillary GC detector for use with the Agilent 8890, 9000, or 7890 GCs.”
Selection

The mainframe weighs 59 kg (64 kg with CI) and the foreline pump adds 22.2 kg, so bench/site load capacity must be planned.

Needed for lab bench structural and handling planning during installation.

Source
“The weight is 59 kg for the turbo pump mainframe and 64 kg for the mainframe with CI. The attached foreline (roughing) pump weighs an additional 22.2 kg.”
Operation

Switching between CI and EI sources takes about 1 hour plus a 1- to 2-hour purge/bakeout, and PCI-to-NCI needs about 2 hours for source cooling.

Sets realistic method-changeover time and throughput expectations.

Source
“Switching back and forth between CI and EI sources takes about 1 hour, however, a 1- to 2-hour wait is required to purge the reagent gas lines and bake out water and other contaminants.”
Safety

Hydrogen is combustible from 4% to 74.2% by volume at atmospheric pressure, so leak testing and supply shutoff are mandatory.

Defines the explosive concentration range requiring leak checks and ventilation.

Source
“At atmospheric pressure, hydrogen is combustible at concentrations from 4% to 74.2% by volume.”
Safety

Hydrogen supply must be off until all connections are made, and inlet/detector fittings must be connected to a column or capped whenever hydrogen is supplied.

Prevents hydrogen accumulation and explosion hazard in the GC oven.

Source
“ensure that the hydrogen supply is turned off until all connections are made, and ensure that the inlet and detector column fittings are either connected to a column or capped at all times when hydrogen gas is supplied to the instrument.”
Installation

The ferrule ID must match the column ID (e.g., 0.40 mm id ferrule for 0.25 mm id columns), so correct ferrule selection is required for a leak-free interface.

Wrong ferrule size causes leaks or column damage at the GC/MS interface.

Source
“0.40 mm id, for 0.25 mm id columns (5181-3323)”
Installation

Intuvo gasket material depends on temperature: polyimide below 350 °C, nickel from 350 °C to 450 °C.

Selecting the wrong gasket material risks failure at high operating temperatures.

Source
“Intuvo gasket, nickel 5/pk, for temperatures 350 °C to 450 °C (5190-9073)”
Installation

The column must extend 4 to 6 mm past the end of the ferrule when positioned in the inlet.

Correct column protrusion is critical for proper inlet sealing and chromatography.

Source
“Position the septum under the column nut so that the column extends 4 to 6 mm past the end of the ferrule.”
Installation

Do not break the column off inside the vacuum manifold, as pieces could be pulled into the turbo pump and damage it.

Prevents costly turbo pump damage during column installation.

Source
“Do not break the column off inside the vacuum manifold. Pieces of column could fall or be pulled into the turbo pump and damage it.”
Operation

The typical GC/MS interface setpoint is 280 °C with limits of 0 °C to 400 °C, and must never exceed the column maximum.

Guides safe interface temperature setting without exceeding column limits.

Source
“The typical setpoint is 280 °C. The limits are 0 °C to 400 °C.”
Operation

During pumpdown the turbo speed should reach 80% within 10 to 15 minutes and eventually 95%, otherwise the foreline pump shuts off and the MS must be power cycled.

Provides pass/fail criteria for verifying correct pumpdown after installation.

Source
“Within 10 to 15 minutes, the turbo pump speed should be up to 80% . The pump speed should eventually reach 95%.”
Operation

After switching from EI to CI or venting, the MS must be purged and baked out at least 2 hours before tuning, with longer bakeout for optimal sensitivity.

Ensures valid CI tuning and sensitivity after source changes.

Source
“After the source is changed from EI to CI or vented for any other reason, the MS must be purged and baked out for at least 2 hours before tuning.”
Maintenance

The standard foreline pump oil trap stops only oil droplets and does not trap toxic chemicals, so toxic work requires venting exhaust outdoors or to a fume hood.

Determines exhaust routing and safety compliance when using toxic solvents.

Source
“This trap stops only pump oil droplets. It does not trap any other chemicals.”
Maintenance

For dirty ion source parts, ultrasonically clean 15 minutes each in methylene chloride, acetone, and methanol in that order; methanol alone suffices for routine cleaning.

Defines the correct cleaning sequence to avoid contaminating source parts.

Source
“For dirty parts, use all three solvents in the order shown, cleaning 15 minutes with each of the following solvents:”

Modelos relacionados