Analysis of Total Aromatic Content

Contenido

  1. Overview p. 2
    Background and purpose of the aromatic content analysis method
  2. pp. 3-4
    Chromatographic results and discussion of aromatic species separation
  3. Procedure pp. 5-6
    Maintaining control of ASTM D5769 and signal-to-noise evaluation
  4. Quality p. 7
    Evaluation of method performance characteristics
  5. p. 8
    Summary of findings on aromatic content analysis
  6. p. 2 table, 17 rows
  7. p. 3 table, 24 rows
  8. p. 5 table, 22 rows

Table 1. GC and GC/MSD configuration, consumables, and method parameters (continued on the next page).

ParameterValue
8850 GC
Sampler7650A Automated Liquid Sampler (ALS)
InletSplit/splitless
ColumnAgilent J&W DB-1, 40 m × 0.18 mm, 0.4 µm (p/n 121-1043E)
Carrier GasHydrogen
Hydrogen Leak SensorHydrogen Sensor Module Series 2 for 8850 GC (p/n G3982A)
5977C GC/MSD
SourceExtractor
Extractor Lens9 mm (G3870-20449)
Acquisition ModeScan
Consumables
Inlet SeptumAdvanced Green, non-stick (p/n 5183-4759)
Inlet LinerDeactivated, low pressure drop split liner w/ glass wool (p/n 5183-4647)
ALS SyringeBlue Line, 5 µL, fixed needle, 23-26s/42/cone (p/n 4513-80206)
Carrier Gas FilterAgilent Gas Clean purifier kit for carrier gas, 1/8 in (p/n CP17976)
Method Parameters
Runtime9 min

Table 1. GC and GC/MSD configuration, consumables, and method parameters (continued from the previous page).

ParameterValue
Inlet, Flow Rate, and ALS
Temperature250 °C
Flow Rate1 mL/min constant flow
Septum Purge3 mL/min
Injection Volume0.3 µL
ModeSplit, 700:1
Gas SaverOn, 20 mL/min after 2 minutes
Oven Program
Initial Temperature35 °C
Initial Hold1 min
Ramp Rate25 °C/min
Final Temperature210 °C
Final Hold1 min
5977C MSD
Transfer Line Temperature250 °C
Solvent Delay2 min
Ionization Voltage70 eV
Source Temperature250 °C
Quad Temperature150 °C
Scan Range40 – 200 amu
Scan Rate8 scans/sec
Tuning Algorithmatune
Gain Factor0.3
Trace Ion Detection (TID)Enabled

Table 3. R2 values for each calibrated aromatic compound.

Peak No.ComponentR2
2Benzene0.99998
3Toluene0.99996
5Ethylbenzene0.99998
6+7p-Xylene / m-Xylene0.99999
8o-Xylene1.00000
9Isopropylbenzene0.99998
10n-Propylbenzene0.99997
111-Methyl-3-ethylbenzene0.99997
121-Methyl-4-ethylbenzene0.99993
131,3,5-Trimethylbenzene0.99996
141-Methyl-2-ethylbenzene0.99996
151,2,4-Trimethylbenzene0.99999
161,2,3-Trimethylbenzene0.99996
17Indan0.99990
18+19n-Butylbenzene/1,4-Diethylbenzene0.99992
201,2-Diethylbenzene0.99994
211,2,4,5-Tetramethylbenzene0.99997
221,2,3,5-Tetramethylbenzene0.99992
24Naphthalene0.99996
25Pentamethylbenzene0.99990
262-Methylnaphthalene0.99996
271-Methylnaphthalene0.99998

Engineering Facts

13 facts · every claim linked to a page
Selection

The 5977C ion source accepts optional 6 mm or 9 mm extraction lenses in addition to the standard 3 mm, enabling method-specific optimization.

Lens choice affects linearity and sensitivity, so it must be decided at system selection/configuration.

Source
“the 5977C GC/MSD EI Extractor ion source offers two optional extraction lenses, either 6 mm or 9 mm in diameter, enabling method‑specific optimization to improve analytical performance.”
Selection

The 9 mm extraction lens was chosen to extend measurement linearity for high-concentration bulk compounds.

