Blue Line, 5 µL, fixed needle, 23-26s/42/cone (p/n 4513-80206)
Carrier Gas Filter
Agilent Gas Clean purifier kit for carrier gas, 1/8 in (p/n CP17976)
Method Parameters
Runtime
9 min
Table 1. GC and GC/MSD configuration, consumables, and method parameters (continued from the previous page).
Parameter
Value
Inlet, Flow Rate, and ALS
Temperature
250 °C
Flow Rate
1 mL/min constant flow
Septum Purge
3 mL/min
Injection Volume
0.3 µL
Mode
Split, 700:1
Gas Saver
On, 20 mL/min after 2 minutes
Oven Program
Initial Temperature
35 °C
Initial Hold
1 min
Ramp Rate
25 °C/min
Final Temperature
210 °C
Final Hold
1 min
5977C MSD
Transfer Line Temperature
250 °C
Solvent Delay
2 min
Ionization Voltage
70 eV
Source Temperature
250 °C
Quad Temperature
150 °C
Scan Range
40 – 200 amu
Scan Rate
8 scans/sec
Tuning Algorithm
atune
Gain Factor
0.3
Trace Ion Detection (TID)
Enabled
Table 3. R2 values for each calibrated aromatic compound.
Peak No.
Component
R2
2
Benzene
0.99998
3
Toluene
0.99996
5
Ethylbenzene
0.99998
6+7
p-Xylene / m-Xylene
0.99999
8
o-Xylene
1.00000
9
Isopropylbenzene
0.99998
10
n-Propylbenzene
0.99997
11
1-Methyl-3-ethylbenzene
0.99997
12
1-Methyl-4-ethylbenzene
0.99993
13
1,3,5-Trimethylbenzene
0.99996
14
1-Methyl-2-ethylbenzene
0.99996
15
1,2,4-Trimethylbenzene
0.99999
16
1,2,3-Trimethylbenzene
0.99996
17
Indan
0.99990
18+19
n-Butylbenzene/1,4-Diethylbenzene
0.99992
20
1,2-Diethylbenzene
0.99994
21
1,2,4,5-Tetramethylbenzene
0.99997
22
1,2,3,5-Tetramethylbenzene
0.99992
24
Naphthalene
0.99996
25
Pentamethylbenzene
0.99990
26
2-Methylnaphthalene
0.99996
27
1-Methylnaphthalene
0.99998
Engineering Facts
13 facts · every claim linked to a page
Selection
The 5977C ion source accepts optional 6 mm or 9 mm extraction lenses in addition to the standard 3 mm, enabling method-specific optimization.
Lens choice affects linearity and sensitivity, so it must be decided at system selection/configuration.
Source
“the 5977C GC/MSD EI Extractor ion source offers two optional extraction lenses, either 6 mm or 9 mm in diameter, enabling method‑specific optimization to improve analytical performance.”
Selection
The 9 mm extraction lens was chosen to extend measurement linearity for high-concentration bulk compounds.
If bulk aromatics like toluene must be quantified, the larger lens is required for adequate linear range.
Source
“the standard 3 mm extraction lens in the 5977C GC/MSD ion source was replaced with a 9 mm extraction lens to extend measurement linearity for bulk compounds.”
Safety
Hydrogen carrier gas requires a dedicated hydrogen leak detector on the 8850 GC.
Hydrogen use mandates leak detection hardware for safe installation and operation.
Source
“The 8850 GC was configured for hydrogen carrier gas and outfitted with the Agilent Hydrogen Leak Detector Series 2.”
Installation
The method uses a specific column (DB-1, 40 m × 0.18 mm, 0.4 µm) and an increased 0.3 µL injection volume.
Column geometry and injection volume are fixed consumable/installation choices that drive separation and sensitivity.
Source
“The column was an Agilent J&W DB-1 40 m × 0.18 mm, 0.4 µm (part number 121-1043E), and the injection volume was increased to 0.3 µL.”
Operation
To attenuate the MSD signal for high-level standards, the Gain Factor was reduced from 1.0 to 0.3.
Gain Factor setting is a key operating parameter to keep high-concentration peaks within linear detector range.
Source
“the voltage applied to the electron multiplier (EM) was reduced by decreasing the Gain Factor from 1.0 to 0.3.”
Operation
The highest calibration standard is run and gain set so the primary-ion EIC maximum is about 1 million counts.
This gives a concrete setup procedure for tuning gain to avoid detector saturation.
Source
“the highest-level calibration standard is run and the gain value set such that the maximum abundance of the extracted ion chromatogram (EIC) of the primary ion of the highest‑responding component is approximately 1 million counts.”
Performance
The method must balance reduced MS sensitivity for bulk components with enough sensitivity to measure 1,4-diethylbenzene at 100 ppm.
Engineers must verify both ends of the dynamic range, not just high-concentration linearity.
Source
“Maintaining control of ASTM D5769 is a balance between reducing the MS sensitivity for high-concentration components while still achieving enough sensitivity to measure 1,4-diethylbenzene at 100 ppm.”
Operation
Lowering the gain factor reduces signal proportionally while S/N stays roughly constant and about ten times the method requirement.
Confirms gain reduction is a safe way to extend upper range without losing required detection sensitivity.
Source
“the signal intensity decreases proportionally with gain factor, while the signal-to-noise ratio remains consistent and approximately ten times higher than the method requirement of 5.”
Maintenance
A common gain factor across multiple systems allows unified response and early maintenance indication via QC signal monitoring.
Supports fleet-wide maintenance planning and consistent system health thresholds.
Source
“This parameter also enables users to unify the response of multiple GC/MS systems running D5769 by setting a common gain factor to facilitate system health monitoring.”
Cumplimiento
The method requires ±5% relative quantitative accuracy per component, except ±10% for 1,2,4,5-tetramethylbenzene and naphthalene.
Defines the acceptance criteria against which QC results must be judged.
Source
“the method requires ± 5% relative quantitative accuracy for each component except for 1,2,4,5-tetramethylbenzene and naphthalene which both require ± 10% relative quantitative accuracy.”
Cumplimiento
Resolution of 1,3,5-trimethylbenzene and 1-methyl-2-ethylbenzene must exceed R = 2; the system achieved R = 4.77.
Critical pair resolution is a pass/fail method criterion that must be verified during setup.
Source
“resulted in R = 4.77 (Figure 4), exceeding the method requirement of R = 2.”
Performance
The RFG total aromatics result (14.46%) matched the ARV (14.82%) with a Z-score of –0.63, within one standard deviation.
Demonstrates the configured system meets interlaboratory accuracy expectations for real gasoline samples.
Source
“the RFG total aromatics result was 14.46% compared to the ARV of 14.82%, with a Z-score of –0.63, well-within one standard deviation of the results obtained by the ILCP participants.”
Performance
The full ASTM D5769 run completes in nine minutes while meeting all method requirements.
Throughput expectation for method selection and lab capacity planning.
Source
“successfully performed ASTM D5769 with a total runtime of nine minutes while meeting all method requirements.”