Hydrocarbon Class Analysis of Conventional and Synthetic AviationTurbine Fuels by ASTM D8396 Flow-modulated GCxGC-FID
Technical Reference Summary
Covered Models & Document Scope
Nota de aplicación que cubre el análisis GCxGC-FID de combustibles de turbina de aviación convencionales y sintéticos según ASTM D8396 utilizando el Agilent Reverse Flow Modulator (RFM) y el Agilent 8890 GC; se comparan dos variantes del método: gas portador hidrógeno y gas portador helio.
Technical Data
| Parameter | Value | Condición |
|---|---|---|
| Sampler | Agilent 7693A Automated Liquid Sampler p.3 ↗ | Hydrogen carrier |
| Sampler | Agilent 7693A Automated Liquid Sampler p.3 ↗ | Helium carrier |
| Inlet | Split/Splitless p.3 ↗ | Hydrogen carrier |
| Inlet | Split/Splitless p.3 ↗ | Helium carrier |
| Column 1 | Agilent DB-17, 20 m × 0.18 mm, 0.18 µm p.3 ↗ | Hydrogen carrier |
| Column 1 | Agilent DB-17, 20 m × 0.18 mm, 0.18 µm p.3 ↗ | Helium carrier |
| Detector | FID p.3 ↗ | Hydrogen carrier |
| Detector | FID p.3 ↗ | Helium carrier |
| Carrier gas | Hydrogen p.3 ↗ | Hydrogen carrier |
| Runtime | 63.33 min p.3 ↗ | Hydrogen carrier |
| Runtime | 63.33 min p.3 ↗ | Helium carrier |
| Column 1 flow | 0.3 mL/min p.3 ↗ | Hydrogen carrier |
| Column 1 flow | 0.3 mL/min p.3 ↗ | Helium carrier |
| Column 2 flow | 15 mL/min p.3 ↗ | Hydrogen carrier |
| Restrictor column flow | 0.36 mL/min p.3 ↗ | Hydrogen carrier |
| Restrictor column flow | 0.36 mL/min p.3 ↗ | Helium carrier |
| Septum purge | 6 mL/min p.3 ↗ | Hydrogen carrier |
| Septum purge | 6 mL/min p.3 ↗ | Helium carrier |
| Injection volume | 0.1 µL p.3 ↗ | Hydrogen carrier |
| Injection volume | 0.1 µL p.3 ↗ | Helium carrier |
| Inlet temperature | 300 °C p.3 ↗ | Hydrogen carrier |
| Inlet temperature | 300 °C p.3 ↗ | Helium carrier |
| Oven initial temperature | 40 °C p.3 ↗ | Hydrogen carrier |
Operating Limits
- Maximum programmable temperature: 300 °C p.8 ↗