If bulk aromatics like toluene must be quantified, the larger lens is required for adequate linear range.

Source
“the standard 3 mm extraction lens in the 5977C GC/MSD ion source was replaced with a 9 mm extraction lens to extend measurement linearity for bulk compounds.”
Safety

Hydrogen carrier gas requires a dedicated hydrogen leak detector on the 8850 GC.

Hydrogen use mandates leak detection hardware for safe installation and operation.

Source
“The 8850 GC was configured for hydrogen carrier gas and outfitted with the Agilent Hydrogen Leak Detector Series 2.”
Installation

The method uses a specific column (DB-1, 40 m × 0.18 mm, 0.4 µm) and an increased 0.3 µL injection volume.

Column geometry and injection volume are fixed consumable/installation choices that drive separation and sensitivity.

Source
“The column was an Agilent J&W DB-1 40 m × 0.18 mm, 0.4 µm (part number 121-1043E), and the injection volume was increased to 0.3 µL.”
Operation

To attenuate the MSD signal for high-level standards, the Gain Factor was reduced from 1.0 to 0.3.

Gain Factor setting is a key operating parameter to keep high-concentration peaks within linear detector range.

Source
“the voltage applied to the electron multiplier (EM) was reduced by decreasing the Gain Factor from 1.0 to 0.3.”
Operation

The highest calibration standard is run and gain set so the primary-ion EIC maximum is about 1 million counts.

This gives a concrete setup procedure for tuning gain to avoid detector saturation.

Source
“the highest-level calibration standard is run and the gain value set such that the maximum abundance of the extracted ion chromatogram (EIC) of the primary ion of the highest‑responding component is approximately 1 million counts.”
Performance

The method must balance reduced MS sensitivity for bulk components with enough sensitivity to measure 1,4-diethylbenzene at 100 ppm.

Engineers must verify both ends of the dynamic range, not just high-concentration linearity.

Source
“Maintaining control of ASTM D5769 is a balance between reducing the MS sensitivity for high-concentration components while still achieving enough sensitivity to measure 1,4-diethylbenzene at 100 ppm.”
Operation

Lowering the gain factor reduces signal proportionally while S/N stays roughly constant and about ten times the method requirement.

Confirms gain reduction is a safe way to extend upper range without losing required detection sensitivity.

Source
“the signal intensity decreases proportionally with gain factor, while the signal-to-noise ratio remains consistent and approximately ten times higher than the method requirement of 5.”
Maintenance

A common gain factor across multiple systems allows unified response and early maintenance indication via QC signal monitoring.

Supports fleet-wide maintenance planning and consistent system health thresholds.

Source
“This parameter also enables users to unify the response of multiple GC/MS systems running D5769 by setting a common gain factor to facilitate system health monitoring.”
Cumplimiento

The method requires ±5% relative quantitative accuracy per component, except ±10% for 1,2,4,5-tetramethylbenzene and naphthalene.

Defines the acceptance criteria against which QC results must be judged.

Source
“the method requires ± 5% relative quantitative accuracy for each component except for 1,2,4,5-tetramethylbenzene and naphthalene which both require ± 10% relative quantitative accuracy.”
Cumplimiento

Resolution of 1,3,5-trimethylbenzene and 1-methyl-2-ethylbenzene must exceed R = 2; the system achieved R = 4.77.

Critical pair resolution is a pass/fail method criterion that must be verified during setup.

Source
“resulted in R = 4.77 (Figure 4), exceeding the method requirement of R = 2.”
Performance

The RFG total aromatics result (14.46%) matched the ARV (14.82%) with a Z-score of –0.63, within one standard deviation.

Demonstrates the configured system meets interlaboratory accuracy expectations for real gasoline samples.

Source
“the RFG total aromatics result was 14.46% compared to the ARV of 14.82%, with a Z-score of –0.63, well-within one standard deviation of the results obtained by the ILCP participants.”
Performance

The full ASTM D5769 run completes in nine minutes while meeting all method requirements.

Throughput expectation for method selection and lab capacity planning.

Source
“successfully performed ASTM D5769 with a total runtime of nine minutes while meeting all method requirements.”